AN-NIR-068
2026-06
Isocyanates quality control with NIRS
Fast, chemical-free determination of isocyanate content for polyurethane production
Summary
Analyzing NCO content in isocyanates as raw materials for polyurethane (PU) is crucial to control properties of the final polymer. Wet analytical methods such as those used in ASTM D7252 are challenging due to the reactivity of these organic species with atmospheric moisture, as well as their toxicity. Furthermore, analysis of NCO content with HPLC (high-performance liquid chromatography) involves many sample preparation steps and chemical reagents, with each measurement taking up to 20 minutes to complete.
This Application Note demonstrates that near-infrared spectroscopy (NIRS) provides a reagent-free and faster alternative to determine NCO content in isocyanates. The OMNIS NIR Analyzer Liquid analyzes isocyanates in a few seconds without any sample preparation.
Experimental equipment
Liquid samples of isocyanates were measured with an OMNIS NIR Analyzer Liquid in transmission mode over a wavelength range of 1000–2250 nm. Reproducible spectrum acquisition was achieved using the built-in temperature control at 30 °C. For convenience, disposable vials with a path length of 8 mm were used. OMNIS Software was used for all data acquisition and prediction model development.
| Equipment | Article number |
|---|---|
| OMNIS NIR Analyzer Liquid | 2.1070.0010 |
| Holder OMNIS NIR, vial, 8 mm | 6.07401.070 |
| Disposable vial, 8 mm, transmission | 6.7402.240 |
| OMNIS Stand-Alone | 6.06003.010 |
| Software license Quant Development | 6.06008.002 |
Result
The NIR spectra (Figure 2), along with the corresponding reference values, were used to create a prediction model for isocyanate quantitative analysis. The quality of the prediction model was evaluated using the correlation diagram shown in Figure 3 which displays a very high correlation between the NIR prediction and the measured values using HPLC. The respective figures of merit (FOM) display the expected precision of a prediction during routine analysis.
Result NCO content in isocyanates
| R2 | SEC (%) | SECV (%) | SEP (%) |
|---|---|---|---|
| 0.983 | 0.30 | 0.33 | 0.39 |
Conclusion
This Application Note shows the feasibility of using NIR spectroscopy for the quality control of isocyanates for PU production. NCO content can be determined in only a few seconds without any sample preparation.
Compared to wet chemical methods (Table 2), the time to result is a major advantage of NIR spectroscopy. Also, no chemicals are required, helping to better adhere to environmental, chemical, and health regulations.
| Parameter | Method | Time to result |
|---|---|---|
| Isocyanate content | HPLC | ∼20 min (preparation) + ∼20 min (HPLC) |