Glass electrode with sodium-selective membrane glass and SGJ.
This ISE has to be used in combination with a reference electrode and is suitable for:
- ion measurements of Na+ (10-5 to 1 mol/L)
- titrations
This electrode is also suitable for difficult matrices, such as fruit/vegetable juices, and samples that contain chlorate or perchlorate.
Bu ürün artık üretilmemekte olup destek hizmeti hala devam etmektedir.
Muadil ürün grubuna git
Daha yakından tanıyın Detail
Ekipmanlar Size özel Sodium-selective glass electrode
Aksesuarlar PDF dosyasını indir

- Opsiyonel parçalar
Electrode cable / 1 m / F
6.2104.020
For connecting electrodes with Metrohm plug-in head G to Metrohm instruments (socket F).
LL ISE reference electrode
6.0750.100
Silver / silver chloride reference electrode with double junction system.This reference electrode is well suited for:automated applications; ion measurements; surfactant titrations; The ground-joint diaphragm, which is insensitive to contamination, offers a constant and reproducible electrolyte outflow. Additionally, the reference electrolyte is gelified for even better signal stability. The sensor is delivered with c(KCl) = 3 mol/L as bridge electrolyte, which can be freely selected and exchanged as needed.