Screen-printed carbon electrode for conductivity measurements

Screen-printed carbon electrode for conductivity measurements

DRP-11COND-U50

  • Overblik over funktioner
  • Tekniske specifikationer

Disposable screen-printed electrodes designed to measure conductivity aqueous media in a typical range of 84 µS cm-1 to 111.8 mS cm-1. Conductivity sensors are recommended for working with microvolumes and are ideal for decentralized and ‘in situ’ assays.

Recommended instuctions to work with these SPEs:

  • Let the sample temper at 25 ºC to obtain accurate measurements
  • Place 100 μL onto the cell with the aid of a micropipette
  • Measure |Z| in ohm at 10 kHz with an EIS potentiostat
  • Subtract resistance value, r0, in ohm to the |Z| to correct contribution of the electrodes

If |Z| = 3638 Ω and r0 = 100 Ω then

|Z| – r0 = 3638 Ω – 100 Ω = 3538 Ω

  • Calculate inverted value and multiply by one million to achieve conductivity in μS

(1/3538 Ω) x 1000000 = 283 μS

  • Multiply the value by the Kcell to obtain final result in μS·cm-1

If Kcell = 5 cm-1 then

283 μS x 5 cm-1 = 1413 μS cm-1