Chuyển hướng tin nhắn
Autolab PGSTAT302N

Autolab PGSTAT302N

PGSTAT302N

Autolab PGSTAT302N

vi

PGSTAT302N

Tổng quan

This high end, high current potentiostat/galvanostat, with a compliance voltage of 30 V and a bandwidth of 1 MHz, combined with our FRA32M module, is specially designed for electrochemical impedance spectroscopy.

The PGSTAT302N is the successor of the popular PGSTAT30. The maximum current is 2 A, the current range can be extended to 20 A with the BOOSTER20A, the current resolution is 30 fA at a current range of 10 nA.

Thông số kỹ thuật

Potential range in volt-10 V to 10 V
Compliance voltage range in volt30
Maximum current in ampère2
Number of current ranges9
Number of current ranges remarks10 nA to 1 A
Modular instrumenttrue
Maximum number of modules8
Analog integratorfalse
Multichannel instrumentfalse
Maximum number of channels1
Potential and current accuracy0.2 % or 0.2 % of current range
Potential resolution0.3 µV (gain 1000)
Current resolution0.0003% (of current range)
Maximum bandwidth in Hz1 MHz
Input impedance in Ohm1 TOhm
Dimensions in mm (W/H/D)520/160/420
Dimensions remarkWithout cables and accessories
Weight of the instrument in kg18
Sản phẩm này đã ngừng sản xuất, nhưng vẫn được hỗ trợ.
    • Autolab dummy cell
    • AUT.DUMCELL.S
    • Dummy cell for instrument testing.
    • <p>Dummy cell for instrument testing.</p>
    • PCS
    • 1
    • 50 cm BNC cable
    • CABLE.BNC.50
    • 50 cm BNC cable for diagnostics purposes.
    • <p>50 cm BNC cable for diagnostics purposes. </p>
    • PCS
    • 1
    • Monitor cable for N series Autolab
    • CABLE.MONITOR
    • Monitor cable for modular Autolab systems, providing connections for external equipments (Potential output (Eout), Current output (iout) and Potential input (Ein)).
    • <p>Monitor cable for modular Autolab systems, providing connections for external equipments (Potential output (E<sub>out</sub>), Current output (i<sub>out</sub>) and Potential input (E<sub>in</sub>)).</p>
    • PCS
    • 1
    • Power cable
    • CABLE.PWR
    • Standard power cable for Autolab instruments and accessories.
    • <p>Standard power cable for Autolab instruments and accessories.</p>
    • PCS
    • 1
    • Cell cable
    • CELLCBL.30.RE.S
    • Standard cell cable, 1.5 m, with connection for reference electrode (RE) and sense electrode (S).
    • <p>Standard cell cable, 1.5 m, with connection for reference electrode (RE) and sense electrode (S).</p>
    • PCS
    • 1
    • Cell cable
    • CELLCBL.30.WE.S
    • Standard cell cable, 1.5 m, with connection for counter electrode (CE), working electrode (WE) and ground.
    • <p>Standard cell cable, 1.5 m, with connection for counter electrode (CE), working electrode (WE) and ground.</p>
    • PCS
    • 1
    • Advanced software for electrochemical research
    • NOVA
    • NOVA is the package designed to control all the Autolab instruments with USB interface.Designed by electrochemists for electrochemists and integrating over two decades of user experience and the latest .NET software technology, NOVA brings more power and more flexibility to your Autolab potentiostat/galvanostat.NOVA offers the following unique features:Powerful and flexible procedure editor; Clear overview of relevant real-time data; Powerful data analysis and plotting tools; Integrated control for external devices like Metrohm Liquid Handling devices; Download the latest version of NOVA
    • <p>NOVA is the package designed to control all the Autolab instruments with USB interface.<br /><br />Designed by electrochemists for electrochemists and integrating over two decades of user experience and the latest .NET software technology, NOVA brings more power and more flexibility to your Autolab potentiostat/galvanostat.<br /><br />NOVA offers the following unique features:</p><ul><li>Powerful and flexible procedure editor</li><li>Clear overview of relevant real-time data</li><li>Powerful data analysis and plotting tools</li><li>Integrated control for external devices like Metrohm Liquid Handling devices</li></ul><p><a href="https://www.metrohm.com/en/service/software-center/nova.html">Download the latest version of NOVA</a></p>
    • PCS
    • 1
    • Autolab dummy cell
    • AUT.DUMCELL.S
    • Dummy cell for instrument testing.
    • <p>Dummy cell for instrument testing.</p>
    • PCS
    • 1
    • 50 cm BNC cable
    • CABLE.BNC.50
    • 50 cm BNC cable for diagnostics purposes.
    • <p>50 cm BNC cable for diagnostics purposes. </p>
    • PCS
    • 1
    • Monitor cable for N series Autolab
    • CABLE.MONITOR
    • Monitor cable for modular Autolab systems, providing connections for external equipments (Potential output (Eout), Current output (iout) and Potential input (Ein)).
    • <p>Monitor cable for modular Autolab systems, providing connections for external equipments (Potential output (E<sub>out</sub>), Current output (i<sub>out</sub>) and Potential input (E<sub>in</sub>)).</p>
    • PCS
    • 1
    • Power cable
    • CABLE.PWR
    • Standard power cable for Autolab instruments and accessories.
    • <p>Standard power cable for Autolab instruments and accessories.</p>
    • PCS
    • 1
    • Cell cable
    • CELLCBL.30.RE.S
    • Standard cell cable, 1.5 m, with connection for reference electrode (RE) and sense electrode (S).
    • <p>Standard cell cable, 1.5 m, with connection for reference electrode (RE) and sense electrode (S).</p>
    • PCS
    • 1
    • Cell cable
    • CELLCBL.30.WE.S
    • Standard cell cable, 1.5 m, with connection for counter electrode (CE), working electrode (WE) and ground.
    • <p>Standard cell cable, 1.5 m, with connection for counter electrode (CE), working electrode (WE) and ground.</p>
    • PCS
    • 1
    • Advanced software for electrochemical research
    • NOVA
    • NOVA is the package designed to control all the Autolab instruments with USB interface.Designed by electrochemists for electrochemists and integrating over two decades of user experience and the latest .NET software technology, NOVA brings more power and more flexibility to your Autolab potentiostat/galvanostat.NOVA offers the following unique features:Powerful and flexible procedure editor; Clear overview of relevant real-time data; Powerful data analysis and plotting tools; Integrated control for external devices like Metrohm Liquid Handling devices; Download the latest version of NOVA
    • <p>NOVA is the package designed to control all the Autolab instruments with USB interface.<br /><br />Designed by electrochemists for electrochemists and integrating over two decades of user experience and the latest .NET software technology, NOVA brings more power and more flexibility to your Autolab potentiostat/galvanostat.<br /><br />NOVA offers the following unique features:</p><ul><li>Powerful and flexible procedure editor</li><li>Clear overview of relevant real-time data</li><li>Powerful data analysis and plotting tools</li><li>Integrated control for external devices like Metrohm Liquid Handling devices</li></ul><p><a href="https://www.metrohm.com/en/service/software-center/nova.html">Download the latest version of NOVA</a></p>
    • PCS
    • 1
    • Dual-channel ultra-fast sampling module
    • ADC10M.S
    • The ADC10M module is an ultra-fast sampling module that increases the sampling rate of the Autolab from 50 kSamples/s to 10 MSamples/s giving you the ability to acquire fast transients with intervals down to 100 ns.When combined with the SCAN250 module, you can perform ultra-fast cyclic voltammetry measurements with scan rates up to a practical limit of 250 kV/s, making it a powerful tool for studying fast kinetic processes.
    • <p>The ADC10M module is an ultra-fast sampling module that increases the sampling rate of the Autolab from 50 kSamples/s to 10 MSamples/s giving you the ability to acquire fast transients with intervals down to 100 ns.<br /><br />When combined with the SCAN250 module, you can perform ultra-fast cyclic voltammetry measurements with scan rates up to a practical limit of 250 kV/s, making it a powerful tool for studying fast kinetic processes.</p>
    • PCS
    • 1
    • Black Alligator Clamp
    • ALL.CLIP.BLACK
    • Black alligator clamp for connections to electrodes in the electrochemical cell.  
    • <p><br />Black alligator clamp for connections to electrodes in the electrochemical cell. <br /> </p>
    • PCS
    • 1
    • Red Alligator Clamp 
    • ALL.CLIP.RED
    • Red alligator clamp for connections to electrodes in the electrochemical cell.  
    • <p><br />Red alligator clamp for connections to electrodes in the electrochemical cell. <br /> </p>
    • PCS
    • 1
    • Booster10A
    • Booster10A
    • The Booster10A module increases the maximum current of the PGSTAT100N, PGSTAT128N, PGSTAT302N, PGSTAT204 or M204 to 10 Ampere. The compliance voltage of the system is 20 V in combination with the Booster10A.With its fast response time, the Autolab Booster10A has been optimized to perform electrochemical impedance measurements, in combination with the FRA32M module, on fuel cells, batteries and super-capacitors. The booster is able to handle active as well as passive cells. The Booster10A can be used to measure the charge and discharge characteristics of super-capacitors, perform measurements on fuel cells or perform DC or AC measurements on large area electrodes.
    • <p>The Booster10A module increases the maximum current of the PGSTAT100N, PGSTAT128N, PGSTAT302N, PGSTAT204 or M204 to 10 Ampere. The compliance voltage of the system is 20 V in combination with the Booster10A.<br /><br />With its fast response time, the Autolab Booster10A has been optimized to perform electrochemical impedance measurements, in combination with the FRA32M module, on fuel cells, batteries and super-capacitors. The booster is able to handle active as well as passive cells. The Booster10A can be used to measure the charge and discharge characteristics of super-capacitors, perform measurements on fuel cells or perform DC or AC measurements on large area electrodes.</p>
    • PCS
    • 1
    • True linear scan generator module
    • SCAN250.S
    • When the processes exhibit very fast transient behavior, such as hydrogen adsorption, digital sweep can lead to loss of information regarding the adsorption process. The SCAN250 module, which has the capability of applying a true analog sweep to the sample, was specially designed to overcome this problem. The true linear scan rate range is from 10 mV/s to 250 kV/s. The SCAN250 module in combination with ADC10M is a very powerful tool for studying fast transients.
    • <p>When the processes exhibit very fast transient behavior, such as hydrogen adsorption, digital sweep can lead to loss of information regarding the adsorption process. The SCAN250 module, which has the capability of applying a true analog sweep to the sample, was specially designed to overcome this problem. The true linear scan rate range is from 10 mV/s to 250 kV/s. The SCAN250 module in combination with ADC10M is a very powerful tool for studying fast transients.</p>
    • PCS
    • 1
    • Software development kit
    • SDK
    • The Autolab Software Development Kit (Autolab SDK) is designed to control the Autolab instrument from different external applications such as LabVIEW, Visual Basic for Applications (VBA), scripting etc. With the Autolab SDK the external application can be used to measure complete procedures or control individual Autolab modules.In order to use the Autolab SDK from other applications, these applications must have the possibility to use .NET assemblies or in the case of 'older' applications to use COM assemblies. How to integrate these assemblies is explained in the manual of the application.The Autolab SDK is compatible with Autolab NOVA however it does not require NOVA to be installed.
    • <p>The Autolab Software Development Kit (Autolab SDK) is designed to control the Autolab instrument from different external applications such as LabVIEW, Visual Basic for Applications (VBA), scripting etc. With the Autolab SDK the external application can be used to measure complete procedures or control individual Autolab modules.<br /><br />In order to use the Autolab SDK from other applications, these applications must have the possibility to use .NET assemblies or in the case of &#039;older&#039; applications to use COM assemblies. How to integrate these assemblies is explained in the manual of the application.<br /><br />The Autolab SDK is compatible with Autolab NOVA however it does not require NOVA to be installed.</p>
    • PCS
    • 1
    • Voltage and pH measurement module
    • pX1000
    • The pX1000 allows the measurements of pH or pX values during electrochemical experiments. This module also provides an additional Pt1000 input which allows recording of the temperature during the experiments, either through a Pt1000 sensor or through a combined pH/Pt1000 sensor. The temperature measurement allows automatic pH corrections.The pX1000 module can also be used as an additional differential electrometer, with the same specifications as the main Autolab electrometer. The pX1000 module is compatible with the Metrohm pH and temperature sensors.The user can connect any pH, pX or ‘double’ electrode to the pX1000 module. In case an electrode other than a pH electrode is used, the output is given as the voltage difference that is measured between the electrodes making it possible to connect a detection electrode to perform coulometric titration. The pX1000 module also works as an independent pH meter.
    • <p>The pX1000 allows the measurements of pH or pX values during electrochemical experiments. This module also provides an additional Pt1000 input which allows recording of the temperature during the experiments, either through a Pt1000 sensor or through a combined pH/Pt1000 sensor. The temperature measurement allows automatic pH corrections.<br /><br />The pX1000 module can also be used as an additional differential electrometer, with the same specifications as the main Autolab electrometer. The pX1000 module is compatible with the Metrohm pH and temperature sensors.<br /><br />The user can connect any pH, pX or &lsquo;double&rsquo; electrode to the pX1000 module. In case an electrode other than a pH electrode is used, the output is given as the voltage difference that is measured between the electrodes making it possible to connect a detection electrode to perform coulometric titration. The pX1000 module also works as an independent pH meter.</p>
    • PCS
    • 1