Determination of diethylene glycol, isophthalic acid, carboxyl end group, and more in PET with NIRS

Simultaneously monitor multiple quality parameters in virgin and recycled polyethylene terephthalate without chemicals


Summary

Determining diethylene glycol content, isophthalic acid content, intrinsic viscosity (ASTM D4603), and carboxyl end group (CEG) in polyethylene terephthalate (PET) is a lengthy and challenging process due to the sample’s limited solubility and the need to use different analytical methods. NIR spectroscopy (NIRS) can analyze PET in a few seconds without sample preparation or using any chemical reagents. This study shows that the OMNIS NIR Analyzer provides a fast, cost-efficient solution to simultaneously determine all these QC parameters in PET.

 


Experimental equipment

OMNIS NIR Analyzer Solid with large cup OMNIS NIR and polymer sample.
Figure 1. OMNIS NIR Analyzer Solid with large cup OMNIS NIR and polymer sample.

In this study, 90 samples of virgin PET in pelletized form with different intrinsic viscosity, diethylene glycol, isophthalic acid, and carboxyl end group values were measured with an OMNIS NIR Analyzer Solid (Figure 1) in diffuse reflection mode (1000–2250 nm) using a large cup OMNIS NIR, 100 mm and large holder OMNIS NIR, 100 mm. To compensate for the inhomogeneity of the samples, rotation and multipoint measurement was used.

Additionally, 50 samples of recycled PET pellets and 200 samples of recycled PET flakes were measured under the same conditions as previously mentioned.  

All data acquisition and prediction model development was performed with OMNIS Software.


Configuration


Result

The near-infrared spectra of virgin PET and recycled PET (Figures 2–4), along with the corresponding reference values, were used to create prediction models for quantification of all the quality control parameters listed earlier.

The quality of the prediction models was evaluated using correlation diagrams (Figures 5–10) which display the relation between the NIR prediction and the reference values of intrinsic viscosity, carboxyl end group, isophthalic acid, and diethylene glycol content using different analytical methods. The respective figures of merit (FOM) display the expected precision during routine analysis.

Selection of NIR spectra of virgin PET samples in pellet form. Data was obtained with an OMNIS NIR Analyzer Solid.
Figure 2. Selection of NIR spectra of virgin PET samples in pellet form. Data was obtained with an OMNIS NIR Analyzer Solid.
Selection of NIR spectra of recycled PET samples in pellet form. Data was obtained with an OMNIS NIR Analyzer Solid.
Figure 3. Selection of NIR spectra of recycled PET samples in pellet form. Data was obtained with an OMNIS NIR Analyzer Solid.
Selection of NIR spectra of recycled PET samples in flake form. Data was obtained with an OMNIS NIR Analyzer Solid.
Figure 4. Selection of NIR spectra of recycled PET samples in flake form. Data was obtained with an OMNIS NIR Analyzer Solid.

Result intrinsic viscosity in virgin PET

Correlation diagram and the respective FOMs for the prediction of intrinsic viscosity in virgin PET pellets using an OMNIS NIR Analyzer Solid.
Figure 5. Correlation diagram and the respective FOMs for the prediction of intrinsic viscosity in virgin PET pellets using an OMNIS NIR Analyzer Solid.
R2 CVSEC (dL/g)SECV (dL/g)
0.8681.701.75

Result isophthalic acid content in virgin PET

Correlation diagram and the respective FOMs for the prediction of isophthalic acid content in virgin PET pellets using an OMNIS NIR Analyzer Solid. HPLC was the lab method used to determine isophthalic acid in PET.
Figure 6. Correlation diagram and the respective FOMs for the prediction of isophthalic acid content in virgin PET pellets using an OMNIS NIR Analyzer Solid. HPLC was the lab method used to determine isophthalic acid in PET.
R2 CVSEC (%)SECV (%)
0.9930.060.07

Result diethylene glycol content in virgin PET

Correlation diagram and the respective FOMs for the prediction of diethylene glycol content in virgin PET pellets using an OMNIS NIR Analyzer Solid. The lab method used for diethylene glycol content in PET was extraction followed by analysis by HPLC-MS.
Figure 7. Correlation diagram and the respective FOMs for the prediction of diethylene glycol content in virgin PET pellets using an OMNIS NIR Analyzer Solid. The lab method used for diethylene glycol content in PET was extraction followed by analysis by HPLC-MS.
R2 CVSEC (%)SECV (%)
0.9250.040.04

Result carboxyl end group content in virgin PET

Correlation diagram and the respective FOMs for the prediction of carboxyl end group content in virgin PET pellets using an OMNIS NIR Analyzer Solid. Titration was the lab method used for carboxyl end group analysis of PET.
Figure 8. Correlation diagram and the respective FOMs for the prediction of carboxyl end group content in virgin PET pellets using an OMNIS NIR Analyzer Solid. Titration was the lab method used for carboxyl end group analysis of PET.
R2 CVSEC (meq/kg)SECV (meq/kg)
0.8522.913.30

Result intrinsic viscosity in recycled PET (pellet form)

Correlation diagram and the respective FOMs for the prediction of intrinsic viscosity in recycled PET pellets using an OMNIS NIR Analyzer Solid.
Figure 9. Correlation diagram and the respective FOMs for the prediction of intrinsic viscosity in recycled PET pellets using an OMNIS NIR Analyzer Solid.
R2 CVSEC (dL/g)SECV (dL/g)
0.8810.00700.0079

Result intrinsic viscosity in recycled PET (flakes form)

Correlation diagram and the respective FOMs for the prediction of intrinsic viscosity in recycled PET flakes using an OMNIS NIR Analyzer Solid.
Figure 10. Correlation diagram and the respective FOMs for the prediction of intrinsic viscosity in recycled PET flakes using an OMNIS NIR Analyzer Solid.
R2 CVSEC (dL/g)SECV (dL/g)
0.9050.00880.0102

Conclusion

This study demonstrates the feasibility of using NIR spectroscopy for the analysis of key quality parameters of PET. An important benefit of NIR spectroscopy PET analysis is that no sample preparation is required. When compared to wet chemical methods (Table 1), the time to result is a major advantage of NIR spectroscopy, since all parameters are determined in a single measurement in a few seconds.  

Table 1. Time to result overview for the different parameters analyzed in this study.
ParameterMethodTime to result
Diethylene glycolExtraction + analysis HPLC-MS∼45 min (preparation) + ∼40 min (HPLC)
Isophthalic acidDissolve + HPLC∼45 min (preparation) + ∼40 min (HPLC)
Intrinsic viscosityDissolve + viscometry∼90 min (preparation) + ∼1 min (viscometry)
Carboxyl end groupDissolve + titration∼90 min (preparation) + ∼10 min (titration)