The Metrosep A Supp 5 - 150 in the Microbore version is distinguished for its very good separation properties. The particle size of 5 µm makes a decisive contribution to the separating efficiency of this column. The Metrosep A Supp 5 - 150/2.0 offers the optimum combination of selectivity and capacity with which even complex separation tasks can be solved within a short time. The 2 mm Metrosep A Supp 5 separation columns are packed with the same material as the corresponding 4 mm separation columns. The 150 mm version of this column type is used for universal applications at low eluent consumption.

With its low eluent flow, this column is particularly suitable for IC-MS coupling.

  • Tekniska specifikationer
  • Dokument
  • Alternativ
  • Kromatogram
Carrier material Polyvinyl alcohol with quarternary ammonium groups
Column capacity 12 µmol (Cl-)
Column dimensions 150 x 2.0 mm
Eluents

Carbonate eluent (standard eluent):

Sodium hydrogen carbonate: 1.0 mmol/L; 168 mg/2 L
Sodium carbonate: 3.2 mmol/L; 678 mg/2 L
Housing material PEEK
Max. pressure 20 MPa
Maximum flow 0.21 mL/min
Organic modifier 0 - 100% (particularly acetone, acetonitrile, methanol)
Particle size 5 µm
pH range 3...12
Standard flow 0.18 mL/min
Standard temperature 25 °C
Temperatur range 20...60 °C

Ladda ner hela leveransomfattningen och extra tillbehör som PDF

  • Metrosep A Supp 5 Guard/2.0
    6.1006.600
    Metrosep A Supp 5 Guard/2.0

    The Metrosep A Supp 5 Guard/2.0 reliably protects the 2 mm versions of the Metrosep A Supp 5 and Metrosep A Supp 7 IC anion columns against contamination from the sample or eluent.
    It contains the same separation material as the Metrosep A Supp 5, is also made of PEEK, and is screwed directly onto the respective separation column with nearly no dead volume ("On Column Guard System"). The guard column prolongs the service life of the analytical column, with practically no influence on its chromatographic separating efficiency. The economical price and simple handling make using the A Supp 5 Guard/2.0 highly recommended.

  • Metrosep A Supp 5 S-Guard/2.0
    6.1006.610
    Metrosep A Supp 5 S-Guard/2.0

    The Metrosep A Supp 5 S-Guard/2.0 reliably protects the 2 mm versions of the Metrosep A Supp 5 and Metrosep A Supp 7 IC anion columns against contamination from the sample or eluent.
    It contains the same separation material as the Metrosep A Supp 5, is also made of PEEK, and is connected directly onto the respective separation column by means of capillary connection. The guard column prolongs the service life of the analytical column, with practically no influence on its chromatographic separating efficiency. The economical price and simple handling make using the A Supp 5 S-Guard/2.0 highly recommended.