Foi redireccionado para a sua versão local da página solicitada.

MiPT – Metrohm intelligent Partial Loop Technique

AN-C-117

pt_PT

<p>Calibration of lithium, sodium, ammonium, zinc, potassium, magnesium, and calcium using the partial loop technique and cation chromatography with direct conductivity detection. This technique allows a calibration range of 1:100 (e.g. 1 &mu;g/L to 100 &mu;g/L corresponding to 2 &mu;L to 200 &mu;L injected volume) out of 1 calibration solution. Applying the full range of partial loop injection to the samples one calibration covers a sample concentration range of 1 to 10'000 e.g. 2 &mu;L of a 10 mg/L solution corresponds to the highest calibration level (100 &mu;g/L) while 200 &mu;L of a 1 &mu;g/L solution corresponds to the lowest calibration level.</p>

Calibration of lithium, sodium, ammonium, zinc, potassium, magnesium, and calcium using the partial loop technique and cation chromatography with direct conductivity detection. This technique allows a calibration range of 1:100 (e.g. 1 μg/L to 100 μg/L corresponding to 2 μL to 200 μL injected volume) out of 1 calibration solution. Applying the full range of partial loop injection to the samples one calibration covers a sample concentration range of 1 to 10'000 e.g. 2 μL of a 10 mg/L solution corresponds to the highest calibration level (100 μg/L) while 200 μL of a 1 μg/L solution corresponds to the lowest calibration level.

Contato

Metrohm Portugal

R. Frei Luis de Granada 14G
1500-680 Lisboa

Contato