Dual mode bipotentiostat module

Dual mode bipotentiostat module

BA.MAC.204

Dual mode bipotentiostat module

en

BA.MAC.204

기능 개요

The BA is a dual-mode bipotentiostat module that converts the Autolab into a double channel potentiostat with which measurements on 2 working electrodes can be performed sharing the same counter and reference electrode.

In the Bipotentiostat mode, a fixed potential is applied to the second channel (second Working Electrode) while applying a potential step or a sweep to the first channel (first Working Electrode). In the Scanning Bipotentiostat mode, a potential offset with respect to the first channel is applied to the second channel.

기술 스펙

Potential range in volt10
Maximum current in ampère0.050
Number of current ranges7
Number of current ranges remarks10 mA to 10 nA
Potential and current accuracy0.2 %
Current resolution0.0003 % of current range
    • Rotating ring disk electrode, Gold disk Platinum ring
    • RRDE.AUPT.S
    • The RRDE electrode tip consist of a 5 mm disc of gold with a concentric platinum ring at a distance of 375µm, leading to a theoretical collection efficiency of 24.9%.
    • <p>The RRDE electrode tip consist of a 5 mm disc of gold with a concentric platinum ring at a distance of 375&micro;m, leading to a theoretical collection efficiency of 24.9%.</p>
    • PCS
    • 1
    • Rotating ring disk electrode, Glassy Carbon disk Platinum ring
    • RRDE.GCPT.S
    • The RRDE electrode tip consist of a 5 mm disc of Glassy Carbon with a concentric platinum ring at a distance of 375µm, leading to a theoretical collection efficiency of 24.9%.
    • <p>The RRDE electrode tip consist of a 5 mm disc of Glassy Carbon with a concentric platinum ring at a distance of 375&micro;m, leading to a theoretical collection efficiency of 24.9%.</p>
    • PCS
    • 1
    • Rotating Ring Disk Electrode, Platinum disk Platinum ring
    • RRDE.PTPT.S
    • The RRDE electrode tip consist of a 5 mm disc of platinum with a concentric platinum ring at a distance of 375µm, leading to a theoretical collection efficiency of 24.9%.
    • <p>The RRDE electrode tip consist of a 5 mm disc of platinum with a concentric platinum ring at a distance of 375&micro;m, leading to a theoretical collection efficiency of 24.9%.</p>
    • PCS
    • 1