5-ring conductivity measuring cell c = 0.7 cm-1 with Pt1000 (fixed cable)

5-ring conductivity measuring cell c = 0.7 cm-1 with Pt1000 (fixed cable)

6.0915.100

5-ring conductivity measuring cell c = 0.7 cm-1 with Pt1000 (fixed cable)

en

6.0915.100

Visão geral

5-ring conductivity measuring cell with cell constant c = 0.7 cm-1 (guide value), with integrated Pt1000 temperature sensor and with fixed cable (1.2 m) for connecting to an 856 Conductivity Module.
This sensor is suitable for measurements of medium conductivities (5 µS/cm to 20 mS/cm), e.g., in:

  • drinking water
  • surface water
  • wastewater

 

Especificações técnicas

Shaft materialPEEK
Shaft material (completion)PCTFE/PEEK
Measuring range (theoretical)5...20'000 (ideal)
Measuring unitμS/cm
Temperature range, long-term (°C)0...70
Temperature range, short-term (°C)0...70
Max. installation length (mm)125
SGJ sleeveStandard ground-joint 14/12
Shaft diameter top (mm)12
Shaft diameter bottom (mm)12
Min. immersion depth (mm)34
Electrode plug-in headFixed cable, plug N
Temperature sensorPt1000
Type5-ring, Pt
Cell constant, guide value (1/cm)0.7
Este campo é obrigatório.
    • Sleeve with SGJ 14/12 mm
    • 6.1236.050
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • PCS
    • 1
    • Storage vessel 
    • 6.2008.040
    • Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.
    • <p>Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.</p>
    • PCS
    • 1
    • Sleeve with SGJ 14/12 mm
    • 6.1236.050
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • Sleeve with SGJ 14/12 mm, polyethylene.
    • PCS
    • 1
    • Storage vessel 
    • 6.2008.040
    • Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.
    • <p>Together with 6.2008.050 storage vessel holder. it provides a support for the electrode on 807 Dosing Units.</p>
    • PCS
    • 1
    • KCl ion standard 250 mL
    • 6.2301.060
    • KCl ion standard (c(ion) = 0.1000 +/-0.0005 mol/L), can also be used as standard solution for conductometry 12.87 mS/cm (25 °C).
    • <p>KCl ion standard (c(ion) = 0.1000 +/-0.0005 mol/L), can also be used as standard solution for conductometry 12.87 mS/cm (25 &deg;C).</p>
    • PCS
    • 1
    • Conductivity standard 100 µS/cm 250 mL
    • 6.2324.010
    • Conductivity standard for calibration of conductivity measuring cells with cell constant = 0.1/cm.
    • Conductivity standard for calibration of conductivity measuring cells with cell constant = 0.1/cm.
    • PCS
    • 1
    • Conductivity standard 100 µS/cm, 5 x 30 mL
    • 6.2324.110
    • Conductivity standard for calibration of conductivity measuring cells with cell constant = 0.1/cm.
    • Conductivity standard for calibration of conductivity measuring cells with cell constant = 0.1/cm.
    • PCS
    • 1
    • Flow-through cell for the 6.0915.100/6.0920.100
    • 6.2763.000
    • Flow-through cell for the 6.0915.100/6.0920.100 5-ring conductivity measuring cell
    • <p>Flow-through cell for the 6.0915.100/6.0920.100 5-ring conductivity measuring cell</p>
    • PCS
    • 1