Profitrode (length 12.5 cm)
6.0255.100
Profitrode (length 12.5 cm)
en
6.0255.100
Överblick över funktioner
Combined pH electrode with double-junction construction, installation length 11.3 cm. This electrode is suitable for pH measurements/titrations of samples:
- that contaminate the reference system of the electrode (e.g., galvanic baths, samples containing sulfides)
- for which potassium chloride c(KCl) = 3 mol/L cannot be used as reference electrolyte (e.g., reaction of potassium or chloride with sample)
The electrode is equipped with a flexible ground-joint diaphragm which is insensitive to contamination and which can be replaced when needed.
When c(KCl) = 3 mol/L is used as bridge electrolyte, storage in storage solution is recommended.
The bridge electrolyte can be easily replaced with a suitable electrolyte (e.g., potassium nitrate c(KNO3) = 1 mol/L (6.2310.010)), storage in the used bridge electrolyte.
The Profitrode is available in other lengths under the following article numbers:
- 6.0255.110: Length 17.8 cm
- 6.0255.120: Length 31.0 cm
Tekniska specifikationer
| Shaft material | Glass |
|---|---|
| Measuring unit | pH |
| Temperature range, long-term (°C) | 0 ... 80 |
| Temperature range, short-term (°C) | 0 ... 100 |
| pH range | 0...14 |
| Max. installation length (mm) | 113 |
| SGJ sleeve | flexible ground-joint sleeve |
| Shaft diameter top (mm) | 12 |
| Shaft diameter bottom (mm) | 12 |
| Min. immersion depth (mm) | 30 |
| Electrode plug-in head | Metrohm plug-in head G |
| Indicator electrode shape | Cylinder |
| Membrane glass type | U |
| Membrane resistance | 150 ... 500 |
| Electrode zero point (mV) | ±15 |
| Electrode slope | >0.97 |
| Diaphragm | Ground-joint diaphragm |
| Reference electrolyte type | c(KCl) = 3 mol/L |
| Electrolyte flow rate (μl/h) at 10 cm water column | 5...100 |
| Reference system | LL system |
| Resistance (kOhm) | 5 |