MiPT – Metrohm intelligent Partial Loop Injection Technique

AN-S-269

vi

<p>Calibration of fluoride, chloride, nitrite, bromide, nitrate, phosphate, and sulfate using intelligent partial loop injection technique and anion chromatography with conductivity detection after sequential suppression. This technique allows a calibration range of 1:100 (e.g. 1 &mu;g/L to 100 &mu;g/L corresponding to 2 &mu;L to 200 &mu;L injected volume) out of 1 calibration solution. Applying the full range of partial loop injection to the samples one calibration covers a sample concentration range of 1 to 10'000.</p>

Calibration of fluoride, chloride, nitrite, bromide, nitrate, phosphate, and sulfate using intelligent partial loop injection technique and anion chromatography with conductivity detection after sequential suppression. This technique allows a calibration range of 1:100 (e.g. 1 μg/L to 100 μg/L corresponding to 2 μL to 200 μL injected volume) out of 1 calibration solution. Applying the full range of partial loop injection to the samples one calibration covers a sample concentration range of 1 to 10'000.