元の言語のページへ戻りました

This Bulletin describes the simultaneous determination of gold and copper by potentiometric titration using an Fe(II) solution as titrant. Fe(II) reduces Au(III) directly to the free metal, whereas Cu(II) does not react. By the addition of fluoride ions the Fe(III) is complexed and a shift of the redox potential is effected. Afterwards, potassium iodide is added, thus reducing the Cu(II) to Cu(I), and the free iodine is again titrated with the Fe(II) solution using a Pt Titrode.

Chemical reactions:

Au(III) + 3 Fe(II)  → Au + 3 Fe(III)

2 Cu(II) + 2 I-   → 2 Cu(I) + I2

I2 + 2 Fe(II) → 2 I- + 2 Fe(III)

お問い合わせ

メトロームジャパン株式会社

143-0006 東京都大田区平和島6-1-1
東京流通センター アネックス9階

お問い合わせ