元の言語のページへ戻りました
This bulletin deals with the automated determination of  mixtures of HNO3, HF and H2SiF6 in the range of approximately 200-600 g/L HNO3, 50-160 g/L HF, and 0-185 g/L H2SiF6 using thermometric titration.Etch acid mixtures containing HNO3, HF and H2SiF6 from the etching of silicon substrates can be analyzed in a sequence of two determinations using the 859 Titrotherm. The first determination involves a direct titration with standard c(NaOH) = 2 mol/L, followed by a back titration with c(HCl) = 2 mol/L. This determination yields the H2SiF6 content plus a value for the combined (HNO3+HF) contents. The second determination consists of a titration with c(Al3+) = 0.5 mol/L to determine the HF content. For freshly made up mixtures of HNO3 and HF containing no H2SiF6, a linked two-titration sequence is employed. Results from the two determinations are used by tiamoTM to yield individual results for HNO3, HF and H2SiF6.
お問い合わせ

メトロームジャパン株式会社

143-0006 東京都大田区平和島6-1-1
東京流通センター アネックス9階

お問い合わせ