Visão geral

Combined nitrate-selective electrode with polymer membrane.
This ISE is suitable for:
 

  • ion measurements of NO3- (7*10-6 to 1 mol/L) in aqueous solutions
  • titrations in aqueous solutions


Thanks to a robust/break-proof plastic shaft made of propylene and an impact protection for the polymer membrane, this sensor is mechanically very resistant.
The reference electrolyte used is c(KCl) = 3 mol/L.

Especificações técnicas

Shaft materialPMMA/PP
Measuring range (theoretical)1x10E-6...1
Measuring unitmol/L
Temperature range, long-term (°C)0...40
Temperature range, short-term (°C)0...40
pH range2.5...11
Indicator electrode shapeFlat
Max. installation length (mm)113
SGJ sleeveflexible
Shaft diameter top (mm)12
Shaft diameter bottom (mm)12
Min. immersion depth (mm)10
Electrode plug-in headMetrohm plug-in head G
Indicator electrode typePolymer (NO3)
Membrane materialPolymer
IonNO3-
Reference electrolyte typec(KCl) = 3 mol/L
DiaphragmCeramic pin
Reference systemAg/AgCl/3M KCl
Resistance (kOhm)<10
This field is required.
    • Sleeve with SGJ 14/12 mm
    • 6.1236.050
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • PCS
    • 1
    • Storage vessel 
    • 6.2008.040
    • Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.
    • <p>Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.</p>
    • PCS
    • 1
    • Sleeve with SGJ 14/12 mm
    • 6.1236.050
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • PCS
    • 1
    • Storage vessel 
    • 6.2008.040
    • Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.
    • <p>Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.</p>
    • PCS
    • 1
    • Electrode cable / 1 m / F
    • 6.2104.020
    • For connecting electrodes with Metrohm plug-in head G to Metrohm instruments (socket F).
    • For connecting electrodes with Metrohm plug-in head G to Metrohm instruments (socket F).
    • PCS
    • 1
    • Electrolyte 3 mol/L KCl (250 mL)
    • 6.2308.020
    • Electrolyte solution c(KCl) = 3 mol/L, (for Ag/AgCl reference systems)
    • Electrolyte solution c(KCl) = 3 mol/L, (for Ag/AgCl reference systems)
    • PCS
    • 1
    • Electrolyte 3 mol/L KCl (50 mL)
    • 6.2308.050
    • Electrolyte solution c(KCl) = 3 mol/L (for Ag/AgCl reference systems)
    • Electrolyte solution c(KCl) = 3 mol/L (for Ag/AgCl reference systems)
    • PCS
    • 1