Metrosep A Supp 5 - 150/4.0
6.1006.520
Metrosep A Supp 5 - 150/4.0
en
6.1006.520
Feature overview
The 150 mm version of the Metrosep A Supp 5 is distinguished by its very good separation properties. High plate numbers and excellent peak symmetries facilitate working in the lower µg/L range. The particle size of 5 µm makes a decisive contribution to the separating efficiency of this column. The Metrosep A Supp 5 - 150/4.0 offers the optimum combination of selectivity and capacity, with which even complex separation tasks can be solved within a short time. This property makes the Metrosep A Supp 5 - 150/4.0 one of the best universally applicable standard IC columns.
Tech specs
| ApplicationRange | Applications
|
|---|---|
| Organic Modifier | 0 - 100% (particularly acetone, acetonitrile, methanol) |
| Regeneration | Contamination with low-valency hydrophilic ions: With shifted system peak (regeneration method with column oven): |
| StandardEluent | Carbonate eluent (standard eluent) Sodium hydrogen carbonate: 1.0 mmol/L; 168 mg/2 L Chromate eluent Sodium hydrogen carbonate: 4.0 mmol/L; 672 mg/2 L |
| Storage | In the eluent |
| Capacity | 52 µmol (Cl–) |
| Housing material | PEEK |
| Carrier material | Polyvinyl alcohol with quarternary ammonium groups |
| Particle size | 5 µm |
| Standard flow | 0.7 mL/min |
| Maximum pressure | 15 MPa |
| Maximum flow | 0.8 mL/min |
| pH range | 3–12 |
| Column dimensions | 150 x 4.0 mm |
| Temperature range | 20–60 °C |