Du er blevet omdirigeret til din lokale version af den anmodede side
How to transfer manual titration to autotitration

How to transfer manual titration to autotitration

11. maj 2026

Artikel

Manual titration is a common analysis method in many labs. However, it has several drawbacks, such as subjective endpoint detection, risk for manual errors, or lack of data integrity. In this article, learn the most important steps to consider when converting manual titration to automatic titration, from the choice of the titrator to the optimization of the method itself. 

To jump directly to a topic, click on the link below:

Disadvantages of manual titration

Manual titration involves adding titrant dropwise from a buret to a sample solution. Endpoint detection is typically done visually (by eye) using color indicators. This analytical method remains widely used because it requires only a glass buret and a flask. However, it comes with several notable disadvantages.

  • Subjective endpoint detection: Visual recognition of the color change depends on individual perception. Each person may interpret color intensity differently, which leads to inconsistent results. This issue becomes more pronounced when working with colored or turbid solutions.
  • Drop size accuracy: The precision of the analysis is influenced by the size of each drop. For instance, if the drop size is 50 µL and the total titrant volume is 5 mL, a single drop can cause a variation of approximately 1%.
  • Time-consuming process: Manual titration takes considerable time. This includes not only the titration itself, but also cleaning and refilling the buret and manually calculating the results. These steps limit the number of samples that can be processed during a shift. 
  • Data integrity risks: When performing manual titrations, results are often recorded by hand, transferred to spreadsheets, and then manually entered into a LIMS. Each step introduces potential for human error and requires thorough quality control checks. This process is both tedious and inefficient.

Using an automatic titrator (autotitrator) helps operators to overcome these limitations.

What is autotitration?

Autotitration refers to the automation of one or more steps in the titration process. There are three levels of automation, depending on the setup.

  1. Partial automation: Only a single step is automated. For example, an electrode is used for objective endpoint recognition, or the titrant is added manually using a more precise dosing device.
    Metrohm offers partial automation within the OMNIS platform, allowing full traceability for manual titrations.
  2. Stand-alone automation: A titrator performs the entire titration, including titrant addition, endpoint detection, and result calculation. The operator is still responsible for preparing the sample and cleaning the equipment afterwards. Figure 1 shows a stand-alone titrator.
  3. Fully automated titration: The titrator is combined with an autosampler, allowing full automation of the titration including sample preparation and equipment cleaning. This setup enables the operator to focus on other tasks and significantly reduces human error. 

Learn more about what can be automated in the blog post «Why consider automation – even for simple titrations» and how automation can benefit you in the blog post «Save money by using automated titration systems».

Anatomy of an automatic titrator.
Figure 1. Stand-alone automatic titrator with piston buret (1), titration stand with integrated stirrer (2), electrode (3) for endpoint detection, and display (4) for control and results.

Automated titration offers several advantages. Table 1 provides a comparison between manual and automatic titration.

Table 1. Key differences between manual and automatic titration.
Parameter Manual titration Automatic titration
Titrant addition Manually Automatically with piston buret
Dosing accuracy 50 µL (50 mL buret class AS)
20 µL (10 mL buret class AS)
25 µL (50 mL buret)
7 µL (10 mL buret)
Endpoint detection Visually by eye Using a sensor and mathematical algorithm
Control Manually by operator Integrated touch display or with software
Calculation Manually by operator Integrated in system or with software
Data integrity No Yes (optional)
Automation possibility No Yes

Choice of titrator

When choosing a titrator, you should take several key points into account.

  • GLP/GMP compliance: Autotitrators must be able to document results. This includes providing options for a printer connection or digital storage (e.g., USB flash drive). All Metrohm titrators meet GLP and GMP requirements.
  • FDA compliance: In regulated environments, titrators must comply with FDA Regulation 21 CFR Part 11. 
  • Control options: Titrators can be controlled either via software or an integrated touch display.
  • Sample throughput: If a high number of samples are analyzed daily, an automated system can significantly improve laboratory efficiency. Learn more about automation in the article «Why consider automation – even for simple titrations».
  • Flexibility for different titrations: Switching between titration types, for example from redox to acid-base titration, requires changing both titrant and electrode. Not all titrators support this. Efficiency improves with titration systems that can be extended with additional burets and electrodes.

Metrohm offers several titrator models with different capabilities. Table 2 provides a comparison of the key points mentioned above for each titrator family.

Table 2. Comparison of Metrohm titrator families based on the key selection criteria.
Parameter Eco Titrator Ti-Touch OMNIS Titrator
GLP/GMP compliance Yes Yes Yes
FDA compliance No Yes Yes
Control option Touch display Touch display Software
Automation possibility Yes
(max. 9 samples)
Yes
(max. 24 samples)
Yes
(max. 175 samples, parallel titrations)
Extension for different titrations No Limited
(two sequential titrations)
Extensive
(up to five titrations in parallel)

Choice of sensor

Optrode M2, optical sensor for photometric titrations.
Figure 2. Optrode M2, optical sensor for photometric titrations.

The most crucial step in transferring a manual titration to automatic titration is selecting the appropriate sensor for detecting the equivalence point.

A straightforward option is to use a photometric sensor, which effectively replaces the human eye. This approach is particularly useful when norms or standards specify the use of color indicators. Figure 2 shows an optical sensor for photometric titrations.

Using a potentiometric electrode is often simpler, as it does not require an indicator solution. However, selecting the correct electrode depends on several factors.

  • Chemical reaction: The chemical reaction determines the choice of electrode. For example, acid-base titrations require a pH electrode, while redox titrations require a metal electrode.
  • Sample matrix: The sample matrix can significantly affect the electrode performance. For example, protein-containing samples (e.g., dairy products) require a pH electrode with a diaphragm that is insensitive to proteins.
  • Sample volume: For small sample volumes, specialized micro electrodes are recommended.

More information about different electrodes for various titration types is available in the article «Best practice for electrodes in titration». 

The flyer «Electrodes for titration» offers guidance on selecting the most suitable electrode. Alternatively, the Electrode Finder allows filtering by the reaction type, application area, or specific details, such as sample matrix.

Optimizing the sample size and solution volumes

When performing manual titrations, reaching the endpoint often requires 30 mL or even 40 mL of titrant. Autotitrators, which have a higher dosing accuracy, are typically equipped with 10 mL or 20 mL burets. Because refilling the buret during titration causes errors, the sample size should be adjusted for automatic titrations. 

For automatic titrators, it is generally recommended that the equivalence point (or expected endpoint volume) should lie between 10% and 90% of the total buret volume. Therefore, a critical step when transferring manual titration to autotitration is the optimization of the sample size. The lower titrant consumption can also save costs.

For example, in the assay of calcium hydroxide, the sample size can be reduced from 1.5 g to 0.375 g. This adjustment decreases the titrant consumption by 30 mL per titration [1]. With an estimated 10 titrations per day, this results in annual savings of approximately €4700 per year on titrant alone. See Table 3 for a detailed example calculation.

Table 3. Example calculation of cost savings from reduced titrant consumption when switching from manual to autotitration. Savings may vary by country and reagent choice.
  Manual titration Automatic titration
Titrant 0.05 mol/L EDTA 0.05 mol/L EDTA
Costs of titrant €43.40  per liter * €43.40  per liter *
Sample size 1.5 g 0.375 g
Expected endpoint volume 40 mL 10 mL
Estimated titrations per day 10 10
Costs of titrant per year €6336.40 €1584.10
Cost savings   €4752.30
* (https://www.sigmaaldrich.com/DE/de/product/mm/160320, April 23, 2026 for Germany)
Close-up view of a pH electrode fully immersed in a clear liquid
Figure 3. For accurate pH measurements, both the glass membrane and diaphragm must be immersed in titration solution.

In addition to adjusting the sample size, it might be necessary to modify the amount of diluent (water or solvent) used in the analysis. Accurate results depend on full immersion of the relevant parts of the sensor in the solution. For example, with a pH electrode, both the glass membrane (for measurement) and diaphragm (for reference) must be fully immersed, as shown in Figure 3.

Selecting the right titration mode

Some titration reactions are reversible, e.g., acid-base titrations, while others are irreversible, e.g., redox titrations. The speed of the titration reaction can also vary. Therefore, autotitrators are equipped with different titration modes.

There are three commonly used titration modes:

  1. Endpoint titration: Titrant is added until a predefined endpoint is reached. This mode is commonly used with a specific pH value as the endpoint.
  2. Monotonic titration: A constant volume of titrant is added with each dosing step.
  3. Dynamic titration: The volume of titrant added varies depending on the proximity to the equivalence point. Smaller titrant volumes are added incrementally around the equivalence point.
0:00 / 0:00

This video explains the difference between monotonic and dynamic titration in more detail.

Dynamic mode is recommended for fast titrations, such as acid-base titrations. Monotonic mode is more suitable for slower titrations where the equivalence point is reached abruptly, such as in vitamin C determination, as the dynamic mode would overshoot such equivalence points.

Optimizing the titration setup and method

Optimizing the titration setup and method helps to improve accuracy, speed up the titration, and reduce titrant consumption.

Electrode placement relative to stirring direction. Left: downstream of the buret tip, resulting in inaccurate readings. Right: upstream of the buret tip, allowing titrant to react with analyte before reaching the electrode.
Figure 4. Electrode placement relative to stirring direction. Left: downstream of the buret tip, resulting in inaccurate readings. Right: upstream of the buret tip, allowing titrant to react with analyte before reaching the electrode.

Stirrer setup for higher accuracy

Both stirring speed and sensor placement within the titration beaker directly affect the accuracy of the results. The following points should be considered:

  • Select a stirring speed that ensures thorough mixing without splashing. The optimal speed depends on the sample beaker and stirrer type.
  • Avoid creating a vortex which can leave the electrode hanging in the air and decrease accuracy. 
  • Position the electrode close to the beaker wall and upstream of the buret tip, as illustrated in Figure 4. This setup promotes effective mixing of the titrant with the sample and improves accuracy.

Start volume to speed up the titration

To speed up titrations – especially monotonic titrations – a start volume can be used. This approach resembles the pre-dosing of titrant in manual titration. A pause step should follow the addition of the start volume to allow the titrant to react with the analyte before the titration begins.

Stop volume to save titrant

Applying a stop criterion helps to reduce titrant consumption and minimize waste. The simplest option is a fixed stop volume. This is effective when the equivalence point consistently occurs around the same volume.

For samples with varying analyte concentrations, a more flexible solution is required. In such cases, the maximum number of equivalence points is used as the stop criterion. Additionally, a defined volume should be added afterwards.

In both cases, a stop volume of approximately 1 mL after the equivalence point is recommended.

Validation of the method

The final step in converting manual titration to automatic titration is method validation – especially important in regulated environments. The validation of a titration method includes standardization of the titrant as well as determination of accuracy and precision, linearity, and specificity. For more information about the validation of titration methods, read our blog post «Validation of titration methods».

Conclusion

When transitioning from manual to autotitration, laboratories should consider several key factors. At the same time, this change provides an opportunity to optimize the titration process and reduce waste. The infographic in Figure 5 summarizes the essential steps involved in method transfer:

Summary of key steps for transferring manual titration to autotitration.
Figure 5. Summary of key steps for transferring manual titration to autotitration.
  1. Select a titrator based on requirements regarding compliance, instrument control, sample throughput, and reaction type.
  2. Choose a sensor suitable for the titration reaction, sample matrix, and sample volume.
  3. Adjust the sample size so that the equivalence point lies between 10% and 90% of the buret volume. If necessary, modify the diluent volume to ensure that the sensor is fully immersed.
  4. Select the titration mode based on the titration reaction. Dynamic mode is recommended for fast titrations. Monotonic mode is better suited for slower titrations where the equivalence point is reached abruptly.
  5. Optimize the titration by placing the electrode upstream of the buret tip, selecting a stirring speed that does not lead to vortex formation, and defining start and stop volumes.
  6. Validate the method to ensure reproducibility and reliability.

Detailed examples for steps 2 to 5 are provided in White Paper WP-063 «Recommendations for converting a manual titration procedure into an automated titration procedure».

For more information, watch our webinar «How to convert from manual to automatic titrations».

Reference

[1] Marques, M. R. C.; Pappa, H.; Chang, M.; Spafford, L.; Klein, M.; Meier, L. Recommendations for Converting a Manual Titration Procedure into an Automated Titration Procedure; White Paper WP-063; Metrohm: Herisau, 2021.

Recommendations for converting a manual titration procedure into an automated titration procedure

Click here to download

We have compiled a comprehensive White Paper and several application documents that address typical questions that arise with the change from manual to automated titration. This White Paper is intended to help follow the recommendations by USP.

Author
Meier

Lucia Meier

Technical Editor
Metrohm International Headquarters, Herisau, Switzerland

Kontakt