The importance of using four-terminal sensing for EIS measurements on low-impedance systems

AN-EC-013

en

In this application note, electrochemical impedance spectroscopy (EIS) is used to test a commercial battery connected in two different ways. In the first EIS measurement, the battery is connected in a two-terminal sensing configuration. In the second EIS measurement, the battery is connected in a four-terminal sensing (Kelvin sensing) configuration. The difference in how the leads are connected results in different measured impedance values for the battery.  

In this application note, electrochemical impedance spectroscopy (EIS) is used to test a commercial battery connected in two different ways. In the first EIS measurement, the battery is connected in a two-terminal sensing configuration. In the second EIS measurement, the battery is connected in a four-terminal sensing (Kelvin sensing) configuration. The difference in how the leads are connected results in different measured impedance values for the battery. 

 