Column stability of the Metrosep C 6 - 250/4.0 with Inline Ultrafiltration and Inline Eluent Preparation

AN-C-158

vi

<p>The column stability of the Metrosep C 6 - 250/4.0 was determined in long-term laboratory tests. Two injection series per day were run on each of six days in a row. Each series was comprised of nine tap water injections, three check standard injections and six tap water injections. The IC system was shut down on the seventh day of each series. As a whole, the system ran over 10 weeks and counted a total of 2,150 injections. The results show an outstanding reproducibility and verify the high column stability.</p>

The column stability of the Metrosep C 6 - 250/4.0 was determined in long-term laboratory tests. Two injection series per day were run on each of six days in a row. Each series was comprised of nine tap water injections, three check standard injections and six tap water injections. The IC system was shut down on the seventh day of each series. As a whole, the system ran over 10 weeks and counted a total of 2,150 injections. The results show an outstanding reproducibility and verify the high column stability.