Metrosep A Supp 1 - 250/4.6
6.1005.300
Metrosep A Supp 1 - 250/4.6
en
6.1005.300
Prezentare generală a caracteristicilor
The Metrosep A Supp 1 - 250/4.6 is a universal anion column that is characterized by medium capacity and special selectivity. With this column it is possible to process samples with great differences in concentrations. For example, 4 µg/L of sulfate can be determined in a solution containing 150 g/L sodium chloride. An additional advantage is that bromide does not elute until after the nitrate. Particularly in the area of oxohalide analysis, the A Supp 1 - 250/4.6 impresses with its outstanding separation properties. Pressure fluctuations, constantly changing eluents and large sample throughput do not influence the separating efficiency of this column, even after very long periods. It is the "workhorse" for development and routine laboratories.
Specificații tehnice
| Housing material | PEEK |
|---|---|
| Carrier material | Polystyrene-divinylbenzene copolymer with quaternary ammonium groups |
| Particle size | 7 µm |
| Standard flow | 1.0 mL/min |
| Maximum pressure | 15 MPa |
| Maximum flow | 2.5 mL/min |
| pH range | 1–13 |
| Column dimensions | 250 x 4.6 mm |
Cromatogramă
-
- Metrosep A Supp 1 Guard/4.6
- 6.1005.340
- The Metrosep A Supp 1 Guard/4.6 protects the Metrosep A Supp 1 - 250/4.6 separation column securely against contaminations from particles and bacteria.
- <p>The Metrosep A Supp 1 Guard/4.6 protects the Metrosep A Supp 1 - 250/4.6 separation column securely against contaminations from particles and bacteria. </p>
- PCS
- 1