Dual mode bipotentiostat module
MBA.S
Dual mode bipotentiostat module
en
MBA.S
Prezentare generală a caracteristicilor
The MBA.S is an additional bipotentiostat module which can be installed into a MultiBA (MBA) Autolab potentiostat/galvanostat adding an additional working electrode to the MBA instrument. A maximum of 5 MBA.S modules together with one FRA32M.S module can be installed in one MBA instrument.
Specificații tehnice
| Potential range in volt | ± 10 |
|---|---|
| Maximum current in ampère | ± 0.05 |
| Number of current ranges | 7 |
| Number of current ranges remarks | 10 mA to 10nA |
| Potential and current accuracy | ± 0.2 % |
| Current resolution | 0.0003% of current range |
-
- Rotating ring disk electrode, Gold disk Platinum ring
- RRDE.AUPT.S
- The RRDE electrode tip consist of a 5 mm disc of gold with a concentric platinum ring at a distance of 375µm, leading to a theoretical collection efficiency of 24.9%.
- <p>The RRDE electrode tip consist of a 5 mm disc of gold with a concentric platinum ring at a distance of 375µm, leading to a theoretical collection efficiency of 24.9%.</p>
- PCS
- 1
- Rotating ring disk electrode, Glassy Carbon disk Platinum ring
- RRDE.GCPT.S
- The RRDE electrode tip consist of a 5 mm disc of Glassy Carbon with a concentric platinum ring at a distance of 375µm, leading to a theoretical collection efficiency of 24.9%.
- <p>The RRDE electrode tip consist of a 5 mm disc of Glassy Carbon with a concentric platinum ring at a distance of 375µm, leading to a theoretical collection efficiency of 24.9%.</p>
- PCS
- 1