Cleaning after the determination

Cleaning the oxidation stability accessories

0:00 / 0:00

Watch this video to learn how to clean your oxidation stability accessories. More information can be found below.

Reaction vessel (6.1429.040, 6.1429.050) and air tube (6.2418.100, 6.2418.120, 6.2418.130)

support-and-service/stab-maintenance-after-determination/_1

Reaction vessels and air tubes are disposable and should not be cleaned and reused. This is because:

  • Removing sample residue is extremely time-consuming and therefore costly.
  • Cleaning procedures that are typically applied damage the glass. The altered glass surface impairs the reproducibility and accuracy of the results.
  • Poorly cleaned reaction vessels lead to scattered and, in some cases, even incorrect results.

Reaction vessel cover (6.2753.107)

support-and-service/stab-maintenance-after-determination/_2

During the determination, the plastic of the reaction vessel cover is contaminated by both the adsorption and absorption of sample components.

We recommend that you take the following steps when cleaning after each determination:

  1. Clean reaction vessel cover with hot water, dishwashing detergent, and a brush. Rinse with water.
  2. Clean reaction vessel cover with a solvent (acetone, 2-propanol, or an acetone/ethanol mixture (1:1)) and a brush. Rinse thoroughly with deionized water.
  3. Heat the reaction vessel cover in a drying oven at 80 °C for at least 2 hours. This step is important for releasing and removing absorbed compounds.


Dishwasher-safe

You can also clean the reaction vessel cover in a laboratory dishwasher instead of carrying out steps 1 and 2. Carry out step 3 as described above.

 

Connection tubing (6.1816.010, 6.1839.000)

https://s7e5a.scene7.com/is/image/metrohm/Stab%20Connection%20tubing?qlt=85&ts=1644584116011&dpr=off

During the determination, the plastic of the connection tubing between the reaction vessel and measuring vessel (silicone tubing, Iso-Versinic tubing) is contaminated by both the adsorption and absorption of sample components.

We recommend that you take the following steps when cleaning after each determination:

  1. Clean connection tubing with hot water, dishwashing detergent, and a brush. Rinse with water.
  2. Clean connection tubing with a solvent (acetone, 2-propanol, or an acetone/ethanol mixture (1:1)) and a brush. Rinse thoroughly with deionized water.
  3. Heat the connection tubing in a drying oven at 80 °C for at least 2 hours. This step is important for releasing and removing absorbed compounds.


Dishwasher-safe

You can also clean the connection tubing in a laboratory dishwasher instead of carrying out steps 1 and 2. Carry out step 3 as described above.

Measuring vessel cover (6.0913.130)

support-and-service/stab-maintenance-after-determination/_4

Volatile sample components also get into the measuring cell during the determination. There they adsorb on both the plastic surface of the cover and the conductivity measuring cell.

We recommend that you take the following steps when cleaning after each determination:

  1. Remove protection ring. Clean the measuring vessel cover with hot water, dishwashing detergent, and a brush. Also thoroughly clean both steel electrodes. Rinse with water. Proceed with the protection ring in the same way.
  2. Clean measuring vessel cover with a solvent (acetone, 2-propanol, or an acetone/ethanol mixture (1:1)) and a brush. Rinse thoroughly with deionized water. Clamp the protection ring back in place.

Dishwasher-safe
You can also clean the measuring vessel cover and the protection ring in a laboratory dishwasher instead of carrying out steps 1 and 2

Measuring vessel (6.1428.107, 6.1428.030)

support-and-service/stab-maintenance-after-determination/_5

Volatile sample components also get into the measuring cell during the determination. 

There they adsorb on the surface of the measuring vessel. We recommend that you take the following steps when cleaning after each determination:

Glass measuring vessel (6.1428.030)

  1. Clean measuring vessel with hot water, dishwashing detergent, and a brush. Rinse with water.
  2. Clean measuring vessel with a solvent (acetone, 2-propanol, or an acetone/ethanol mixture (1:1)) and a brush. Rinse thoroughly with deionized water.

Dishwasher-safe
You can also clean the measuring vessel in a laboratory dishwasher instead of carrying out steps 1 and 2.

Polystyrene measuring vessel (6.1428.107)

  1. Clean measuring vessel with hot water, dishwashing detergent, and a brush. Rinse with water.
  2. Clean measuring vessel with a solvent (ethanol or 2-propanol) and a brush. Rinse thoroughly with deionized water. Note! Polystyrene is not chemically resistant to acetone. Do not use acetone for cleaning.

Dishwasher-safe
You can also clean the measuring vessel in a laboratory dishwasher instead of carrying out steps 1 and 2.

Foam barrier (6.1451.010) 

support-and-service/stab-maintenance-after-determination/_6

During the determination, the plastic of the foam barrier is contaminated by both the adsorption and absorption of sample components.

We recommend that you take the following steps when cleaning after each determination:

  1. Clean foam barrier with hot water, dishwashing detergent, and a brush. Rinse with water.
  2. Clean foam barrier with a solvent (acetone, 2-propanol, or an dacetone/ethanol mixture (1:1)) and a brush. Rinse thoroughly with deionized water.
  3. You can also clean the connection tubing in a laboratory dishwasher instead of carrying out steps 1 and 2. Carry out step 3 as described above..

Dishwasher-safe
You can also clean the connection tubing in a laboratory dishwasher instead of carrying out steps 1 and 2. Carry out step 3 as described above.

Oil separator (6.2753.200)

support-and-service/stab-maintenance-after-determination/_7

During the determination, the oil separator (body and base) is contaminated by both the adsorption and absorption of sample components. 

We recommend that you take the following steps when cleaning after each determination:

  1. Clean body and base with hot water, dishwashing detergent, and a brush. Rinse with water.
  2. Clean foam barrier with a solvent (acetone, 2-propanol, or an dacetone/ethanol mixture (1:1)) and a brush. Rinse thoroughly with deionized water.
  3. Heat body and base in a drying oven at 80 °C for at least 2 hours. This step is important for releasing and removing absorbed compounds.

Dishwasher-safe
You can also clean the oil separator (body and base) in a laboratory dishwasher instead of carrying out steps 1 and 2. Carry out step 3 as described above.

Events
3 Nov 8 Dec
Innovation Open House
Experience Metrohm solutions in action and connect with our experts. Discover the latest in ion chromatography, titration, spectroscopy, electrochemistry, and process analytics, including the new OMNIS IC and IONIC Titrator.
4 Nov
From Reaction to Release: Raman and NIR Analysis for Fertilizer Production
Fertilizer production depends on timely chemical measurements, but reference techniques such as titration and ICP-OES can require sample transport, preparation, reagents, specialized analysts, and significant time before results are available. Raman and near-infrared spectroscopy can provide faster, simpler analysis closer to production while reducing reagent use and cost per measurement. Using examples from phosphate, urea, ammonium nitrate fertilizer production this webinar will explore the development of quantitative spectroscopic methods, from collecting representative production samples and reliable reference data through calibration and independent validation. The session will also examine method transfer and deployment across instruments and production sites. Adam J. Hopkins, Ph.D., Product Manager for Spectroscopy at Metrohm USA, will introduce Raman and NIR fundamentals and the Metrohm products, sampling options, and software used for production measurements. Hannah Schorr, Ph.D., Applications Specialist for Spectroscopy at Metrohm USA, will then present the analytical science behind the fertilizer applications, including how quantitative methods are developed, validated, transferred, and maintained for reliable routine analysis.