Metrosep A Supp 5 - 150/4.0

Metrosep A Supp 5 - 150/4.0

6.1006.520

Metrosep A Supp 5 - 150/4.0

en

6.1006.520

Prezentare generală a caracteristicilor

The 150 mm version of the Metrosep A Supp 5 is distinguished by its very good separation properties. High plate numbers and excellent peak symmetries facilitate working in the lower µg/L range. The particle size of 5 µm makes a decisive contribution to the separating efficiency of this column. The Metrosep A Supp 5 - 150/4.0 offers the optimum combination of selectivity and capacity, with which even complex separation tasks can be solved within a short time. This property makes the Metrosep A Supp 5 - 150/4.0 one of the best universally applicable standard IC columns.

Specificații tehnice

ApplicationRange

Applications

  • Standard anions
  • F-, Cl-, Br-, I-
  • ClO2-, ClO3-, ClO4-, BrO3-
  • Cr(VI) (CrO42-)
  • Method development
  • Difficult matrices
  • Difficult separation problems
Organic Modifier

0 - 100% (particularly acetone, acetonitrile, methanol)

Regeneration

Contamination with low-valency hydrophilic ions:

a) Rinse with ultrapure water (25 min at 0.3 mL/min)
b) Rinse with 10x concentrated eluent (100 min at 0.3 mL/min)
c) Rinse with ultrapure water (25 min at 0.3 mL/min)
d) Rinse with eluent (100 min at 0.3 mL/min)

Contaminations with high-valency hydrophobic ions and organic contaminations:

a) Rinse with ultrapure water (25 min at 0.3 mL/min)
b) Rinse with 5% acetonitrile (20 min at 0.3 mL/min)
c) Rinse with 100% acetonitrile (60 min at 0.3 mL/min)
d) Rinse with 50% acetonitrile (10 min at 0.3 mL/min)
e) Rinse with ultrapure water (50 min at 0.3 mL/min)
f) Rinse with eluent (100 min at 0.3 mL/min)

With shifted system peak (regeneration method with column oven):

a) Rinse with concentrated eluent 1 mol/L Na2CO3 (25 min at 0.4 mL/min)
b) Maintain at 45 - 50 °C for 10 - 12 hours (without rinsing)
c) Rinse with standard eluent (at least 40 min at 0.4 mL/min)

StandardEluent

Carbonate eluent (standard eluent)

Sodium hydrogen carbonate: 1.0 mmol/L; 168 mg/2 L
Sodium carbonate: 3.2 mmol/L; 678 mg/2 L

Chromate eluent

Sodium hydrogen carbonate: 4.0 mmol/L; 672 mg/2 L
Sodium carbonate: 12.8 mmol/L; 2712 mg/2 L

StorageIn the eluent
Capacity

52 µmol (Cl)

Housing materialPEEK
Carrier materialPolyvinyl alcohol with quarternary ammonium groups
Particle size5 µm
Standard flow0.7 mL/min
Maximum pressure15 MPa
Maximum flow0.8 mL/min
pH range3–12
Column dimensions150 x 4.0 mm
Temperature range20–60 °C

Cromatogramă

  • Cromatogramă
  • Cromatogramă
  • Cromatogramă
    • Metrosep A Supp 5 Guard/4.0
    • 6.1006.500
    • The Metrosep A Supp 5 Guard/4.0 reliably protects the Metrosep A Supp 5 and 7 IC anion columns against contamination from the sample or the eluent.It contains the same separation material as the Metrosep A Supp 5, is also made of PEEK, and is screwed directly onto the respective separation column with virtually no dead volume ("On Column Guard System"). The guard column prolongs the service life of the analytical column, with practically no influence on its chromatographic separating efficiency. The economical price and simple handling make using the A Supp 5 Guard/4.0 highly recommended.
    • <p>The Metrosep A Supp 5 Guard/4.0 reliably protects the Metrosep A Supp 5 and 7 IC anion columns against contamination from the sample or the eluent.<br/><br/>It contains the same separation material as the Metrosep A Supp 5, is also made of PEEK, and is screwed directly onto the respective separation column with virtually no dead volume ("On Column Guard System"). The guard column prolongs the service life of the analytical column, with practically no influence on its chromatographic separating efficiency. The economical price and simple handling make using the A Supp 5 Guard/4.0 highly recommended.</p>
    • PCS
    • 1
    • Metrosep A Supp 5 S-Guard/4.0
    • 6.1006.540
    • The Metrosep A Supp 5 S-Guard/4.0 reliably protects the Metrosep A Supp 5 and 7 IC anion columns against contaminations from the sample or eluent.
    • <p>The Metrosep A Supp 5 S-Guard/4.0 reliably protects the Metrosep A Supp 5 and 7 IC anion columns against contaminations from the sample or eluent.</p>
    • PCS
    • 1