iUnitrode with Pt1000
6.0278.300
iUnitrode with Pt1000
en
6.0278.300
Prezentare generală a caracteristicilor
Intelligent, combined pH electrode with integrated memory chip for storing sensor data and Pt1000 temperature sensor. This electrode is particularly suitable:
- for pH measurements and titrations in difficult, viscous, or alkaline samples
- at elevated temperatures
- for long-term measurements
The fixed ground-joint diaphragm is insensitive to contamination.
Reference electrolyte: c(KCl) = 3 mol/L, storage in storage solution.
Alternatively: reference electrolyte for measurements at T>80°C: Idrolyte, storage in Idrolyte.
iTrodes can be connected to Titrando, Ti-Touch or 913/914 meters.
Specificații tehnice
| Shaft material | Glass |
|---|---|
| Measuring unit | pH |
| Temperature range, long-term (°C) | 0...80 |
| Temperature range, short-term (°C) | 0...100 |
| pH range | 0...14 |
| Max. installation length (mm) | 125 |
| Shaft diameter top (mm) | 12 |
| Shaft diameter bottom (mm) | 12 |
| Min. immersion depth (mm) | 25 |
| Electrode plug-in head | Metrohm plug-in head K |
| Temperature sensor | Pt1000 |
| Indicator electrode shape | Cylinder |
| Membrane glass type | U |
| Membrane resistance | 150...500 |
| Electrode zero point (mV) | ±15 |
| Electrode slope | >0.97 |
| Diaphragm | Fixed ground-joint diaphragm |
| Reference electrolyte type | c(KCl) = 3 mol/L |
| Electrolyte flow rate (μl/h) at 10 cm water column | 5...30 |
| Reference system | LL system |
| Resistance (kOhm) | 5 |
-
- Sleeve with SGJ 14/12 mm
- 6.1236.050
- Sleeve with SGJ 14/12 mm, polyethylene.
- Sleeve with SGJ 14/12 mm, polyethylene.
- PCS
- 1
- Storage vessel
- 6.2008.040
- Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.
- <p>Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.</p>
- PCS
- 1
-
- Sleeve with SGJ 14/12 mm
- 6.1236.050
- Sleeve with SGJ 14/12 mm, polyethylene.
- Sleeve with SGJ 14/12 mm, polyethylene.
- PCS
- 1
- Storage vessel
- 6.2008.040
- Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.
- <p>Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.</p>
- PCS
- 1
-
- 854 iConnect
- 2.854.0010
- The 854 iConnect – Electrode cable and measuring amplifier for intelligent "iTrodes" electrodes (cable length 150 cm).
- <p>The 854 iConnect – Electrode cable and measuring amplifier for intelligent "iTrodes" electrodes (cable length 150 cm).</p>
- PCS
- 1
- Electrolyte 3 mol/L KCl (250 mL)
- 6.2308.020
- Electrolyte solution c(KCl) = 3 mol/L, (for Ag/AgCl reference systems)
- Electrolyte solution c(KCl) = 3 mol/L, (for Ag/AgCl reference systems)
- PCS
- 1
- Idrolyte (250 mL)
- 6.2308.040
- Idrolyte (for 6.0224.100 Electrode and Unitrode at T >80°C)
- Idrolyte (for 6.0224.100 Electrode and Unitrode at T >80°C)
- PCS
- 1
- Storage solution
- 6.2323.000
- Storage solution for all combined pH glass electrodes with reference electrolyte c(KCl) = 3 mol/L
- Storage solution for all combined pH glass electrodes with reference electrolyte c(KCl) = 3 mol/L
- PCS
- 1
- pHit kit
- 6.2325.000
- Maintenance kit for pH electrodes The kit contains:50 mL cleaning solution; 50 mL 3M KCl solution; 50 mL storage solution; 2 Storage vessels; Instructions for use;
- <p>Maintenance kit for pH electrodes </p><p>The kit contains:</p><ul><li>50 mL cleaning solution</li><li>50 mL 3M KCl solution</li><li>50 mL storage solution</li><li>2 Storage vessels</li><li>Instructions for use</li></ul>
- PCS
- 1
- Cleaning solution 3 x 50 mL
- 6.2325.100
- Reliable measuring results over long periods of time can be guaranteed only if the pH glass membrane and the diaphragm receive preventive and regular care. Cleaning by etching with toxic chemicals or applying mechanical treatment to the diaphragm is not only complicated and expensive, it also accelerates the ageing of the pH glass electrode.The cleaning solution was developed for easy and gentle cleaning of pH glass electrodes. Regular use can considerably prolong their service life.This cleaning solution is also part of the pHit kit.
- <p>Reliable measuring results over long periods of time can be guaranteed only if the pH glass membrane and the diaphragm receive preventive and regular care. Cleaning by etching with toxic chemicals or applying mechanical treatment to the diaphragm is not only complicated and expensive, it also accelerates the ageing of the pH glass electrode.<br/>The cleaning solution was developed for easy and gentle cleaning of pH glass electrodes. Regular use can considerably prolong their service life.<br/><br/>This cleaning solution is also part of the pHit kit.</p>
- PCS
- 1