Cationic components in developer solution applying a flow gradient

AN-C-162

en

This Application Note describes the determination of N,N-diethylhydroxylamine (DEHA), triisopropanolamine (TIPA) and a cationic color developing component (CDC) in a developer solution. The analysis is performed on a high-capacity column such as Metrosep C - 250/4.0 with subsequent direct conductivity detection. To reduce the residence time of the strongly retained color developer components, the column flow rate is increased after the elution of the amines.

This Application Note describes the determination of N,N-diethylhydroxylamine (DEHA), triisopropanolamine (TIPA) and a cationic color developing component (CDC) in a developer solution. The analysis is performed on a high-capacity column such as Metrosep C - 250/4.0 with subsequent direct conductivity detection. To reduce the residence time of the strongly retained color developer components, the column flow rate is increased after the elution of the amines.