AN-RA-011
2025-07
Operando Raman characterization of oxygen evolution reaction (OER) catalysts
Gain insight into the electrocatalytic behavior of Ni-based electrocatalysts through Raman spectroelectrochemistry
Summary
Operando Raman characterization provides direct insight into the structural evolution of catalysts for the oxygen evolution reaction. Specifically, nickel‑based materials undergo potential‑dependent transformations that define their catalytic activity. Tracking these changes through Raman spectroelectrochemistry makes it possible to identify the formation of active phases and correlate them with electrochemical performance. By linking vibrational signatures with catalytic behavior, operando Raman studies support the development of efficient Ni‑based electrocatalysts.
Raman characterization of Ni-based electrocatalysts
Analytical grade potassium hydroxide (sourced from VWR) was used as received. An aqueous solution was prepared using ultrapure water (Direct-Q® 5 system, Millipore).
Initially, the sample was pretreated by applying 15 conditioning cycles to improve the contact between the TCP surface and the electrolyte. The potential was scanned from 0.00 V to +0.70 V at a scan rate of 0.01 V/s in 1 mol/L KOH. Raman spectroelectrochemical measurements were performed afterwards to obtain deeper insight into the catalytic behavior of the material. During these experiments, amperometric pulses at different potentials were applied while Raman spectra were concurrently recorded. Multipulsed Amperometric Detection (MAD) protocol was configured to apply 600 s steps from 0.00 V to +0.60 V in increments of +0.10 V (Figure 2a). Raman spectra were simultaneously obtained using an integration time of 30 s (Figure 2b). As the electrochemical pulses and spectral acquisition are perfectly synchronized, the spectra corresponding to each applied potential can be clearly distinguished.
As shown in Figure 2b, two new Raman bands are observed at 475 and 550 cm⁻¹ during the experiment. These bands emerge from +0.50 V onward and are characteristic of NiOOH [1,2]. Specifically, they correspond to the Eg bending vibration and the A1g stretching vibration modes of the Ni–O bonds in NiOOH. The formation of this active oxyhydroxide phase is a key step in nickel‑based electrocatalysts since it is directly involved in promoting the oxygen evolution reaction (OER).
Conclusion
In-situ Raman spectroelectrochemistry is an excellent technique to detect the real-time species evolution of Ni-based catalysts in the OER. For instance, the electrochemical conversion of Ni(OH)2 into NiOOH is demonstrated by the detection of the characteristic bands at 475 and 550 cm-1. Monitoring these vibrational signatures provides detailed information about the formation and stability of the active oxyhydroxide phase, which is essential for understanding how nickel-based catalysts work.
Beyond clarifying the mechanistic steps involved in oxygen evolution, this methodology also offers valuable information for the design of new electrocatalysts. The correlation of potential‑dependent structural changes with catalytic performance can be extended to other important oxidation processes, including the urea oxidation reaction, ammonia oxidation, and related electrochemical transformations where nickel-based catalysts may also play a central role.
Authors
This Application Note was developed in collaboration with Composite Materials Group (María González-Ingelmo, Zoraida González Arias, and Victoria García Rocha) of Instituto de Ciencia y Tecnología del Carbono INCAR-CSIC (Oviedo, Spain).
References
- Yang, X.; Zhang, H.; Yu, B.; et al. An Unveiled Electrocatalysis Essence of NiCo Hydroxides through in Situ Raman Spectroscopy for Urea Oxidation. Energy Tech 2022, 10 (5), 2101010. DOI:10.1002/ente.202101010
- Yan, Z.; Sun, H.; Chen, X.; et al. Anion Insertion Enhanced Electrodeposition of Robust Metal Hydroxide/Oxide Electrodes for Oxygen Evolution. Nat Commun 2018, 9 (1), 2373. DOI:10.1038/s41467-018-04788-3
Related Application Notes
AN-RA-006 New strategies for obtaining the SERS effect in organic solvents
AN-RA-007 Enhancement of Raman intensity for the detection of fentanyl
AN-RA-008 Easy detection of enzymes with the electrochemical-SERS effect
AN-RA-009 Comparison of SPELEC RAMAN and standard Raman microscopes
AN-RA-010 SERS detection of pesticides using screen-printed electrodes