Metrosep A Supp 16 - 100/4.0
6.1031.410
Metrosep A Supp 16 - 100/4.0
en
6.1031.410
Prezentare generală a caracteristicilor
The Metrosep A Supp 16 is a high capacity separation column based on a surface-functionalized polystyrene-divinylbenzene copolymer. The functional groups are bonded covalently. The morphology of the anion exchanger results in unique selectivity. In addition, this column type is noteworthy for its high mechanical and chemical resilience.
The column is well-suited to applications with a high ionic load but which require only relatively low resolution. Using the Metrosep A Supp 16 - 100/4.0 to determine bromate in water by means of the triiodide method (EPA 326, DIN EN ISO 11206) is another of its numerous applications.
Specificații tehnice
| Housing material | PEEK |
|---|---|
| Carrier material | Polystyrene-divinylbenzene copolymer with quaternary ammonium groups |
| Particle size | 4.6 µm |
| Standard flow | 0.8 mL/min |
| Maximum pressure | 20 MPa |
| Maximum flow | 1.2 mL/min |
| pH range | 0–14 |
| Column dimensions | 100 x 4.0 mm |
| Temperature range | 10–70 °C |
Cromatogramă
-
- Threaded stopper
- 6.2744.060
- For UNF 10/32. Stopper for IC, e.g. for the sealing of columns.
- <p>For UNF 10/32. Stopper for IC, e.g. for the sealing of columns.</p>
- PCS
- 2
-
- Threaded stopper
- 6.2744.060
- For UNF 10/32. Stopper for IC, e.g. for the sealing of columns.
- <p>For UNF 10/32. Stopper for IC, e.g. for the sealing of columns.</p>
- PCS
- 2
-
- Metrosep A Supp 16 Guard/4.0
- 6.1031.500
- The Metrosep A Supp 16 Guard/4.0 reliably protects the Metrosep A Supp 16 analytical separation columns against contamination. Thanks to the "On Column Guard System", the guard column is very easy to handle. The guard column screws easily onto the analytical column. No tools are required.
- <p>The Metrosep A Supp 16 Guard/4.0 reliably protects the Metrosep A Supp 16 analytical separation columns against contamination. Thanks to the "On Column Guard System", the guard column is very easy to handle. The guard column screws easily onto the analytical column. No tools are required.</p>
- PCS
- 1
- Metrosep A Supp 16 S-Guard/4.0
- 6.1031.510
- The Metrosep A Supp 16 S-Guard/4.0 reliably protects the Metrosep A Supp 16 analytical separation columns against contamination. The guard column is connected to the separation column simply using capillary connections. No tools are required.
- <p>The Metrosep A Supp 16 S-Guard/4.0 reliably protects the Metrosep A Supp 16 analytical separation columns against contamination. The guard column is connected to the separation column simply using capillary connections. No tools are required.</p>
- PCS
- 1