元の言語のページへ戻りました
Ecotrode

Ecotrode

6.0232.100

Ecotrode

ja

6.0232.100

特長

水性酸/塩基滴定のための複合pH 電極。
参照内部液: c(KCl) = 3 mol/L、保存液で保管。

技術仕様

SGJ sleeveフレキシブルな摺り接手
Membrane glass typeT
Resistance (kOhm)3
Shaft materialGlass
Measuring range (theoretical)0 ... 14
Measuring unitpH
Temperature range, long-term (°C)0 ... 80
Temperature range, short-term (°C)0 ... 80
Max. installation length (mm)125
Shaft diameter top (mm)12
Shaft diameter bottom (mm)12
Min. immersion depth (mm)20
Electrode plug-in headMetrohm plug-in head G
Indicator electrode typeGlass electrode
Indicator electrode shapeHemisphere
Membrane resistance150 ... 400
Electrode zero point (mV)0 ± 15 mV (pH = 7)
Electrode slope> 0.97
Asymmetry potential± 15
Internal reference electrode typeLL system
DiaphragmCeramic pin
Reference electrolyte typec(KCl) = 3 mol/L
Bridge electrolyte typec(KCl) = 3 mol/L
Electrolyte flow rate (μl/h) at 10 cm water column10 ... 25
Reference systemLL system
こちらの製品は販売終了していますが、サポートは継続しています。
代替製品
このフィールドは必須です。
    • 保存容器 
    • 6.2008.040
    • ホルダー 6.2008.050 とともに使用すると 807 Dosing Unit の電極用のホルダーになります。
    • <p>ホルダー 6.2008.050 とともに使用すると 807 Dosing Unit の電極用のホルダーになります。</p>
    • PCS
    • 1
    • 保存容器 
    • 6.2008.040
    • ホルダー 6.2008.050 とともに使用すると 807 Dosing Unit の電極用のホルダーになります。
    • <p>ホルダー 6.2008.050 とともに使用すると 807 Dosing Unit の電極用のホルダーになります。</p>
    • PCS
    • 1
    • Storage solution
    • 6.2323.000
    • Storage solution for all combined pH glass electrodes with reference electrolyte c(KCl) = 3 mol/L
    • PCS
    • 1
    • pHit kit
    • 6.2325.000
    • Maintenance kit for pH electrodes The kit contains:50 mL cleaning solution; 50 mL 3M KCl solution; 50 mL storage solution; 2 Storage vessels; Instructions for use;
    • PCS
    • 1
    • 洗浄液 3 x 50 mL
    • 6.2325.100
    • Reliable measuring results over long periods of time can be guaranteed only if the pH glass membrane and the diaphragm receive preventive and regular care. Cleaning by etching with toxic chemicals or applying mechanical treatment to the diaphragm is not only complicated and expensive, it also accelerates the ageing of the pH glass electrode.The cleaning solution was developed for easy and gentle cleaning of pH glass electrodes. Regular use can considerably prolong their service life.This cleaning solution is also part of the pHit kit.
    • PCS
    • 1
    • Electrode cable / 1 m / F
    • 6.2104.020
    • For connecting electrodes with Metrohm plug-in head G to Metrohm instruments (socket F).
    • PCS
    • 1