Chuyển hướng tin nhắn
Conductivity measuring cell c = 0.1 cm-1 with Pt1000 (fixed cable)

Conductivity measuring cell c = 0.1 cm-1 with Pt1000 (fixed cable)

6.0914.040

Conductivity measuring cell c = 0.1 cm-1 with Pt1000 (fixed cable)

vi

6.0914.040

Tổng quan

Conductivity measuring cell made of stainless steel with cell constant c = 0.1 cm-1, with integrated Pt1000 temperature sensor and with fixed cable (4 x B (4 mm)).
This sensor is suitable for measurements of low conductivities (0 uS/cm to 300 uS/cm), e.g., in deion. water or for measurements in accordance with USP 645>.

The sensor can be connected to an 856 Conductivity Module with the adapter box (6.2103.160).

Thông số kỹ thuật

SGJ sleeveFlexible SGJ sleeve
Shaft materialStainless steel 18/13
Measuring range (theoretical)0 ... 300
Measuring unituS/cm
Temperature range, long-term (°C)0...70
Temperature range, short-term (°C)0...70
Max. installation length (mm)125
Shaft diameter top (mm)12
Shaft diameter bottom (mm)12
Min. immersion depth (mm)35
Electrode plug-in headFixed cable 4xB (4 mm)
Temperature sensorPt1000
TypeStainless steel
Cell constant, guide value (1/cm)0.1
This field is required.
    • Sleeve with SGJ 14/12 mm
    • 6.1236.050
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • PCS
    • 1
    • Sleeve with SGJ 14/12 mm
    • 6.1236.050
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • PCS
    • 1
    • Flow-through measuring vessel
    • 6.1420.100
    • For 12 mm dia. electrodes (pH, metal, conductivity)
    • For 12 mm dia. electrodes (pH, metal, conductivity)
    • PCS
    • 1
    • Conductivity standard 100 µS/cm, 5 x 30 mL
    • 6.2324.110
    • Conductivity standard for calibration of conductivity measuring cells with cell constant = 0.1/cm.
    • Conductivity standard for calibration of conductivity measuring cells with cell constant = 0.1/cm.
    • PCS
    • 1
    • Conductivity standard 100 µS/cm 250 mL
    • 6.2324.010
    • Conductivity standard for calibration of conductivity measuring cells with cell constant = 0.1/cm.
    • Conductivity standard for calibration of conductivity measuring cells with cell constant = 0.1/cm.
    • PCS
    • 1