Sie wurden zu Ihrer lokalen Version der angefragten Seite umgeleitet

Scope:

Determination of cuprous ion in the presence of ferrous ion in electrochemical copper leach solutions

Principle:

Acidified aliquots of electrochemically-derived copper leach feed solution containing iron are titrated against standard potassium dichromate to a second exothermic endpoint. The presence of ferrous ion appears to protect the cuprous ion against oxidation. The cuprous content is determined from the difference between the first (ferrous) and second (cuprous) endpoints

Cr2O72− +14H+ +6e− ⇔2Cr3+ +7H2O [Cu+ ⇔Cu2+ +e−]×6

Cr2O72− +14H+ +6Cu+ ⇔6Cu2+ +2Cr3+ +7H2O

Bestimmung von Kupferionen in Anwesenheit von Eisenionen in elektrochemischen Kupferauslaugungslösungen.

Kontakt