You have been redirected to your local version of the requested page

Total base number in motor oil with thermometric titration

AN-H-064

en

The determination of the total base number (TBN) in motor oils is accomplished by means of titration with a standard solution made up of trifluoromethanesulfonic acid in glacial acetic acid and isobutyl vinyl ether as reagent for improved end point identification.

The total base number (TBN) provides important information about the remaining amount of additives present in motor oil. Together with other parameters, TBN can be used to estimate the remaining lifetime of an oil. Thermometric titration is ideal for analyzing TBN in motor oils – especially used motor oils – as the titration is not affected by colored samples and the sensor does not require any conditioning or extra cleaning steps.

The total base number is determined by titrating the motor oil sample with a standard solution of trifluoromethanesulfonic acid in glacial acetic acid. Isobutyl vinyl ether is used as a reagent for better endpoint detection.