Vous avez été redirigé vers la version locale de la page demandée

Scope:

Determination of cuprous ion in the presence of ferrous ion in electrochemical copper leach solutions

Principle:

Acidified aliquots of electrochemically-derived copper leach feed solution containing iron are titrated against standard potassium dichromate to a second exothermic endpoint. The presence of ferrous ion appears to protect the cuprous ion against oxidation. The cuprous content is determined from the difference between the first (ferrous) and second (cuprous) endpoints

Cr2O72− +14H+ +6e− ⇔2Cr3+ +7H2O [Cu+ ⇔Cu2+ +e−]×6

Cr2O72− +14H+ +6Cu+ ⇔6Cu2+ +2Cr3+ +7H2O

Détermination de l'ion cuivreux en présence de l'ion ferreux dans des solutions de lixiviation de cuivre.

Contact