Ion-selective electrode, F

Ion-selective electrode, F

6.0502.150

Ion-selective electrode, F

en

6.0502.150

Visão geral

Fluoride-selective electrode with crystal membrane.
This ISE has to be used in combination with a reference electrode and is suitable for:
 

  • ion measurements of F- (10-6 mol/L to sat.)
  • ion measurements in small sample volumes (minimum immersion depth 1 mm)
  • titrations


Thanks to the robust/break-proof plastic shaft made of EP, this sensor is mechanically very resistant.

Especificações técnicas

Shaft materialEP
Measuring range (theoretical)1x10-6 ... Sat.
Measuring unitmol/L
Temperature range, long-term (°C)0 ... 80
Temperature range, short-term (°C)0 ... 80
pH range5 ... 7
Indicator electrode shapeFlat
Max. installation length (mm)125
SGJ sleeveFlexible SGJ sleeve
Shaft diameter top (mm)12
Shaft diameter bottom (mm)12
Min. immersion depth (mm)1
Electrode plug-in headMetrohm plug-in head G
Indicator electrode typeCrystal (F)
Este campo é obrigatório.
    • LL ISE reference electrode
    • 6.0750.100
    • Silver / silver chloride reference electrode with double junction system.This reference electrode is well suited for:automated applications; ion measurements; surfactant titrations; The ground-joint diaphragm, which is insensitive to contamination, offers a constant and reproducible electrolyte outflow. Additionally, the reference electrolyte is gelified for even better signal stability. The sensor is delivered with c(KCl) = 3 mol/L as bridge electrolyte, which can be freely selected and exchanged as needed.
    • <p>Silver / silver chloride reference electrode with double junction system.<br/>This reference electrode is well suited for:</p><ul><li>automated applications</li><li>ion measurements</li><li>surfactant titrations</li></ul><p>The ground-joint diaphragm, which is insensitive to contamination, offers a constant and reproducible electrolyte outflow. Additionally, the reference electrolyte is gelified for even better signal stability. The sensor is delivered with c(KCl) = 3 mol/L as bridge electrolyte, which can be freely selected and exchanged as needed.</p>
    • PCS
    • 1
    • Electrode cable / 1 m / F
    • 6.2104.020
    • For connecting electrodes with Metrohm plug-in head G to Metrohm instruments (socket F).
    • For connecting electrodes with Metrohm plug-in head G to Metrohm instruments (socket F).
    • PCS
    • 1