Metrosep C 3 - 150/4.0

Metrosep C 3 - 150/4.0

6.1010.420

Metrosep C 3 - 150/4.0

en

6.1010.420

Feature overview

The innovative substrate on a polyvinyl alcohol base increases selectivity for monovalent and divalent cations significantly. The peak forms on this cation column are highly symmetrical.

The middle separation column of the Metrosep-C-3 product range is particularly suitable for rapid separations of standard cations and certain transition metal cations as well as for separating mid-sized organic amines.

Tech specs

ApplicationRange

Applications

  • Li +, Na+ , K+ , Rb+ , Cs+ , Mg2+ , Ca2+ , Sr2+ , Ba2+ , NH4+ , Mn2+, Co2+, Ni2+ , Zn2+
  • Organic amines
  • Low limits of detection
  • Matrix with high pH

 

Organic Modifier

0 - 50% acetonitrile,
0 - 30% acetone, no methanol

RegenerationAdd 30% acetonitrile to the standard eluent
StandardEluent

Nitric acid eluent (standard eluent)

Nitric acid (c = 1 mol/L): 5.0 mmol/L; 10 mL/2 L
(column temperature 40 °C)

Nitric acid eluent modified

Nitric acid (c = 1 mol/L): 2.5 mmol/L; 5 mL/2 L
(column temperature 40 °C)

Storage

For 1 - 3 days in the eluent; in ultrapure water for longer storage. Recommended temperature: 4 - 8 °C.

Capacity

18 µmol (K+)

Housing materialPEEK
Carrier materialPolyvinyl alcohol with carboxyl groups
Particle size5 µm
Standard flow1.0 mL/min
Maximum pressure15 MPa
Maximum flow1.5 mL/min
pH range2–12
Column dimensions150 x 4.0 mm
Temperature range20–40 °C

Chromatogram

  • Chromatogram
  • Chromatogram
  • Chromatogram
    • Metrosep C 3 Guard/4.0
    • 6.1010.450
    • The Metrosep C 3 S-Guard/4.0 contains the C-3 column material and is used to protect Metrosep C-3 cation columns. Particles and contamination are reliably retained, considerably prolonging the service life of the analytical separation column. The Metrosep C 3 Guard/4.0 also works based on the "On Column Guard System" and is screwed directly onto the separation column with virtually no dead volume.
    • <p>The Metrosep C 3 S-Guard/4.0 contains the C-3 column material and is used to protect Metrosep C-3 cation columns. Particles and contamination are reliably retained, considerably prolonging the service life of the analytical separation column. The Metrosep C 3 Guard/4.0 also works based on the &#34;On Column Guard System&#34; and is screwed directly onto the separation column with virtually no dead volume.</p>
    • PCS
    • 1
    • Metrosep C 3 S-Guard/4.0
    • 6.1010.460
    • The Metrosep C 3 S-Guard/4.0 contains the C-3 column material and is used to protect Metrosep C-3 cation columns. Particles and contamination are reliably retained, considerably prolonging the service life of the analytical separation column.
    • <p>The Metrosep C 3 S-Guard/4.0 contains the C-3 column material and is used to protect Metrosep C-3 cation columns. Particles and contamination are reliably retained, considerably prolonging the service life of the analytical separation column.</p>
    • PCS
    • 1