Applications
- AN-PAN-1001Online analysis of hydrogen sulfide and ammonia in sour water stripper
This Process Application Note details the simultaneous online analysis of H2S and NH3 in sour water which was previously treated in the sour water stripper (SWS). The method includes automatic cleaning and calibration. Fast and accurate results are continuously supplied for process control.
- AN-PAN-1016Online analysis of silica in boiler feed water of power plants
Excessive silica concentrations in the boiler feed water can lead to deposits on turbine blades and must therefore be avoided. Silica analysis is carried out via differential photometry using a leading-edge technology thermostatic cuvette module for non-sample contact at the detector. Typical concentration ranges for silica are 0–50 ppb and 0–1 ppm or higher.
- AN-S-353Routine drinking water analysis
Ion chromatography (IC) is the method of choice to determine the concentration of common ions in water. This information is crucial as drinking water must meet certain standards to guarantee health (e.g., nitrite and nitrate), as well as technical suitability (e.g., corrosiveness of chloride and sulfate). The Eco IC is an ion chromatograph suitable for economical routine water analysis. Using an A Supp 17 anion column, the analysis of major anions in drinking waters is robust and can be performed at ambient temperatures without additional temperature conditioning.
- 8.000.6021Water analysis
A complete tap water analysis includes the determination of the pH value, the alkalinity and the total water hardness. Both the pH measurement and the pH titration by means of a standard pH electrode suffer from several drawbacks. First, the response time of several minutes is too long and, above all, the stirring rate significantly influences the measured pH value. Unlike these standard pH electrodes, the Aquatrode Plus with its special glass membrane guarantees rapid, correct and very precise pH measurements and pH titrations in solutions that have a low ionic strength or are weakly buffered. Total water hardness is ideally determined by a calcium ion-selective electrode (Ca ISE). In a complexometric titration, calcium and magnesium can be simultaneously determined up to a calcium/magnesium ratio of 10:1. Detection limits for both ions are in the range of 0.01 mmol/L.
- AB-221Standard methods in water analysis
This Bulletin gives a survey of standard methods from the field of water analysis. You will also find the analytical instruments required for the respective determinations and references to the corresponding Metrohm Application Bulletins and Application Notes. The following parameters are dealt with: electrical conductivity, pH value, fluoride, ammonium and Kjeldahl nitrogen, anions and cations by means of ion chromatography, heavy metals by means of voltammetry, chemical oxygen demand (COD), water hardness, free chlorine as well as a few other water constituents.
- AN-Q-006Online analysis of trace anions in borated water of a pressurized water reactor (PWR)
Water of the primary cycle of pressurized water reactors (PWR) contains boron for neutron absorption. The high borate content interferes with the direct analysis of trace anions. Inline Neutralization combined with variable preconcentration and Inline Matrix Elimination (MiPCT-ME) allows to remove boron as boric acid before injection.
- AN-PAN-1038Power generation: Analysis of the m value (Alkalinity) in cooling water
One way to maximize heat transfer efficiency and reduce costs in a power plant is by controlling the water chemistry in the cooling circuit. This cooling water is kept alkaline to maintain the protective oxide layer on the metal piping throughout the water circuit. However alkalinity above the recommended range increases the probability of scale formation (deposition), so it is buffered with carbonate (CO32-) and bicarbonate ions (HCO3-). Titration of the cooling water to pH 4.5 gives the so-called "M-Alkalinity" (methyl orange alkalinity), a measure of total alkalinity. Below this pH, there is no more alkalinity present, only free acid (H+), carbonic acid (H2CO3), and CO2.
- AN-S-301Comprehensive water analysis with VoltIC pro I
VoltIC pro I is the perfect combination of voltammetry and ion chromatography for the fully automated analysis of anions, cations, and heavy metals (e.g., Zn, Cd, Pb, Cu): comprehensive water analysis on a single system.
- AB-178Fully automated analysis of water samples
The determination of the physical and chemical parameters as electrical conductivity, pH value, p and m value (alkalinity), chloride content, the calcium and magnesium hardness, the total hardness, as well as fluoride content are necessary for evaluating the water quality. This bulletin describes how to determine the above mentioned parameters in a single analytical run.Further important parameters in water analysis are the permanganate index (PMI) and the chemical oxygen deman (COD). Therefore, this Bulletin additionally describes the fully automated determination of the PMI according to EN ISO 8467 as well as the determination of the COD according to DIN 38409-44.
- AN-S-387Comprehensive water analysis with TitrIC flex II
The TitrIC flex II system is the perfect combination of titration, direct measurement, and ion chromatography for fully automated analysis of all key parameters. These include pH, conductivity, hardness, anions, cations, as well as the calculation of the ion balance: comprehensive water analysis from one system.
- AN-S-320Fast IC: Drinking water analysis in three minutes
Fast IC means a high sample throughput. This is attained with short columns, relatively high flows and strong eluents. Applied to drinking water analysis this means: determining chloride, nitrate and sulfate within 3 minutes.
- AN-PAN-1044Online trace analysis of amines in the alkaline water-steam circuit of power plants
Neutralizing amines are added to adjust pH levels within the water-steam circuit of power plants to avoid corrosion-inducing conditions. This preventive maintenance can reduce costly and critical downtimes due to corrosion, however frequent monitoring of the amine chemistry is necessary to ensure conditions stay optimal. The 2060 IC Process Analyzer featuring the Metrohm intelligent Partial Loop Technique (MiPT) option is ideal for this application, with the ability to measure trace amounts of the analytes precisely and reliably through an automated method. The benefit of using IC is that multiple analytes can be monitored simultaneously, and here the ability to measure the presence of sodium next to the high ammonium or amine concentrations could indicate that cooling water is seeping into the circuit, indicating a problem upstream.
- AN-PAN-1013Online analysis of boric acid in the cooling water of pressurized water reactors
Boric acid requires precise monitoring in the primary circuit to control nuclear reactor reactivity. The 2060 TI Process Analyzer monitors boric acid online in near-real time.
- TA-044pH value, conductivity and titration in water and soil analysis
The rapid growth of the Earth's population has led to massive increases in the consumption of energy and resources and in the production of consumer products and chemicals. It is estimated that 17 million chemical compounds are currently on the market, of which 100,000 are produced on a large industrial scale. Many of these enter the environment. This leads to a demand for sensitive analytical procedures and high-performance analytical instruments.pH value, conductivity and oxygen requirement are important characteristics in water and soil analysis. The first two of these can be determined rapidly; for the third, the titration that is used is also the one used in numerous single determinations. This article describes several important standard-compliant determinations in water and soil analysis.
- AN-K-017Water in expandable polystyrene – Oven system with closed sample vials simplifies analysis
The presence of water in expandable polystyrene (EPS) can have a negative impact on the thermal insulation properties, as it increases thermal conductivity. If EPS is exposed to a high moisture environment, additional water may be absorbed, which can further affect thermal insulation.Direct analysis of the moisture content by Karl Fischer titration requires the water to be extracted from the EPS, which involves several time-consuming steps. Therefore, determination of the water content with an oven system is preferred. As EPS expands when heated, the use of sample boats, as required by ASTM D6869, is not possible, as the EPS will contaminate the oven system. This Application Note describes the determination of water content in EPS using an oven system with closed sample vials. A determination takes about 7 to 14 min depending on the water content of the sample and the sample size.
- AN-Q-005Online analysis of trace anions in power plant water matrices
A setup that allows online sampling is crucial for immediate and contamination-free analysis of power plant water samples. This application recommends a setup that facilitates simultaneous anion/cation determinations. Automated inline sample preparation combines variable preconcentration (MiPCT) and calibration with a single multi-ion standard. AN-Q-004 displays the respective cation results.
- AN-Q-004Online analysis of trace cations in power plant water matrices
A setup that allows online sampling is crucial for immediate and contamination-free analysis of power plant water samples. This application recommends a setup that facilitates simultaneous anion/cation determinations. Automated inline sample preparation combines variable preconcentration (MiPCT) and calibration with a single multi-ion standard. AN-Q-005 shows the respective anion results.
- AN-C-147Fast analysis of cations in tap water using Metrosep C 4 - 100/2.0
Reducing the analysis time is a demanding task because it is accompanied by a parallel reduction of peak resolution. With a Microbore column 100 mm in length, standard cations in tap water can be determined in only 5 minutes. Strontium can also be determined by simply extending the run time to 6.5 min.
- 410000053-AQuantitative Analysis of a Water-soluble Polymer Using the i-Raman EX Spectrometer
Quantitation of the functionalization of a water-soluble polymer was achieved using a portable Raman spectrometer. The Raman spectrum provides strong, unique bands for both the initial and fully reacted polymer. This enables development of a simple, robust quantitative analysis of the percent polymer functionalization. This method is now routinely used in a manufacturing plant's quality control laboratory.
- AN-CIC-034Fast analysis of AOX in waters by CIC
Combustion ion chromatography (CIC) measures AOX (adsorbable organically bound halogens, i.e., AOCl, AOBr, AOI) and AOF as well as CIC AOX(Cl) according to DIN 38409-59 and ISO 18127.
- AN-U-049Analysis of bromate in drinking water according to ISO 11206 and EPA 317
Ion chromatography with PCR and UV/VIS detection provides a highly specific and sensitive method for bromate analysis, meeting EPA Method 317 and ISO 11206 requirements.
- AN-T-205Fully automated water analysis by OMNIS
The determination of the physical and chemical parameters as electrical conductivity, pH value, alkalinity, the calcium and magnesium hardness as well as the total hardness are necessary for evaluating the water quality. A fast and accurate determination in tap water is realized using an automated OMNIS System working in parallel on different workstations. An 856 Conductivity Module with Dosinos extends the system.
- AN-S-347Comprehensive water analysis with VoltIC Professional 1
VoltIC Professional 1 is the perfect combination of voltammetry and ion chromatography for the fully automated, simultaneous analysis of anions, cations, and heavy metals (e.g., Zn, Cd, Pb, Cu). The multiple-parameter analysis uses the same "Liquid Handling" elements and a shared sample changer, thus saving on space and costs.
- WP-087Green alternative methods for voltammetric analysis in different water matrices
This White Paper presents four different «green» sensors: the scTRACE Gold, screen-printed electrodes, the glassy carbon electrode, and the Bi drop electrode from Metrohm that can be used to determine low concentrations of heavy metals in different sample matrices, such as boiler feed water, drinking water, and sea water.
- WP-021Water analysis in the field: Determining arsenic, mercury, and copper
Heavy metals such as arsenic and mercury find their way into the ground water in many regions of the world, either through natural processes or as the result of human activities. Limit values are exceeded many times over, particularly for arsenic in drinking water, in many areas. This calls for a rigorous monitoring of water quality. The present whitepaper focuses on field determinations of arsenic, mercury, and copper – directly at the sampling site.
- AN-S-323Fast IC: Drinking water analysis including fluoride in less than seven minutes
Fast IC means short run times and a high sample throughput. This is attained using short columns and strong eluents. Drinking water (including fluoride) is analyzed on the Metrosep A Supp 5 - 100/4.0 under the same conditions as in AN-S-322.
- AN-Q-007Online analysis of chloride and sulfate in supercritical water-steam cycles
A combination of the 850 Professional IC and the 872 Extension Module Liquid Handling opens the field of Metrohm’s online monitoring by IC. In this application, Inline Preconcentration is coupled to Matrix Elimination (MiPCT-ME). By removing excess matrix components, corrosive anions can be sensitively determined. Additionally, this technique allows automated calibration using a single multi-ion standard solution. Online trace analysis for chloride and sulfate is possible for several different sample lines.
- AN-C-133Tap water analysis for cations using Metrohm intelligent Partial Loop Technique (MiPT)
Partial loop injection is a well known way of sample introduction to HPLC. In ion chromatography, it is not yet used to a large extent. Liquid handling with Metrohm's Dosino technology now enables to use partial loop injection on a highly reproducible and accurate level. It includes multi-level calibration out of one standard solution. This Application Note shows its use for parallel anion and cation determination in tap water applying one single Sample Processor. The anion results are shown in Application Note S–287.
- AN-S-287Tap water analysis for anions using Metrohm intelligent Partial Loop Technique (MiPT)
Partial loop injection is a well known way of sample introduction to HPLC. In ion chromatography it is not yet used to a large extent. Liquid handling with Metrohm's Dosino technology now enables to use partial loop injection on a highly reproducible and accurate level. It includes multi-level calibration out of one standard solution. This AN shows its use for parallel anion and cation determination in tap water applying one single Sample Processor. The cation results are shown in Applicatin Note C-133.
- AN-M-010Speciation analysis of arsenic and selenium in drinking water using IC-ICP/MS
The maximum contaminant concentrations (Maximal Contaminant Level, MCL) of inorganic arsenic and selenium species in drinking water should not exceed 10 and 50 µg/L, respectively. Given that each of the two elements occurs in two oxidation levels – trivalent and pentavalent – a separation step is necessary prior to ICP/MS detection. This Application Note shows the simultaneous determination of the two arsenic (arsenite and arsenate) and selenium species (selenite and selenate). Separation takes place on the Metrosep Dual 3 - 100/4.0 column.
- AN-PAN-1032Monitoring corrosion in power plants with online process analysis
Corrosion in the water-steam circuit of power plants leads to shorter lifetimes of most metal components and potentially dangerous situations. Flow Accelerated Corrosion (FAC) is a specific case, leading to thinned pipes and elevated Fe concentrations in the circuit. Additionally, metal transport issues such as with Cu from copper heat exchangers can lead to deposition on the high pressure turbine blades, decreasing their efficiency. Current methods can monitor but not prevent these issues, and analysis times are extremely long (up to three weeks). In combination with the power plant’s Distributed Control System (DCS), online monitoring of Fe and Cu with the 2060 Process Analyzer from Metrohm Process Analytics ensures that corrosion can be controlled before it affects the power plant efficiency, ultimately decreasing downtime and lowering maintenance costs. Results are offered within 20 minutes, allowing fast adjustments to the water-steam circuit to protect company assets.
- 8.000.6033Analysis of energetic materials in various water and soil samples using HPLC and LC-MS
In modern days, a new breed of energetic (explosive) materials is emerging. Traditional aromatic nitrates are still in use, but there is dire need of analytical techniques for energetic materials in the chemical class of peroxides, azo etc. This presentation will demonstrate the use of a modern HPLC system with traditional detector (DAD) and also coupled with mass spectrometry for the analysis of abovementioned various classes of energetic materials.
- AN-PAN-1043Online trace analysis of cations in the primary circuit of nuclear power plants
In power plants, corrosion is the primary factor leading to costly and critical downtimes. In a nuclear power plant, a separate water circuit known as the Pressurized Water Reactor (PWR) ensures radioactive material stays contained while still transferring heat and energy to the other circuits. Boric acid and lithium hydroxide are added specially to the PWR circuit in amounts which can complicate other analytical measurements. Lithium prevents corrosion and must be monitored, along with other cations such as zinc, nickel, and ammonium. In order to measure these cations online at sub-µg/L range in a single analysis, the 2060 IC Process Analyzer is offered with combined Inline Preconcentration and Inline Matrix Elimination. Several cations can be analyzed in a single injection, with automated sample preparation making precise and reliable measurements easier.
- 8.000.6013Analysis of produced water contaminants by ion chromatography
The analytical challenge treated by the present work consists in detecting sub-ppm quantities of bromide, sulfate, aliphatic monocarboxylic acids and several alkaline earth metals in the presence of very high concentrations of sodium and chloride. Bromide, sulfate, acetate and butyrate can be reliably determined by suppressed conductivity detection. Due to matrix effects, propionate can only be detected qualitatively. This drawback can be overcome by coupling the ion chromatograph (IC) to a mass spectrometric (MS) detector. This results in reduced matrix interferences and significantly enhanced sensitivities. The cations magnesium, barium and strontium are determined by non-suppressed conductivity detection.
- WP-045When HPLC fails: IC in food, water, and pharmaceutical analysis
High-Performance Liquid Chromatography (HPLC) and Ion Chromatography (IC) are commonly used in the pharma, food, and environmental sectors to analyze samples for specific components and to verify compliance with norms and standards. However, users of HPLC may run into the limitations of this technique, e.g., when analyzing standard anions or certain pharmaceutical impurities. This white paper outlines how such challenges can be overcome with IC.
- AN-PAN-1030Monitoring of chromate in wastewater streams
Chromium is extracted from chromite ore and is an important part in the production of stainless steel. Chromium is mainly divalent, trivalent and hexavalent in its compounds. In contrast to chromium(III), which is an important trace element and one that is only sparingly soluble in water, hexavalent chromium is extremely toxic and very water-soluble. Cr(VI) is furthermore an important raw material for industry. It must be determined rapidly and precisely in the lower µg/L range in wastewater. Metrohm Applikon offers an array of process analyzers for the analysis of wastewater streams which determine chromium precisely and reproducibly using photometry.
- AN-PAN-1040Ammonia in cooling water of thermal power plants
Thermal power plants require enormous amounts of water, using high purity steam at high pressure to rotate turbines. A separate cooling water circuit is implemented, which helps to form a vacuum when the steam condenses after the turbines. Maintaining this vacuum with optimal condensation parameters is critical for the power plant efficiency. The copper condensers are susceptible to corrosion by ammonia, leading to an upper limit of 2 mg/L NH3 set by EPRI in cooling water. Small cracks in the condenser combined with the large pressure differential between the steam circuit and the cooling water circuit will contaminate the high purity water in the boiler, causing major problems and necessitating a shutdown for plant maintenance. Monitoring NH3 online in cooling water with a process analyzer can signal early problems in a plant before significant intermediation is necessary.
- AN-PAN-1009Online analysis of ammonia, nitrate, and nitrite in wastewater
This Process Application Note deals with online measurements of ammonia, nitrite, and nitrate in wastewater treatment plants. These nitrogen compounds are analyzed simultaneously using a drift-free colorimetric measurement in a multi-parameter process analyzer from Metrohm Process Analytics.
- AN-PAN-1045Online monitoring of copper corrosion inhibitors in cooling water
Copper is used widely in industrial cooling water systems for its heat transfer properties, although it is susceptible to corrosion. Corrosion can cause a loss of efficiency and eventually a failure of equipment, leading to costly maintenance, replacement, and downtime. Corrosion inhibitors (triazoles) can be added to the water chemistry, which form sparingly soluble protective layers on the surface of the metal. Triazole concentrations must be maintained to protect the copper, which necessitates regular concentration determinations in cooling water. The 2060 IC Process Analyzer with UV/VIS detection is well-suited for this application, able to precisely and reliably measure multiple ionic and UV-active compounds simultaneously in cooling water.
- AN-PAN-1036Online determination of alkalinity and hardness in process and make up water for beer production
Beer is a popular beverage consumed by millions of people for enjoyment, despite its humble beginnings as a water purification technique in pre-modern times. Brewing beer requires large amounts of water which must adhere to strict alkalinity, hardness, and pH parameters to ensure uniformity in flavor and appearance between each batch. Alkalinity is introduced by carbonates and hydroxides in water which raise and buffer the pH. Hardness, balanced to a large degree by the alkalinity, comes from Ca and Mg ions, mainly present as hydrogen carbonates. Depending on the concentration ranges, the 2035 Process Analyzer or the 2060 Process Analyzer are ideally suited for the fully automatic execution of these important analyses, as well as additional parameters like pH or conductivity. These process analyzers can signal the plant’s distribution system to correct the water chemistry, ensuring consistent product quality. In addition to alkalinity and water hardness, numerous other parameters can also be determined (pH, conductivity, etc.).
- AN-C-096Trace analysis of monovalent cations in an ethanolamine matrix (secondary cycle of a pressurized water reactor) using Metrohm Inline Sample Preparation with an 800 Dosino for liquid handling
Methylamine (MMA), dimethylamine (DMA), and trimethylamine (TMA) in methylpyrrolidone using Metrohm Inline Matrix Elimination.
- AN-M-016Resolving haloacetic acids in water
During drinking water disinfection with chlorine, chloramine, or ozone, potentially toxic halogenated byproducts can be formed. The disinfectants can react with naturally occurring bromide and/or organic matter in the source water and form one of the most common and highly toxic disinfection byproducts (DBPs): haloacetic acids (HAAs). To protect human health, maximum tolerable levels of HAA in drinking waters are regulated (EPA 816-F-09-004). The EPA Method 557 specifies the analysis of HAAs beside bromate and dalapon by ion chromatography coupled to tandem mass spectroscopy (IC-MS/MS) with LODs varying from 0.02–0.11 µg/L. However, even with single MS, a high sensitivity is achieved to determine the current MCLs within an adequate accuracy. This Application Note describes the analysis of bromate, chlorite, monochloroacetic acid (MCAA), monobromoacetic acid (MBAA), bromochloroacetic acid (BCAA), bromodichloroacetic acid (BDCAA), dibromoacetic acid (DBAA), dichloroacetic acid (DCAA), tribromoacetic acid (TBAA), chlorodibromoacetic acid (CDBAA), and trichloroacetic acid (TCAA) with IC/MS. The Metrohm Driver 2.1 for EmpowerTM offers the analysis as a single software solution with EmpowerTM.
- AN-T-084Total, calcium, and magnesium hardness in water samples
ASTM D8192 describes the photometric titration of the total hardness, calcium hardness, and magnesium hardness in water with an optical sensor for objective endpoint indication, increasing precision and reliability. The method is suitable for both colored and colorless samples such as groundwater, surface water, wastewater, and drinking water. Using a fully automated OMNIS system equipped with an Optrode ensures that the sample preparation and analysis are repeatable.
- AN-PAN-1042Online trace analysis of anions in the primary circuit of nuclear power plants
Measures to monitor or prevent corrosion are crucial in nuclear power plants, where significant risks to health and safety can occur if corrosion is left unchecked. Anions corrode metals under high temperature and pressure, therefore their concentrations must be monitored at all times. The analytical challenge in the primary circuit is detection of anions in the μg/L range alongside gram quantities of boric acid and lithium hydroxide. Precise, reliable trace analysis requires the method to be automated as much as possible. The 2060 IC Process Analyzer from Metrohm Process Analytics can measure several anions from a single injection, with combined Inline Preconcentration and Inline Matrix Elimination to measure low anion concentrations precisely and reliably time after time.
- AN-PAN-1039Ortho- and total phosphate phosphorus analysis online according to EN ISO 6878
Phosphorus removal is essential in waste water treatment plants to ensure the environmental balance is not upset by discharged effluent. In the treatment facility it is important to know the bioavailable o-phosphate phosphorus (o-PO4-P) concentration in the influent stream either to feed bacteria or to calculate the amount of reagents needed for chemical treatment. For environmental compliance monitoring purposes, treated effluent is monitored for TP – the sum of all insoluble and dissolved phosphates present. With the Metrohm Process Analytics 2035 TP Analyzer (complete with integrated compact digestion cuvette photometer module), you can keep track of both o-PO4-P and TP according to DIN EN ISO 6878:2004-09 around the clock.
- AN-PAN-1049Online determination of bromate and other disinfection byproducts in drinking & bottled water with IC
Drinking water which has been disinfected via the ozonation process can contain undesirable levels of bromate, a carcinogen, via oxidation of bromide in the raw water. Already several agencies including the World Health Organization have recommended concentration limits for bromate set in place to limit its risks to our health. Ion chromatography is mentioned in several analytical standards for the determination of disinfection byproducts (DBP) including bromate, such as EPA 300.1, 317.0, 321.8, 326.0, ASTM D6581, ISO 11206, and ISO 15061. Monitoring trace levels of bromate online means higher throughput and less time spent performing manual laboratory tests, and ensures quality drinking water is produced.
- AN-PAN-1003Online analysis of amines concentration in carbon capture plants
Carbon capture systems strip carbon dioxide from flue gases. Online analysis of amines and carbon dioxide enhances amine usage efficiency and reduces operational costs.
- WP-081Fast determination of AOX in waters according to DIN 38409-59
Monitor adsorbable organic halogens (AOX) in water using combustion ion chromatography (CIC) for precise analysis of AOCl, AOBr, AOI, and total AOX.
- AN-T-032Sulfide and hydrogen sulfide in water
This Application Note presents a potentiometric titration method for trace H2S analysis in water on an OMNIS system using silver nitrate and an Ag Titrode.
- AN-PAN-1066Online analysis of cadmium in incineration plant wet gas scrubbers
Incineration flue gas requires treatment such as wet scrubbing. The 2060 VA Process Analyzer monitors heavy metals in the scrubbing water, ensuring compliance.
- AN-S-236Drinking water quality by EPA 300.1
Fast and reliable analysis of drinking water by combining EPA method 300.1 Parts A and B in a single IC run.
- AN-S-404Water quality testing with EPA 300.1
The Metrosep A Supp 21 column and 948 Continuous IC Module, CEP enable efficient, automated single-run analysis of major anions and disinfection byproducts in water.
- AN-M-017IC-MS/MS analysis of trifluoroacetic acid according to DIN 38407-53
The new DIN draft standard 38407-53 outlines TFA analysis in water using direct injection LC-MS/MS, enabling quantification from 0.1–3.0 μg/L as shown in this Application Note.
- AN-V-229Antimony(III) in drinking water
The toxicity of antimony depends on its oxidation state: antimony(III) is more toxic than antimony(V). Due to its carcinogenicity, EU legislation specifies 5 µg/L and the World Health Organization (WHO) sets a maximum concentration of 20 µg/L as the Sb(III) limit value in drinking water.Straightforward determination using anodic stripping voltammetry provides a fast (analysis time under 10 minutes) and an ultra-sensitive tool for monitoring the antimony(III) concentration in drinking water. Measurements can be performed in the laboratory with the 884 Professional VA, or alternatively in the field with the 946 Portable VA Analyzer.
- AN-I-009Cyanide in water
Cyanides are used in some industrial processes, but if not handled carefully, they could contaminate the wastewater. In an acidic or neutral environment, this contaminated wastewater can form highly toxic hydrogen cyanide gas. Furthermore, the cyanide salts could also poison the environment and enter the ground water system. Therefore, it is essential to monitor the content of cyanide in effluent water. Cyanides can be easily determined with a cyanide ion-selective electrode. This application note presents a method for cyanide analysis according to APHA Method 4500-CN and ASTM D2036.
- 8.000.6062Water determination in various plastics
The presence of excessive water in plastics adversely affects the performance of polymeric goods which is why water determination is of crucial importance. This article describes the accurate and straightforward determination of the water content using the Karl Fischer Oven Method in ten different plastic types that are not amenable to direct Karl Fischer titration. The experiments revealed that besides the determination of the oven temperature, sample preparation is one of the most important steps of the analysis, especially in case of hygroscopic plastic samples.
- WP-010Glyphosate and AMPA in drinking water
For the first time, glyphosate determination and that of its primary metabolite AMPA in drinking water using IC with pulsed amperometric detection (flexIPAD) in the low µg/L range are shown. Compared to HPLC analysis with a mass-selective detector, it is a very cost-effective method for determining the glyphosate and AMPA content in water and foodstuffs. With a detection limit at approx. 1 µg/L, compliance with limit values for glyphosate can be monitored in the USA, Canada, and Australia, among others.
- AN-NIR-077Moisture analysis in caprolactam
Caprolactam is an important polymer used for the production of Nylon 6, which is the base material for industrial fibers. Due to its commercial significance, many different synthesis methods have been developed over the years. Caprolactam is hygroscopic and water soluble, therefore it is important to have a reliable analysis technique for water determination. Analyzing the water content by conventional methods requires each sample to be weighed, dissolved, heated, and titrated. Compared to the primary method, near-infrared spectroscopy (NIRS) offers unique advantages: it generates reliable results within seconds, but it does not need any sample preparation nor does it create chemical waste.
- AN-PAN-1050Inline moisture analysis in fluid bed dryers by near-infrared spectroscopy
In the pharmaceutical industry, the fluid bed granulator/dryer is an integral point in the manufacture of powdered materials. Residual moisture must be kept within certain specifications to avoid fracturing of particles or caking (stickiness) of the bulk material. Current methods are slow and cumbersome, which can lead to damaged or degraded product. The ability to monitor the residual moisture content inline after drying is possible with near-infrared spectroscopy (NIRS). The 2060 The NIR Analyzer offers fast, reagent-free, nondestructive analysis of residual moisture of powders with a fluid bed probe specifically designed for these applications.
- AN-K-070Water in petroleum products
Moisture in petroleum products causes several issues: corrosion and wear in pipelines and storage tanks, an increase in debris load resulting in diminished lubrication, blocked filters, or even harmful bacterial growth. As a result, increased water content can lead to infrastructure damage, higher maintenance costs, or even unwanted downtimes.Coulometric Karl Fischer titration is the method of choice for low water content in petroleum products. Using a Karl Fischer oven to vaporize the water present in the sample prior to titration not only greatly reduces matrix interferences, it can also be fully automated. This allows a reliable and cost-efficient analysis of the water content according to ASTM D6304 (Procedure B) in products such as diesel, hydraulic oil, lubricant, additive, turbine oil, and base oil.
- AN-T-073Fully automated determination of conductivity, pH value, and alkalinity in tap water including sample preparation
The automated system Basic water analysis determines conductivity, pH value, and alkalinity in all kind of water samples. The high degree of automation (e.g., automated sample addition, automated calibration as well as automated titer and cell constant determination) minimizes errors and guarantees an outstanding reproducibility.
- AN-S-342Trace perchlorate in drinking water according to ISO 19340 applying Annex B
Perchlorate in water is mainly due to anthropogenic sources such as fertilizers, fireworks, rocket fuel, etc. Trace analysis of perchlorate in water samples is a critical task. The high content of standard anions leads to large peaks that interfere with the very small perchlorate peak. In the heart-cut technique, the perchlorate fraction – widely freed of interfering anions – is re-injected onto the column thus providing a sharp peak.
- AB-123Determination of manganese in water samples by anodic stripping voltammetry
"A sensitive methods to determine manganese is described. It is primarily suitable for the investigation of ground, drinking and surface waters, in which the concentration of manganese is important. The method can naturally also be used for trace analysis in other matrices.Manganese is determined in an alkaline borate buffer by the anodic stripping voltammetry (ASV). Interference by intermetallic compounds is prevented by the addition of zinc ions in the sample. The limit of determination lies at b(Mn) = 2 µg/L."
- AN-K-020Water in flour (meal)
The water content of flour is determined according to Karl Fischer. To shorten the analysis times and to obtain more precise results, the determinations are carried out at 50 °C.
- AN-K-049Water in plastic pellets – Interference-free determination based on ASTM D6869
The water content, also called moisture content, of plastics is an important quality parameter, as it affects the properties and processability of some plastics. A high water content can lead to degradation of the plastic by hydrolysis or cause surface imperfections. Additionally, it can affect the physical properties of some plastics.For this analysis, the oven technique is used, as volatile compounds present in plastics will interfere, if the water content is directly determined by coulometric Karl Fischer titration. The water content determination in polycarbonate pellets, performed with the 885 Compact Oven Sample Changer and 899 Coulometer, is described in this Application Note.
- AN-NIR-104Analysis of green coffee beans by near-infrared spectroscopy
Near-infrared spectroscopy (NIRS) is a fast, chemical-free analytical method for the simultaneous analysis of density, water activity, and moisture of green coffee beans.
- AN-S-357Waste water from a waste water cleaning plant: Anion determination with a Metrosep A Supp 4 - 250/2.0
The microbore Metrosep A Supp 4 - 250/2.0 column is particularly suitable for the analysis of anions in critical samples. A waste water sample is being analyzed in the current application. The sample requires only one filtration prior to injection on the Metrosep A Supp 4 - 250/2.0. The anions are quantified with the application of conductivity detection following sequential suppression.
- WP-075Simple determination of haloacetic acids (HAAs) in potable water with ion chromatography hyphenated to mass spectrometry
Haloacetic acids (HAAs) are commonly produced as disinfection byproducts (DBPs) from water treatment processes. Some HAAs are regulated by the authorities and have been classified as potentially carcinogenic. They have traditionally been analyzed by gas chromatography (GC), a technique that requires time-consuming sample extraction and derivatization, leading to higher costs per analysis. Ion chromatography hyphenated to mass spectrometry (e.g., single or triple quadrupole MS systems) is a powerful tool that can handle many challenging analytical tasks such as measuring μg/L levels of HAAs in potable water samples. This White Paper explains the benefits of using this hyphenated technique for the accurate measurement of HAAs in potable water.
- AN-V-221Cadmium and lead in drinking water with a Bi drop electrode
To reduce the toxic effects of cadmium on the human body, as well as to limit the neurotoxic effects of lead, the provisional guideline values in the World Health Organization’s «Guidelines for Drinking-water Quality» are set to a maximum concentration of 3 µg/L for cadmium and 10 µg/L for lead. The completely mercury-free Bi drop electrode takes the next step towards converting voltammetric analysis into a non-toxic approach for heavy metal detection. Using this environmentally friendly sensor for anodic stripping voltammetry (ASV) allows the simultaneous determination of Cd and Pb in drinking water. The outstanding sensitivity is more than sufficient to monitor the provisional WHO guideline values.
- AN-K-068Water content determination and acid-base titration in parallel
The water content determination by volumetric Karl Fischer titration is one of the most important analyses worldwide. Using an OMNIS system consisting of an OMNIS Titrator and an OMNIS Sample Robot, the fully automatic analysis of water content is possible in various products and matrices. The OMNIS Sample Robot is capable of running several different titrations in parallel. In this Application Note, we present the results of a volumetric Karl Fischer titration run in parallel to an aqueous acid-base titration on the same system. The water content is not influenced by the parallel running aqueous titration, allowing the combination of potentiometric titrations and Karl Fischer titrations on the same automated system.
- AN-CS-013Cations in deionized water and calculation of LOD and MDL of MiPCT
Trace cation analysis in high purity water (sub-μg/L range) requires cation chromatography after sequential suppression and intelligent Preconcentration Technique (MiPCT). Trace cations in deionized water (DI) are determined and the method detection limit (MDL according to US EPA) as well as the limit of detection (LOD = 3 x S/N) is calculated. MDL and LOD are very similar in the lowest ng/L range for this setup with 6 mL preconcentration volume.
- AN-C-174Rapid determination of cations in drinking water on a microbore separation column
Cation analysis in drinking water is a routine task in ion chromatography and can be achieved with a variety of separating columns. The use of a microbore Metrosep C 6 - 250/2.0 column with a high eluent concentration makes it possible to reduce analysis time to less than 12 minutes. Very symmetrical peaks with high sensitivity for the divalent cations are also achieved. Direct conductivity detection is applied.
- AN-M-007Determination of urea in ultrapure water using IC-MS
Urea is not a typical analyte for ion chromatography. In combination with MS, however, IC is the method of choice for the trace analysis of urea in ultrapure water. This Application Note shows the determination of urea concentrations in the ppb range using the Metrosep C 6 - 250/4.0 column.
- WP-056Determining dissolved oxygen in water – Titration or direct measurement?
«Dissolved oxygen» describes the amount of oxygen molecules (O2) which are dissolved in a liquid phase under certain conditions. In this white paper, two different methods for the analysis of dissolved oxygen, titration and direct measurement, are compared and contrasted to help analysts determine which method is more suitable for their specific applications. Here, we primarily focus on the determination of dissolved O2 in water. However, the same principle applies for other liquid phases such as non-alcoholic or alcoholic beverages.
- AN-C-135Cations in drinking water using Metrosep C 4 - 150/4.0 column according to ISO 14911.
Drinking water analysis is strongly regulated by standards. In this Application Note, the cation determination according to ISO 14911 is shown. The Metrosep C 4 - 150/4.0 is the optimum separation column for this purpose.
- AN-S-295Anions in boiler water including sulfur speciation (sulfite and sulfate)
Boiler water analysis is an important task in power plant applications. Under the given conditions, the "Metrosep A Supp 10 - 100/4.0" column separates sulfite and sulfate without any organic modifier in the eluent. Even without any stabilizer, sulfite can be determined with a high reproducibility.
- AN-C-143Ammonia in addition to standard cations in maritime pore water
Maritime pore water contains sodium in the percentage range. The analysis of ammonia in this kind of sample requires a high column capacity and an exceptionally good separation of sodium and ammonia. These requirements are completely fulfilled by a 2 µL injection to the high-capacity Metrosep C 6 - 250/4.0 column.
- AB-231Determination of zinc, cadmium, lead, copper, thallium, nickel and cobalt in water samples by anodic and adsorptive stripping voltammetry according to DIN 38406-16
The standard method postulated by DIN 38406 Part 16 describes the determination of Zn, Cd, Pb, Cu, Tl, Ni, and Co in drinking, ground, surface and precipitation (e.g. rain) water. Because the presence of organic substances in the water samples can strongly interfere with the voltammetric determination, a pretreatment with UV digestion using hydrogen peroxide is necessary. This digestion ensures the elimination of all organic substances without introduction of blank values. These methods can, of course, also be applied for trace analysis in other materials, e.g. trace analysis in the production of semiconductor chips based on silicon. Zn, Cd, Pb, Cu, and Tl are determined on the HMDE by means of anodic stripping voltammetry (ASV), Ni and Co by means of adsorptive stripping voltammetry (AdSV).
- WP-048Utilizing online chemical analysis to optimize propylene oxide production
Propylene oxide (PO) is a major industrial product used in assorted industrial applications, mainly for the production of polyols (the building blocks for polyurethane plastics). Several production methods exist, with and without co-products. This white paper lays out opportunities to optimize PO production for safer and more efficient processes, higher quality products, and substantial time savings by using online process analysis instead of laboratory measurements.
- AN-K-069Water in E-liquids – Fully automatic determination by volumetric Karl Fischer titration
The vaping and electronic cigarette industries are growing. The mixtures used in these products are usually called e-liquid, e-fluid, or e-juice. Toensure the quality of these e-liquids, testing the most important parameters is required. One important quality control parameter is water or moisture content.Water/moisture content determination by Karl Fischer titration (KFT) is an established and reliable procedure. Compared to other methods the advantages of KFT are its accuracy, speed, and selectivity. For high water content samples, such as e-liquids, volumetric KFT is the method of choice.In this Application Note a system for the fast and reliable determination of the water content in E-liquids is presented. This fully automated system performs the analysis including system preparation, blank, titer, and sample determination completely unattended. Hence, the workload of the operator is reduced to only weighing in the sample and placing the sealed sample vessels on the system.
- AN-NIR-055Determination of water content in moisturizing skin creams using near-infrared spectroscopy
Near-infrared spectroscopy (NIRS) was used for quality control of skin creams. A model for the quantification of the water content was developed based on Karl Fischer titration (KF), enabling fast and reliable atline analysis and final product quality control.
- AN-NIR-079Moisture analysis in fertilizer products
Moisture content is one of the most commonly measured properties of fertilizers. Globally, regulations for different fertilizers vary, but local legal limits ensure that the maximum amount of water must not be exceeded. Next to gravimetric methods, Karl Fischer titration is often used for accurate moisture determination.Compared to these methods, near-infrared spectroscopy (NIRS) offers unique advantages: it generates reliable results within seconds, and at the same time does not create chemical waste. This Application Note explains how NIRS can offer fast, reagent-free analysis of moisture content in various fertilizer products.
- AN-PAN-1010Online analysis of sulfuric acid and zinc sulfate in the viscose process
Monitoring sulfuric acid and zinc sulfate in the viscose wet-spinning process is essential. Online potentiometric titration and colorimetric analysis are recommended for this purpose.
- AN-P-052Trace analysis of cyanide and sulfide in aqueous samples – DC amperometric determination after ion chromatographic separation
Sulfide and cyanide are toxic anions. Their trace determination in any kind of water samples, especially in wastewater, is requied for safety reasons. However, metal traces present in the eluent can mask target anions due to complexation. The addition of a stronger complexing agent to the eluent mask these metal cations enabling interference free determaination. This application is mainly used for the analysis of cyanide and/or sulfide in water. However, it also fulfills the requirements of ASTM D2036 for the determination of total, amenable, weak acid dissociable cyanides. The determination of cyanide and sulfide require an alkaline eluent and amperometric detection. This Application Note describes a new column/eluent combination for optimized separation. The combination consists of the Metrosep A Supp 10 - 100/4.0 column and a sodium hydroxide eluent containing a trace of EDTA for transition metal complexation. This yields in better peak shape and detection limits below 0.1 μg/L.
- AN-S-378Anions in refining process water applying tandem conductivity and UV detection
ASTM D8234 describes the determination of anions in high saline water by applying suppressed conductivity followed by UV/VIS detection. This combination enables the determination of e.g. nitrite by UV detection. With conductivity detection, this quantification is not possible or difficult due to the very large chloride peak. The actual sample is a refining process liquid with a high chloride content. As the sample solution also contains organic material, Inline Dialysis is applied to protect the analytical column. The combination of the two detection modes and the Inline Dialysis option reduces manual sample preparation and substantially increases the accuracy of the analysis.
- EB-004Near-infrared and Raman spectroscopy for polymer analysis: An introduction
This e-book explains how Raman and near-infrared (NIR) spectroscopy enable rapid, nondestructive polymer analysis, ensuring high quality while reducing costs and waste.
- WP-047Optimizing the chlor-alkali process through online chemical analysis
This White Paper explores the critical role of advanced online and inline process analysis in brine chlorine operations, emphasizing their advantages over traditional methods.
- AN-PAN-1047Monitoring water content in refined products inline with NIR spectroscopy
A safer way to monitor moisture content in crude distillation unit overhead fractions is with inline near-infrared spectroscopy using the 2060 The NIR-Ex Analyzer.
- AN-PAN-1053Monitoring of DOTP production via esterification with inline analysis
This Process Application Note presents a way to closely monitor multiple parameters simultaneously during the dioctyl terephthalate production process with near-infrared spectroscopy.
- AN-S-395Forensic examination analysis with IC
Forensic institutes examine terrorist attacks and warfare agents via trace detection analysis of the used explosives and their residuals. Of particular importance is the acquisition of «chemical fingerprints» for criminal investigation departments and governmental security agencies. Institutes for public health and environmental protection analyze such compounds that can contaminate the underlying soil and infiltrate ground water.Forensic investigation with ion chromatography (IC) using suppressed conductivity detection allows a sensitive and robust determination of anionic contaminants such as chlorate, thiosulfate, thiocyanate, and perchlorate next to the common inorganic anions over a broad concentration range.
- AN-PAN-1048Inline moisture analysis in a pilot scale granulation process by NIRS
This Process Application Note provides a detailed account of the inline assessment of moisture during a pilot scale granulation process using a 2060 The NIR Analyzer.
- AN-S-354Anions in wastewater with the Eco IC
The Eco IC is an entry-level instrument that is particularly suitable for routine operations and water analysis. It is equipped with a conductivity detector and can be used both with and without chemical suppression. This Application Note describes the determination of anion content with the Metrosep A Supp 17 - 250/4.0 column. This column model is particularly suitable for water analysis at room temperature.
- WP-078Adsorbable organic fluorine (AOF) for screening of PFAS in waters
Learn about PFAS, their impact on water quality, EU Directive 2020/2184, and the benefits of AOF measurement using combustion ion chromatography (CIC).
- AN-T-213Ozone in water
Water treatment with ozone (O3) is a common procedure for the disinfection of swimming pools. It is important that a sufficient but not excessive amount of O3 is produced to disinfect the water. Otherwise, the remaining ozone could enter the swimming water, which could irritate the respiratory system or the skin of bathers.Ozone is also used in drinking and waste water treatment because it is significantly more effective than chlorine at inactivating or killing viruses and bacteria. This application note describes a method to determine the ozone concentration in water by potentiometric titration according to DIN 38408-3.
- AN-I-014Bromide in water
Bromide is ubiquitous in sea water, where it is present in concentrations of around 65 mg/L. By contrast, the maximum bromide concentration in drinking and ground water is usually less than 0.5 mg/L. A higher bromide content may indicate a contamination of the water caused by fertilizer, road salt or industrial waste water. This Application Note describes the determination of the bromide content in water via direct measurement with a Br ion-selective electrode in accordance with ASTM D1246.
- AN-I-030Dissolved oxygen in tap water
In municipal water supplies, higher dissolved oxygen (DO) content is desirable because it improves the taste of drinking water. However, high DO levels also speed up corrosion in water pipes. For this reason, industries utilize water with as little DO as possible, and add scavengers such as sodium sulfite to remove any oxygen from a water supply. Municipal water supply pipes are normally coated inside with polyphosphates to protect the metal from contact with oxygen, thus allowing higher DO contents. Therefore, monitoring the DO content online in a water supply is important to assess its DO content to either improve taste or minimize pipe corrosion. Using an optical sensor, such as the O2-Lumitrode, allows a fast and reliable determination according to ISO 17289.
- AN-T-204Permanganate index in water
The permanganate index (PMI) is a sum parameter that indicates the total load of oxidizable organic and inorganic matter in water. The substances concerned are mainly humic materials/acids that are primarily formed when dead organic material present in soil is further broken down and released into water sources. As it is an indicator of the water quality, testing of the PMI for drinking water is obligatory in many countries.For the determination, it is necessary to heat the stabilized water sample to 95 °C and higher for a stipulated time. Afterwards, the amount of permanganate that has remained after the reaction with the sample is determined titrimetrically. This sample preparation step requires considerable manual effort.In this Application Note, a fully automated procedure for the determination of the PMI according to GB/T 11892 is described, including all sample preparation steps. The gains in productivity because of a reduced manual workload are considerable.
- AN-V-213Copper in drinking water
Higher levels of copper in drinking water are usually caused by corrosive action of water leaching copper from copper pipes. While copper is an essential nutrient for the human organism, ingestion of higher concentrations have an adverse effect on human health. The current World Health Organization’s «Guidelines for Drinking-water Quality» recommend a maximum concentration of 2000 μg/L. With a limit of detection (LOD) of 0.5 μg/L, anodic stripping voltammetry is a viable, less sophisticated alternative to atomic absorption spectroscopy (AAS) for the determination of copper in drinking water. While AAS (and competing methods) can only be performed in a laboratory, anodic stripping voltammetry can be used conventionally in the laboratory or alternatively in the field with the 946 Portable VA Analyzer. The determination is carried out on the scTRACE Gold electrode.
- 8.000.6010Inline sample preparation – An effective tool for ion analysis in pharmaceutical products
By means of azide analysis in Irbesartan a simple, fast, precise and accurate ion chromatographic method for the determination of traces of inorganic contaminants in pharmaceuticals is described. Traces of toxic azides in pharmaceutical products can accurately be determined in the sub-ppb range after Metrohm Inline Matrix Elimination using isocratic ion chromatography (IC) with suppressed conductivity detection. While the azide anions are retained on the preconcentration column, the interfering pharmaceutical matrix is washed away by a transfer solution, ideally consisting of 70% methanol and 30% ultrapure water. The analytical setup provides a well-resolved azide peak and thus alleviates the common drawback of excipient interferences, especially from the nitrate anion. Calibration with azide standards is linear over the range of 5…80 ppb, providing a coefficient of determination of 0.9995. The limit of detection (LOD) and the limit of quantification (LOQ) of azide in Irbesartan are 5 and 30 µg/L respectively; the relative standard deviations (RSD) for the peak area, peak height and retention time being smaller than 3.9%. Robustness testing involved variation of column oven temperature and composition of the transfer solution and, in terms of peak area, provided RSDs smaller than 2.8% and 3.1% respectively.
- AN-I-028Dissolved oxygen in surface water
Oxygen diffuses into water sources from the air via aeration, however several factors can reduce the dissolved oxygen (DO) content in water. First, as water warms up, oxygen is released into the atmosphere. Secondly, oxygen is consumed by bacteria and other microorganisms which feed on organic material. Finally, plants can also consume oxygen in certain situations.Human-induced alterations can have a negative influence on surface water when DO values fall below crucial limits for maintaining the life supporting capacity of freshwater ecosystems. Therefore, monitoring the DO content in surface water by an optical sensor to assess its quality is important.
- AN-PAN-1056Online monitoring of sodium in industrial power plants
In power plants, corrosion is the greatest enemy. If corrosive impurities are present in the circuit streams (e.g., chlorides and hydroxides), deposition of an insulating layer of scale on the heat transfer surfaces occurs, resulting in costly and critical downtimes. To ensure high throughput of power plants, online analysis of critical parameters such as sodium is highly advantageous for safety, protection, and process optimization. With the 2035 Process Analyzer from Metrohm Process Analytics, operators gain the information they need to accurately identify trends, reduce downtimes, and address operational issues before costly problems arise.
- AN-I-013Sulfide in ground and waste water
Even in low concentration, sulfide ions cause odor and corrosion problems in ground water and waste water. They can release hydrogen sulfide in acidified water, which is toxic in even minuscule amounts. This Application Note describes the determination of sulfide concentration in water via direct measurement with the Ag/S-ISE in accordance with ASTM D4658.
- AN-V-234Tellurium(IV) in drinking water
Tellurium is one of the elements recently identified as technologically critical for photovoltaic conversion, quantum dots, as well as in thermoelectric technology, and has the potential to become a new emergent contaminant. Until now there is no guideline value in the World Health Organization’s «Guidelines for Drinking-water Quality» and in the European Drinking Water Directive for tellurium(IV) concentration in drinking water.To monitor the tellurium(IV) levels in drinking water, anodic stripping voltammetry (ASV) performed on the unmodified scTRACE Gold is recommended. This method allows determination of tellurium(IV) in the concentration range between 1 µg/L and 60 µg/L when using a 90 s deposition time. The scTRACE Gold electrode does not need extensive maintenance such as mechanical polishing. Measurements can be performed in the laboratory with the 884 Professional VA or alternatively in the field with the 946 Portable VA Analyzer.
- AN-V-226Zinc in drinking water with a glassy carbon electrode
No health-based guideline value exists for zinc. However, to maintain good quality municipal drinking water, the United States Environmental Protection Agency (US-EPA) set a maximum concentration of 5 mg/L as the limit value. Typical concentrations in surface and ground waters are between 10–40 μg/L Zn, with values up to 1 mg/L in tap water. Anodic stripping voltammetry (ASV) on the ex-situ mercury film modified glassy carbon electrode provides a less complex alternative to atomic absorption spectroscopy (AAS) for zinc determination in drinking water.
- AN-V-179Iron in boiler feed water
The iron concentration in boiler feed water has to be monitored to ensure reliable and safe operation of the water-steam circuit. Various guidelines set limits for the maximum iron content.The concentration of total iron in boiler feed water can be determined with high sensitivity using adsorptive stripping voltammetry (AdSV) using 2,3- dihydroxynaphthalene (DHN) as complexing agent. Voltammetry is a viable, less sophisticated alternative to atomic absorption spectroscopy (AAS) or inductive couple plasma (ICP) for the determination of iron with only a moderate investment in hardware required and low running costs.
- AN-T-131Calcium, magnesium, and total hardness in water
Water hardness is often determined photometrically using two different indicators and while performing the determination at two different pH values. Additionally, the determination itself is subjective, as the color change is determined by the analyst and not by an analytical device.This application note introduces a more robust option to easily assess calcium, magnesium, and total hardness in water by using the Cu-ISE and two different titrants. Sample preparation is identical for both analyses and can therefore be automated without any issues.
- AN-V-233Selenium(IV) in drinking water
The difference between the toxic and essential levels of selenium to human health are very slight. Therefore, the current provisional guideline value for selenium(IV) in the World Health Organization’s «Guidelines for Drinking-water Quality» and in the European Drinking Water Directive is set to a maximum concentration of 10 µg/L.The anodic stripping voltammetric (ASV) technique performed on the unmodified scTRACE Gold can be used to determine concentrations as low as 0.5 µg/L selenium with a 30 s deposition time. These limits can be lowered even further by increasing the deposition time. The linear range at 30 s deposition time ends at approximately 100 μg/L. The scTRACE Gold electrode does not need extensive maintenance such as mechanical polishing. Measurements can be performed in the laboratory with the 884 Professional VA or alternatively in the field with the 946 Portable VA Analyzer. This method is suited for manual or automated systems.
- AN-V-215Zinc in drinking water with the scTRACE Gold
Zinc is an essential trace element for humans. Excessive intake of zinc in higher concentrations can be harmful, however. There is no guideline value for zinc in the World Health Organization’s «Guidelines for Drinking-water Quality» because typical levels usually found in drinking water are of no concern. Anodic stripping voltammetry is a viable, less sophisticated alternative to atomic absorption spectroscopy (AAS) for the determination of zinc in drinking water. While AAS (and competing methods) can only be performed in a laboratory, anodic stripping voltammetric determinations can be used conventionally in the laboratory or alternatively in the field using with 946 Portable VA Analyzer. The determination is carried out on the scTRACE Gold electrode.
- AN-V-223Nickel and cobalt in drinking water with a Bi drop electrode
The main sources of nickel pollution are electroplating, metallurgical operations, or leaching from pipes and fittings. Catalysts for the petroleum and chemical industries are major application fields for cobalt. In both cases, the metal is either released directly, or via the waste water-river pathway into the drinking water system. Therefore in the EU the legislation specifies 20 µg/L as the limit value for the Ni concentration in drinking water.The simultaneous and straightforward determination of nickel and cobalt is based on adsorptive stripping voltammetry (AdSV). The unique properties of the non-toxic Bi drop electrode combined with AdSV results in an excellent performance in terms of sensitivity.
- AN-V-227Chromium(VI) in drinking water with a glassy carbon electrode
The guideline value for chromium in the World Health Organization’s (WHO) «Guidelines for Drinking-water Quality» is 50 µg/L. It should be noted here that chromium concentrations are often expressed as total chromium and not as chromium(III) or (VI). Chromium(VI) is responsible for changes in genetic material, and is found in significantly lower concentrations than Cr(III). Therefore an extremely sensitive method is required to monitor Cr(VI) in drinking water.The powerful adsorptive stripping voltammetry (AdSV) technique on the ex-situ mercury film modified glassy carbon electrode using DTPA as complexing agent can be used to determine such low concentrations.
- AN-T-074Conductivity, pH value, alkalinity, and hardness in tap water
In this application note, a fully automated system is presented which allows the determination of several parameters according to various standards within one analysis. These include conductivity (ISO 7888, EN 27888, ASTM D1125, EPA 120.1), the pH value (EN ISO 10523, ASTM D1293, EPA 150.1), alkalinity (EN ISO 9963, ASTM D1067, EPA 310.1), and Ca/Mg content (ISO 6059, ASTM D1126, EPA 130.2). Additionally, the system transfers the required sample volume into an external titration vessel for the analysis, reducing manual sample preparation. Furthermore, all sensors can be automatically calibrated and the titer of each titrant can also be determined.
- AN-V-225Cadmium and lead in drinking water with a glassy carbon electrode
To reduce the toxic effects of cadmium on the kidneys, skeleton, and the respiratory system, as well as the neurotoxic effects of lead, the provisional guideline values in the World Health Organization’s (WHO) «Guidelines for Drinking-water Quality» are set to a maximum concentration of 3 µg/L for cadmium and 10 µg/L for lead.The powerful anodic stripping voltammetry (ASV) technique on the ex-situ mercury film modified glassy carbon electrode is more than sufficient to monitor the proposed WHO guidelines for Cd and Pb in drinking water.
- AN-I-036Sodium content in water using an ion-selective electrode
Groundwater contains many minerals, but can be contaminated by sodium-rich leachate from landfills. Accurate Na determination in water is possible following AOAC 976.25 using the Na-ISE.
- AN-V-212Mercury in mineral water
Mercury and its compounds are toxic. The highest risk is posed by chronic poisoning with mercury compounds ingested with food. A significant part of the mercury present in the environment is of anthropogenic origin. Considerable sources are coal-fired power plants, steel, and nonferrous metal production, waste incineration plants, the chemical industry, or artisanal gold mining where the use of elemental mercury for the extraction of gold from the ore is still common. The guideline value for inorganic mercury in the World Health Organization’s «Guidelines for Drinking-water Quality» is set to 6 μg/L.With a limit of detection (LOD) of 0.5 μg/L, anodic stripping voltammetry is a viable, less sophisticated alternative to atomic absorption spectroscopy (AAS).While AAS (and competing methods) can only be performed in a laboratory, anodic stripping voltammetry can be used conventionally in the laboratory or alternatively in the field with the 946 Portable VA Analyzer. The determination is carried out on the scTRACE Gold electrode.
- AN-M-015Trace haloacetic acids, dalapon, and bromate measurement in drinking water
Chlorinating drinking water can form carcinogenic byproducts. EPA Method 557 enables µg/L-level quantification of haloacetic acids using Metrohm IC-MS/MS technology.
- AN-T-075Conductivity, pH value, alkalinity, and chloride in tap water
In this application note, a fully automated system is presented which allows the determination of several parameters according to various standards within one analysis. These include conductivity (ISO 7888, EN 27888, ASTM D1125, EPA 120.1), pH value (EN ISO 10523, ASTM D1293, EPA 150.1), alkalinity (EN ISO 9963, ASTM D1067, EPA 310.1), and chloride content (ISO 9297, ASTM D512, EPA 325.3). Additionally the system transfers the required volume of sample into an external titration vessel, further reducing manual sample preparation. Furthermore, all sensors can be calibrated automatically and the titer of each titrant can also be determined.
- AN-T-076Conductivity, pH value, alkalinity, hardness, and chloride in tap water
In this application note, a fully automated system is presented which allows the determination of several parameters according to various standards within one analysis. These include conductivity (ISO 7888, EN 27888, ASTM D1125, EPA 120.1), pH value (EN ISO 10523, ASTM D1293, EPA 150.1), alkalinity (EN ISO 9963, ASTM D1067, EPA 310.1), Ca/Mg (ISO 6059, ASTM D1126, EPA 130.2), and chloride (ISO 9297, ASTM D512, EPA 325.3). Additionally the system transfers the required volume of sample into external titration vessels for the different analyses, reducing manual sample preparation. Furthermore, all sensors can be automatically calibrated and the titer of each titrant can also be determined.
- AN-PAN-1011Online analysis of permanganate absorption number (PAN)
Permanganate absorption number (PAN) analysis per ISO 8660 ensures caprolactam purity, a precursor of Nylon 6. This application describes real-time, continuous PAN monitoring.
- AB-266Determination of titanium by adsorptive stripping voltammetry
This Application Bulletin describes the determination of titanium by adsorptive stripping voltammetry (AdSV) using mandelic acid as complexing agent. The method is suitable for the analysis of ground, drinking, sea, surface and cooling waters, in which the concentration of titanium is of importance. The methods can, of course, also be used for the trace analysis in other matrices.The limit of detection is approx. 0.5 µg/L.
- AN-NIR-096Water in diesel with NIRS
This Application Note highlights near-infrared spectroscopy as a faster, cost-effective alternative to KF titration for predicting water content in diesel fuel.
- AN-T-214Kjeldahl nitrogen in waste water
Nitrogen-based compounds are widely distributed in the environment and are essential growth nutrients for photosynthetic organisms. Therefore, it is important to monitor and control the amount of nitrogen compounds which are released into the environment.In this Application Note, a method to determine the nitrogen content in water by Kjeldahl digestion and distillation followed by a photometric or potentiometric titration according to ASTM D3590 is presented. The universality, precision, and reproducibility of the Kjeldahl method have made it the internationally recognized method for e.g. estimating the protein content in many matrices and it is the standard method to which all other methods are judged against.
- AN-NIR-078Automated moisture analysis in pharmaceutical peptides
The quantification of residual moisture in lyophilized pharmaceutical peptides is an important measure for quality control in the pharmaceutical industry. For development purposes, such measurements are necessary and routinely performed during stability studies and to optimize the freeze-drying process (lyophilization). Currently, Karl Fischer titration is widely used for moisture determination in routine analysis. However, this method is time consuming and destroys the sample during analysis. This Application Note shows that near-infrared spectroscopy (NIRS) is a fast, reagentless, non-destructive method to determine moisture content in lyophilized pharmaceutical products.
- AN-PAN-1059Online analysis of strontium and barium in high purity brine
This Process Application Note describes a method to determine the strontium and barium concentration in brine as early detectors ofmembrane fouling via online process ion chromatography. Using this multiparameter analytical technique can help reduce the risk of premature membrane fouling and avoid unexpected maintenance and high utility costs with 24/7 automated analysis.
- AN-NIR-121Water content in propylene glycol monomethyl ether (PGME)
Near-infrared spectroscopy (NIRS) can determine water content in PGME (propylene glycol monomethyl ether) within seconds as shown in this Application Note.
- AN-PAN-1035Online analysis of indigo, hydrosulfite, and other parameters in textile dye baths
The size of the indigo molecule makes it difficult to dye synthetic fibers, but the large pores of cellulose (such as in cotton) accept it readily. Indigo is insoluble in water, so it must first be reduced to the water-soluble leuco-indigo form by sodium hydrosulfite in a strong alkaline bath. Good circulation within the bath is imperative for consistent dye coverage, but care must be taken not to introduce any oxygen. Fabrics must be oxidized between dips in the dye bath in order to set the indigo within the pores of the fibers, but multiple dips are necessary for darker, uniform coverage. Many parameters need to be monitored and controlled to ensure high quality of the end product: the pH value for proper NaOH (alkali) dosage, the concentrations of both hydrosulfite and indigo, as well as the temperature of the bath and even the redox potential.
- AN-NIR-134Paprika powder analysis with NIR spectroscopy
This study shows how NIR spectroscopy simultaneously measures capsaicin content, Scoville Heat Units, water activity, ASTA color, and ash content in paprika powder samples.
- AN-NIR-132Almond analysis with near-infrared (NIR) spectroscopy
This Application Note shows how NIR spectroscopy is used to determine the water content (moisture content), protein content, and fat content in whole and ground almonds.
- AB-430Determination of uranium by adsorptive stripping voltammetry according to DIN 38406-17
This Application Bulletin describes the methods for the determination of uranium by adsorptive stripping voltammetry (AdSV) according to DIN 38406 part 17. The method is suitable for the analysis of ground, drinking, sea, surface and cooling waters, in which the concentration of uranium is of importance. The methods can, of course, also be used for the trace analysis in other matrices.Uranium is determined as chloranilic acid complex. The limit of detection in samples with low chloride concentration is about 50 ng/L and in seawater about 1 µg/L. Matrices with high chloride content can only be analyzed after reduction of the chloride concentration by means of a sulfate-loaded ion exchanger.
- AN-NIR-073Determination of water activity in tablets with the OMNIS NIR Analyzer
Water activity is an important parameter to measure for non-sterile pharmaceutical quality and stability. The OMNIS NIR Analyzer provides this data within seconds.
- AN-K-074Water content in crude oil determined with Karl Fischer titration
To determine water in crude oil, ASTM D4928 recommends coulometric Karl Fischer titration with the oven method, allowing full automation for high reproducibility.
- AN-K-053Determination of the water content in toothpaste with MATi 10
This Application Note describes the automated determination of the water content in toothpaste using volumetric Karl Fischer titration (MATi 10).
- AN-NIR-122Quantification of water in lactose with the OMNIS NIRS Analyzer
This Application Note shows how easy it is to determine water content in the pharmaceutical excipient lactose with reagent-free near-infrared spectroscopy.
- AN-PAN-1006Online analysis of zinc, sulfuric acid, and iron during zinc refining
This Process Application Note is dedicated to the online analysis of zinc, iron and sulfuric acid in several stages of the zinc production process. Additionally, traces of germanium, antimony, as well as several transition metals (e.g., Ni, Co, Cu, Cd) can be precisely determined (<50 µg/L) in the purification filtrates and reactor trains.
- WP-093Significant cost savings through dynamic ventilation during nitrification in wastewater treatment
This White Paper gives an overview of the energy-intensive nitrification process that converts ammonia into less harmful nitrogen compounds at wastewater treatment plants (WWTPs). It displays the results from a field test at a WWTP, showcasing the positive influence of single-method process analyzers on the efficiency of the nitrification process.
- AN-U-056Nitrite, bromide, and nitrate in artificial sea water applying direct UV/VIS detection
Seawater analysis with conductivity detection is difficult due to the high excess of chloride. Especially analyzing for nitrite and bromide, UV/VIS detection is preferred as chloride is not interfering with nitrite at 218 nm. This AN shows the determination of all three UV-absorbing anions in an artificial seawater.
- AB-199Determination of sulfide and sulfite by polarography
Sulfide and sulfite can be determined polarographically without any problems. For sulfide, polarography is performed in an alkaline solution, for sulfite in a slightly acidic primary solution. The method is suitable for the analysis of pharmaceuticals (infusion solutions), wastewater/flue gas water, photographic solutions, etc.
- WP-023Karl Fischer titration and near-infrared spectroscopy in perfect synergy
Metrohm has shaped moisture analysis for more than half a century. Find out about new developments in water analysis and learn how near-infrared spectroscopy in combination with Karl Fischer titration can increase your sample throughput and boost your productivity.
- AB-431Determination of iron, copper and vanadium by adsorptive stripping voltammetry
This Application Bulletin describes the voltammetric determination of the elements iron, copper and vanadium. Fe as well as Cu and V can be determined as catechol complex at the HMDE by adsorptive stripping voltammetry (AdSV). Fe(II) and Fe(III) are determined as Fe(total) with the same sensitivity for both species in either phosphate buffer or PIPES electrolyte. Cu and V can be determined in PIPES buffer.The methods are primarily suitable for the investigation of ground, drinking and surface waters, in which the concentration of these metals is important. But the methods can naturally also be used for trace analysis in other matrices.The limit of detection for all three elements in PIPES buffer is 0.5 ... 1 µg/L, for iron in phosphate buffer it is approx. 5 µg/L.
- AB-242Determination of tungsten by anodic stripping voltammetry at the Ultra Trace graphite RDE
The method described allows the determination of W(VI) traces in the range 0.2 to 50 µg/L (ppb). Traces of organic compounds present in the samples (e.g. natural waters) interfere. They have to be removed by UV digestion (e.g. 705 UV Digester). Interference by Fe(III) up to a concentration of 100 mg/L is eliminated by reduction to Fe(lI) with ascorbic acid. If the amount of Cu(II) in the sample exceeds the amount of W(VI) by a factor of 200 or more, the Cu ions have to be bound with thiourea. Moreover, the concentration of Cu(II) should not exceed 5 mg/L. The determination is made by adsorptive stripping analysis in the DP mode.
- AN-S-302Anions in small sample volumes by using the intelligent Pick-up Injection Technique (MiPuT)
The determination of anions in tap water is a simple IC application. Here it is used to present Metrohm's intelligent Pick-up Technique (MiPuT). MiPuT enables the injection of volumes of minimum size from very small sample quantities. In the present case, two volumes of 10 µL from a sample 100 µL in size are used for anion and cation analysis, respectively. The calibration takes place through the injection of various volumes of a single standard solution. AN-C-141 describes the corresponding cation determination.
- WP-014High productivity and profitability in IC environmental analysis
Brad Meadows is Vice President and Lab Director at the US company BSK Labs, which runs a number of environmental laboratories and service centers. Brad is an analytical chemist and has been working in the management of analysis laboratories for 15 years. He shared his experiences with Metrohm ion chromatography with us in the form of some concrete facts and figures.
- AN-C-141Cations in small sample volumes by using the intelligent Pick-up Injection Technique (MiPuT)
The determination of cations in tap water is a simple IC application. Here it is used to present Metrohm's intelligent Pick-up Technique (MiPuT). MiPuT enables the injection of volumes of minimum size from very small sample quantities. In the present case, two volumes of 10 µL from a sample 100 µL in size are used for anion and cation analysis, respectively. The calibration takes place through the injection of various volumes of a single standard solution. AN-S-302 describes the corresponding anion determination.
- AB-207Determination of silver by anodic stripping voltammetry at the carbon RDE
This Application Bulletin describes the stripping analysis of Ag at the rotating disk electrode (RDE) with glassy carbon tip (GC) or Ultra Trace graphite tip. In routine operation, the determination limit lies at approx. 10 μg/L Ag, with careful work 5 μg/L Ag can be obtained. After appropriate digestion, silver determination is also possible with samples containing a relatively high proportion of organic substances (e.g. wine, foodstuffs etc.). The method has been developed primarily for water samples (well, ground and wastewater, desilvering solutions of the photographic industry).
- TA-020IC-MS and IC-ICP/MS analysis in the environment
This article describes the coupling of ion chromatography with mass spectrometry (IC-MS) and plasma mass spectrometry (IC-ICP/MS) for the trace analysis of potentially hazardous compounds in the environment.
- AN-S-032ppt analysis (ng/L) in ion chromatography
Determination of chloride, nitrate, and sulfate in ultrapure water after sample preconcentration using anion chromatography with conductivity detection after chemical suppression.
- WP-039Ion chromatography – addressing the latest challenges in environmental analysis
The analytical challenges of environmental analysis increase in difficulty from year to year. As well as analysis of particularly toxic types of metals such as chromium(VI), highly diverse and partially persistent organic fluorine compounds (e.g., trifluoroacetic acid) are presently in focus. The analysis of toxic oxohalides such as bromate and perchlorate is also a current subject of investigation.
- AB-133Determination of ammonia with the ion-selective electrode – Tips and tricks for a reliable determination according to common standards
Although the known photometric methods for the determination of ammonia/ammonium are accurate, they require a considerable amount of time (Nessler method 30 min, indophenol method 90 min reaction time). A further disadvantage of these methods is that only clear solutions can be measured. Opaque solutions must first be clarified by time-consuming procedures. These problems do not exist with the ion-selective ammonia electrode. Measurements can be easily performed in waste water, liquid fertilizer, and urine as well as in soil extracts. Especially for fresh water and waste water samples several standards, such as ISO 6778, EPA 350.2, EPA 305.3 and ASTM D1426, describe the analysis of ammonium by ion measurement. In this Application Bulletin, the determination according to these standards is described besides the determination of other samples as well as some general tips and tricks on how to handle the ammonia ion selective electrode. Determination of ammonia in ammonium salts, of the nitric acid content in nitrates, and of the nitrogen content of organic compounds with the ion-selective ammonia electrode is based on the principle that the ammonium ion is released as ammonia gas upon addition of excess caustic soda:NH4+ + OH- = NH3 + H2OThe outer membrane of the electrode allows the ammonia to diffuse through. The change in the pH value of the inner electrolyte solution is monitored by a combined glass electrode. If the substance to be measured is not present in the form of an ammonium salt, it must first be converted into one. Organic nitrogen compounds, especially amino compounds are digested according to Kjeldahl by heating with concentrated sulfuric acid. The carbon is oxidized to carbon dioxide in the process while the organic nitrogen is transformed quantitatively into ammonium sulfate.
- AN-S-242Trace analysis of anions in the primary cycle of a nuclear power plant (PWR) using Metrohm Inline Sample Preparation
Determination of fluoride, glycolate, acetate, formate, chloride, nitrite, nitrate, and sulfate in the primary cycle water of a pressurized water reactor (PWR) using anion chromatography with conductivity detection after chemical suppression calibrated with Metrohm Inline Calibration.
- AN-V-082Different chromium species in sea water
Cr(III) forms an electrochemically active complex with diethylenetriaminepentaacetic acid (DTPA), so does Cr(VI) after in situ reduction on the surface of the HMDE. Depending on the sample preparation procedure and the waiting time after the addition of the complexing agent, the different chromium species can be differentiated:Total active chromium [total concentration of Cr(VI) and free Cr(III)]:The measurement is carried out immediately after the addition of DTPA.; Cr(VI): Between the addition of DTPA and the start of the analysis a minimum waiting time of 30 min is necessary. During this waiting time the Cr(III)-DTPA complex becomes electrochemically inactive.; Cr(III): The difference between the total active Cr and Cr(VI).; Totalchromium: Determination of total active Cr after UV digestion.;
- AN-NIR-131NIRS analysis of forages
NIR spectroscopy enables fast, reliable analysis of key quality parameters in alfalfa forage (e.g., protein, fiber, and moisture) without any sample preparation.
- AB-358Analysis of residual moisture in a lyophilized pharmaceutical product by near-infrared spectroscopy (NIRS)
This Application Bulletin describes the method of near-infrared spectroscopy in diffuse reflection for the purpose of determining residual moisture in a lyophilized pharmaceutical product. Numerous sample vials containing freeze-dried pharmaceuticals were spiked with varying amounts of water for calibration purposes. The resulting differences in the absorption wavelengths of the OH-oscillation were correlated with the water content determined by Karl Fischer titration using the algorithm of multiple linear regression (MLR).
- AB-413Textile analysis using near-infrared spectroscopy
The present Application Bulletin describes applications that use near-infrared spectroscopy. Each application describes the used and alternatively deployable spectrometer as well as analysis conditions and results and, where available, information on feasibility studies.
- AB-076Polarographic determination of nitrilotriacetic acid (NTA) and ethylenediaminetetraacetic acid (EDTA) according to DIN 38413 part 5
According to the described method, NTA and EDTA can be determined in mass concentrations of 0.05 mg/L up to 25 mg/L in polluted water and wastewater.At first NTA and EDTA are converted to the corresponding Bi complexes by addition of Bi3+ ions at a pH value of 2.0. As these Bi complexes have significantly different peak potentials, they can be determined simultaneously by DP polarography. The interfering anions nitrite, sulfite, and sulfide are removed from the sample by acidification and purging. Interfering cations are removed by cation exchange; any NTA or EDTA heavy metal complexes present in the sample are disintegrated during this procedure. To remove surfactants and other organic components interfering with the analysis, the sample solution is run through a column filled with non-polar adsorber resin.
- AN-U-078Silicate in cooling agents
Scale formation is a critical issue in cooling systems leading to system damage, which generates enormous operational losses. One important component of scale forming is silica. Amorphous silica and metal silicates especially tend to build up scale. Therefore, it is crucial to know the silica concentration in cooling agents. By ion chromatography with UV/VIS detection and PCR, it is possible to determine both free and total silicate content. Sample dilution in ultrapure water and direct injection delivers the free silicate concentration. The total silicate content is determined after hydrolysis of amorphous silica by sample dilution in alkaline eluent, and injection after a reaction time (e.g. 4 hours) prior to the analysis.
- AN-NIR-105Roasted coffee analysis by near-infrared spectroscopy
Near-infrared spectroscopy (NIRS) is a fast, chemical-free alternative analytical technology for caffeine and moisture analysis in roasted coffee beans and grounds.
- AN-NIR-127Feed analysis with near-infrared spectroscopy (NIRS)
NIR spectroscopy enables fast, reagent-free analysis of fat, moisture, protein, fiber, ash, and starch in animal feed, streamlining quality control with no sample prep.
- AN-NIR-123Flour analysis by near-infrared spectroscopy (NIRS)
NIR spectroscopy offers fast, chemical-free analysis of ash, protein, moisture, and rheological properties in flour – ideal for routine lab or atline quality control.
- 8.000.6101Analysis of Anions and Oxoanions using Ion Chromatography Mass Spectrometry (IC-MS)
Ion chromatography tackles difficult separation problems of various ionic species and typically works with conductivity detection. Mass detection as a secondary independent detector significantly lowers the detection limits and confirms the identity of analytes even when coeluting. This poster describes how the combination of IC-MS and automated sample preparation techniques cope with the analysis of anions and oxoanions in challenging matrices such as soil or explosion residues.
- AN-CIC-033Monitoring PFASs in water sources
AOF (adsorbable organic fluorine) is used to screen for per- and polyfluorinated alkyl substances in aqueous matrices via pyrohydrolytic combustion and ion chromatography.
- TA-016Determination of mercury and arsenic using speciation analysis (IC-ICP/MS)
The combination of ion chromatography and inductively coupled plasma mass spectrometry (IC-ICP/MS) is ideally suited for the detection of species of arsenic and mercury in their various oxidation levels and forms of chemical bonding. However, some species – as in the case of mercury – are reciprocally converted into one another during sample preparation, thus making a determination of the initial concentrations of the heavy metal species impossible. This article shows how these interconversions can be calculated with isotope dilution analysis and IC-ICP/MS in accordance with EPA method 6800.
- AN-PAN-1029Monitoring peracetic acid (PAA) in a beverage bottling facility
Precise online monitoring of peracetic acid (PPA) for beverage bottling requires a reliable stainless steel process analyzer.
- AN-NIR-053Moisture analysis of ethanol-hydrocarbon blends by Vis-NIR spectroscopy
This Application Note shows that visible near-infrared spectroscopy (Vis-NIRS) can determine water content in ethanol-hydrocarbon blends. Vis-NIRS is a fast alternative to conventional lab methods: it accelerates raw material inspection, process monitoring, and final product control.
- AN-NIR-016Near-infrared spectroscopy for monitoring a single-pot granulator
This Application Note describes the utilization possibilities of a new sensor design that permits, in combination with an NIRS XDS Process Analyzer, the determination of solvent residues in a High-Shear Granulator during the drying phase. This system configuration reduces the scattering of the density distribution of the powder samples so that it is possible, directly in the process, to model the water and solvent content precisely.
- WP-071Improving the corn to ethanol fermentation process with near-infrared spectroscopy (NIRS)
The fermentation of corn starch to produce ethanol is a complex biochemical process that requires monitoring of many different parameters (e.g., solids, pH, sugar profile, glycerol, lactic and acetic acid, and water and ethanol content). Traditional laboratory analysis using primary methods (e.g. Karl Fischer titration) takes about an hour to complete and is a limiting step for increasing plant capacity and efficiency. As a fast and non-destructive analytical technique, near-infrared spectroscopy (NIRS) can replace routine laboratory analysis, decreasing operating costs and increasing plant efficiency and capacity. This White Paper describes the capabilities of the modern analytical method near-infrared (NIR) spectroscopy for monitoring and improving the fermentation process of corn to ethanol.
- AN-NIR-129Dried pet food analysis by near-infrared spectroscopy (NIRS)
This study shows how a pre-calibrated NIRS instrument is used for multiparameter analysis of several pet food quality indicators like protein, moisture, fat, and ash.
- WP-009Determination of anions in tap water in accordance with US EPA Method 300
This article describes a simple method for the determination of seven standard anions (fluoride, chloride, nitrite, bromide, nitrate, phosphate and sulfate) in accordance with US EPA Method 300 Part A. An IC system is extended to include Inline Ultrafiltration and Inline Eluent Preparation for the analysis.
- AB-069Titrimetric methods for the chemical analysis of pasta
The quality of egg-based pasta is primarily determined by its egg content. Also of importance, however, is the water content, which influences the storage life of the product, as well as the degree of acidity which, in the case of high values, indicates undesirable acidification during processing or drying. A check of the chloride content shows whether salt has been added to the pasta.
- AB-121Determination of nitrate with the ion-selective electrode
It has been known for years that consuming too much nitrates from foodstuffs can result in cyanosis, particularly for small children and susceptible adults. According to the WHO standard, the hazard level lies at a mass concentration c(NO3-) ≥ 50 mg/L. However, more recent studies have shown that when nitrate concentrations in the human body are too high, they can (via nitrite) result in the formation of carcinogenic and even more hazardous nitrosamines.Known photometric methods for the determination of the nitrate anion are time-consuming and prone to a wide range of interferences. With nitrate analysis continually increasing in importance, the demand for a selective, rapid, and relatively accurate method has also increased. Such a method is described in this Application Bulletin. The Appendix contains a cselection of application examples where nitrate concentrations have been determined in water samples, soil extracts, fertilizers, vegetables, and beverages.
- AN-V-176Total selenium in drinking water after reduction of Se(VI) to Se(IV) with the 909 UV Digester
The Se(IV) concentration can be determined by cathodic Stripping Voltammetry (CSV) in an ammonium sulfate electrolyte. The analysis also functions in the presence of Cu. Se(IV) is determined in the first step. In order to register the entire content of Se, Se(VI) species are first reduced to Se(IV). This is handled by the 909 UV Digester at a pH value of between 7 and 9. The method requires practically no reagents and permits selenium speciation.
- AN-T-227Determination of sodium lactate
Sodium lactate is a salt form of lactic acid used in many regulated industries—therefore an accurate determination of the lactate content is required and is already covered in several norms. One such monograph by the US Pharmacopoeia (USP) results in high accuracies and well-defined titration curves but uses titrants and solvents that are more costly than necessary. In comparison, the presented modified method from Metrohm requires a 1:1 mixture of water and acetone and uses aqueous hydrochloric acid as titrant, resulting in an estimated cost reduction of 40% per titration compared to the USP method (USP–NF 2021, Issue 2). Furthermore, the time needed for each analysis is reduced to just 12% of the USP method (excluding blank determination). This Application Note presents both methods to determine lactate content and shows the results obtained on an OMNIS system.
- AN-PAN-1031Effectively monitoring hydrogen peroxide online in salmon farms
Salmon farms are becoming ever more popular. Strong population growth increases the demand for salmon which has an effect on their cultivation. Ever more fish live in limited space in open net cages. Parasites such as salmon lice can multiply quickly due to the high density of fish. One possibility of fighting the lice is to use baths with weak concentrations of H2O2. The salmon are treated in these concentrations for up to 20 minutes until the lice fall away and die. While it is true that hydrogen peroxide decomposes relatively quickly, it is nonetheless deadly for salmon in higher concentrations. The concentration must therefore always lie within the specifications during treatment. Our process analyzer requires less than two minutes per titration. This means that it permits more concentration determinations per treatment and always guarantees the correct dosage.
- AN-PAN-1002Online monitoring of cyanide and gold in gold leaching solution
Gold leaching by cyanidation requires precise monitoring of cyanide and gold. Online process analyzers perform such measurements, improving safety and compliance.
- AN-PAN-1025Online analysis of ammonia in ammonia-saturated brine
In the Solvay process, ammonium hydrogen carbonate and table salt are converted to sodium hydrogen carbonate and ammonium chloride. Heating the former yields sodium carbonate (soda), an important raw material for the soap and glass industries. Ammonia is an incipient and is regenerated almost completely through conversion of the ammonium chloride with lime milk (Ca(OH)2).A Metrohm process analyzer monitors the ammonia content in the saturated table salt solution after absorption tower, thus guaranteeing a good product yield in the carbonization tower. Additional parameters which can be determined with the analyzer in the Solvay process include: alkalinity, carbonate, chloride, calcium oxide and carbon dioxide.
- AN-PAN-1060Inline process monitoring of moisture content in tetrahydrofuran
This Process Application Note presents a method to accurately monitor low levels of moisture in tetrahydrofuran (THF) in «real-time» safely, reliably, and optimally with a 2060 The NIR Analyzer from Metrohm Process Analytics. Due to the hazardous and hygroscopic nature of THF, a single explosion-proof inline process analyzer is the preferred solution for industries to reduce chemical treatment, improve product quality, and increase profits.
- AN-PAN-1034Analysis of Bayer aluminate liquors using thermometric titration
The Bayer Process is the method used to refine alumina from bauxite ore, as smelting aluminum directly from alumina is much more cost- and energy-effective. In this process, "aluminate liquors" are created by digesting the crushed bauxite with CaO and NaOH at high temperatures. Additionally, the CaO causticizes carbonate which forms in the alkaline solution from organic degradation and CO2 absorption from the atmosphere. Contaminations are removed at various steps in the process, and the liquor is filtered from the alumina crystals before it is recycled back to the digestion step. Before the spent liquor can be reused, a determination of the concentrations of the total hydroxyl (“caustic”), carbonate, and alumina values is required. These parameters can be determined quickly via thermometric titration with the 2035 Process Analyzer.
- AN-PAN-1051Inline process monitoring of the moisture content in propylene oxide
This Process Application Note presents a method to closely monitor low levels of moisture in propylene oxide safely and reliably by using a single explosion-proof inline process analyzer.
- AN-RS-018Trace Detection of Malachite Green in Stream Water
Malachite green (MG) is a textile dye with effective fungicidal properties, however it is acutely toxic and its metabolites persist in the flesh of fish and mammals, making it a threat to the human food chain. The EU has concluded that contaminated foods containing levels higher than 2 μg/g MG constitute a credible health risk, and several countries have banned malachite green as an aquaculture additive. Despite tight regulation, seafood products contaminated with MG continue to find their way to consumers.Using Misa (Metrohm Instant SERS Analyzer) to ensure food safety, the rapid and highly sensitive detection of malachite green is achieved in a facile assay format.
- AN-PAN-1007Online analysis of peroxide in the HP-PO process
Monitoring hydrogen peroxide content online in the HP-PO production process requires a rugged explosion-proof solution like the 2060 TI Ex Proof Process Analyzer.
- AN-PAN-1018Online analysis of acids, bases, and aluminum in anodizing baths
Anodizing metal surfaces improves resistance against corrosion and wear. Etching baths can be monitored precisely online with the 2060 TI Process Analyzer or 2026 HD Titrolyzer.
- AN-T-103Fully automated determination of aluminum in aqueous solution
In many countries, the aluminum concentration in water is limited to 0.2 mg/L. This application note shows how the analysis of aluminum in water can be done fully automatically by complexometric titration with EDTA.
- AN-PAN-1068Online analysis of copper, tin, and zinc in white bronze baths by XRF
This Application Note explains how the 2060 XRF Process Analyzer enables real-time chemical monitoring of copper, tin, and zinc concentrations in white bronze plating baths.
- AN-T-223Analysis of electroplating baths
Electroplating processes are used in several different industry sectors to protect the surface quality of various products against corrosion or abrasion and significantly improve their working life. It is essential to check the bath composition on a regular basis to ensure that the process is operating correctly. Typical examples of electroplating baths include alkaline degreasing baths or acidic or alkaline baths containing metals e.g. copper, nickel, or chromium, or components like chloride and cyanide. It is crucial that the chosen analysis technique fulfills high safety standards for these kinds of analyses and produces reliable results. The OMNIS Sample Robot system automatically pipettes and analyzes aggressive electroplating bath samples on different workstations, increasing the safety in the lab. This provides more reliable results in comparison to manual titration and is more time efficient as different parameters can be analyzed in parallel.
- AN-EIS-003Electrochemical Impedance Spectroscopy (EIS) Part 3 – Data Analysis
Here, the most common circuit elements for EIS are introduced which may be assembled in different configurations to obtain equivalent circuits used for data analysis.
- AN-NIR-117Analysis of moisture, ash, carbon, and volatile content in coal by NIRS
Conventional methods used to analyze moisture, ash, fixed carbon, and volatile content in coal samples, are time consuming and costly. Near-infrared (NIR) spectroscopy is excellently suited to determine all parameters simultaneously in less than one minute without any sample preparation.
- AB-141Analysis of edible fats and oils – The seven most important parameters for quality control
As the determination of the exact content of individual glycerides in fats and oils is difficult and time-consuming, several fat sum parameters or fat indices are used for the characterization and quality control of fats and oils. Fats and oils are not only essential for cooking, they are also an important ingredient in pharmaceuticals and personal care products, such as ointments and creams. Consequently, several norms and standards describe the determination of the most important quality control parameters. This Application Bulletin describes eight important analytical methods for the following fat parameters in edible oils and fats:Determination of water content in accordance with the Karl Fischer method; Oxidation stability in accordance with the Rancimat method; Iodine value; Peroxide value; Saponification value; Acid value, free fatty acids (FFA); Hydroxyl number; Traces of nickel using polarography; Special care is taken to avoid chlorinated solvents in these methods. Also, as many of the mentioned methods as possible are automated.
- AB-083Sodium analysis by ion-selective electrode
This document explains how to measure Na ion concentration in diverse matrices with a sodium ion-selective electrode (Na-ISE) using direct measurement and standard addition.
- AN-PAN-1041Inline monitoring of free isocyanate (%NCO) content in polyurethane
Polyurethanes are a class of synthetic polymers formed by reacting liquid di/polyisocyanates and polyols with a catalyst and various additives in a reactor. Polymer properties are modified with stepwise additions of these chemicals at different points in time depending on whether the process has reached an equilibrium. NCO functional groups from unreacted isocyanates must be quenched at the end for a finished product, and this parameter must be known to determine the proper chemical mixing ratio. A fast, non-destructive real-time measurement of %NCO can be obtained by using NIR spectroscopy with a probe seated directly in the reactor.
- AN-S-391Anions in diesel applying advanced Inline Matrix Elimination
Anions in diesel, especially biodiesel, may cause harmful deposits in the engine. Determination with ion chromatography requires the transfer of the diesel anions into an aqueous solution, injectable to the IC. A typical method to transfer the anions into water is via Inline Extraction with subsequent Inline Dialysis prior to the injection (see AN-C-101 for a respective analysis of cations). In the actual Matrix Elimination method, diesel diluted with isopropanol is injected into an isopropanol stream and passed through a preconcentration column. Isopropanol washes off the diesel, and a subsequent rinsing step with ultrapure water removes excess isopropanol.
- TA-023Investigation of MRT contrast media containing gadolinium by means of IC-ICP/MS analysis
This article describes the investigation using ion chromatography and subsequent inductively coupled plasma mass spectronomy (ICP/MS) to determine the extent to which the iron(III) flocculation carried out in the context of wastewater treatment releases toxic gadolinium(III) ions as the result of recomplexing.
- 8.000.6014Determination of anions and cations in aerosols by ion chromatography
The study of adverse effects of air pollution requires semi-continuous, rapid and accurate measurements of inorganic species in aerosols and their gas phase components in ambient air. The most promising instruments, often referred to as steam collecting devices, are the Particle-Into-Liquid-Sampler (PILS) coupled to wet-chemical analyzers such as a cation and/or anion chromatograph (IC) and the Monitoring instrument for AeRosols and GAses (MARGA) with two integrated ICs. Both instruments comprise gas denuders, a condensation particle growth sampler as well as pump and control devices. While PILS uses two consecutive fixed denuders and a downstream growth chamber, the MARGA system is composed of a Wet Rotating Denuder (WRD) and a Steam-Jet Aerosol Collector (SJAC). Although the aerosol samplers of PILS and MARGA use different assemblies, both apply the technique of growing aerosol particles into droplets in a supersaturated water vapor environment. Previously mixed with carrier water, the collected droplets are continuously fed into sample loops or preconcentration columns for on-line IC analysis. While PILS has been designed to sample aerosols only, MARGA additionally determines water-soluble gases. Compared to the classical denuders, which remove gases from the air sample upstream of the growth chamber, MARGA collects the gaseous species in a WRD for on-line analysis. In contrast to the gases, aerosols have low diffusion speeds and thus neither dissolve in the PILS denuders nor in the WRD. Proper selection of the ion chromatographic conditions of PILS-IC allows a precise determination, within 4 to 5 minutes, of seven major inorganic species (Na+, K+, Ca2+, Mg2+, Cl-, NO3- and SO4 2-) in fine aerosol particles. With longer analysis times (10-15 minutes) even airborne low-molecular-weight organic acids, such as acetate, formate and oxalate can be analyzed. MARGA additionally facilitates the simultaneous determination of HCl, HNO3, HNO2, SO2 and NH3.PILS and MARGA provide semi-continuous, long-term stand-alone measurements (1 week) and can measure particulate pollutants in the ng/m3 range.
- AN-PAN-1015Calcium and sulfate in flue-gas desulfuration
This Process Application Note deals with the online monitoring of calcium and sulfate in flue gas scrubbing solutions using titration. Other contaminants that can be measured are sulfite, chloride, and chlorine. Low concentrations of heavy metals such as cadmium, zinc, copper, and lead can be measured in the ppb/ppm range with the ADI 2045VA Process Analyzer using voltammetry.
- AN-SEC-004Spectroelectrochemical analysis of electrochromic materials
Poly(3,4-ethylenedioxythiophene) (PEDOT) is one of the most promising ICPs due to its high conductivity, electrochemical stability, catalytic properties, high insolubility in almost all common solvents and interesting electrochromic properties (transparent in the doped state and colored in the neutral state). In this Application Note, PEDOT film is evaluated by spectroelectrochemical techniques.
- AN-T-025Hydrogen peroxide content in aqueous solutions
Peroxides are often used for disinfection and water treatment purposes due to their antiseptic properties. Lower concentrations between 0.3–3% are used in households, while higher concentrations can be used for sterilization purposes. Additionally, peroxides are utilized as oxidizing and bleaching agents. Peroxides, perborates, and percarbonates can easily be determined by titration. This application note presents two titration methods for peroxide analysis: ASTM D2180 for concentrated hydrogen peroxide solutions, and a second method for trace determination of hydrogen peroxide, suitable for concentrations as low as 0.4 mg/L.
- 8.000.6011Ion chromatographic determination of anions, cations and organic acids in biofuels
Quality and process control of biofuels require straightforward, fast and accurate analysis methods. Ion chromatography (IC) is at the leading edge of this effort. Traces of anions in a gasoline/ethanol blend can accurately be determined in the sub-ppb range after Metrohm Inline Matrix Elimination using anion chromatography with conductivity detection after sequential suppression. While the analyte anions are retained on the preconcentration column, the interfering organic gasoline/bioethanol matrix is washed away.Detrimental alkali metals and water-extractable alkaline earth metals in biodiesel are determined in the sub-ppm range using cation chromatography with direct conductivity detection applying automated extraction with nitric acid and subsequent Metrohm Inline Dialysis. Unlike high-molecular substances, ions in the high-ionic strength matrix diffuse through a membrane into the low-ionic water acceptor solution. In biogas reactor samples, low-molecular-weight organic acids stem from the biodegradation of organic matter. Their profile allows important conclusions concerning conversion in the anaerobic digestion reaction. Volatile fatty acids and lactate can be accurately determined by using ion-exclusion chromatography with suppressed conductivity detection after inline dialysis or filtration.
- AN-NIR-019Human stool analysis by near-infrared spectroscopy
This Application describes the determination of moisture, nitrogen, and fat in stool samples using near-infrared spectroscopy. These parameters are of great importance in medical diagnostics.
- AN-NIR-128Milk powder analysis by near-infrared spectroscopy
Near-infrared spectroscopy saves time and resources by simultaneously measuring key quality parameters like lactose, moisture, fat, and protein content in milk powder.
- AN-NIR-130Multiparameter hops analysis by near-infrared spectroscopy (NIRS)
NIRS can simultaneously measure several quality parameters in hops like cohumulone, hop oils, and moisture content, the hop storage index (HSI), and alpha and beta acids.
- AN-NIR-144Whey permeate analysis with NIRS
NIRS simultaneously measures important quality parameters in whey permeate (i.e., ash, phosphate, lactose, protein, pH, and moisture) without any sample preparation.
- WP-077Robust multiparameter analysis of infant and follow-on formulas with ion chromatography (IC)
Free white paper gives comprehensive overview of how to reliably assess the quality of infant formula with ion chromatography.
- AN-NIR-076Quality control of polyvinyl alcohol
Polyvinyl alcohol (PVA) is a linear polymer, used in a variety of medical products (e.g. eye drops). Here, the degree of alcoholysis is an important index for the water solubility, viscosity, and adhesion of the product. The degree of alcoholysis is defined as the percentage of hydroxyl functional groups compared to the total functional groups accessible in the molecule. Conventional alcoholysis determination can take up to six hours per sample. Compared to the primary method, analysis with near-infrared spectroscopy (NIRS) only takes one minute. The following application note describes the determination of the degree of alcoholysis by NIRS.