重定向消息

聚乙烯的质量监控

AN-NIR-081

2020-03

zh

Reliable determination of PE density within one minute using NIRS

聚乙烯 (PE) 的密度测定(ASTM D792)通常是一个具有挑战性的过程,因为很难再现。由于样品的不均匀性,当须分析较大的样品时,通过傅立叶变换红外光谱进行测量可能会出现问题。本应用说明表明,在可见光和近红外光谱区工作的 DS2500 近红外固体分析仪为测定聚乙烯密度提供了可靠而快速的解决方案。无需样品制备或化学试剂,可见近红外光谱法可在一分钟内分析较大的、不均匀的聚乙烯样品。

Determination of the density of polyethylene (PE) (ASTM D792) is normally a challenging procedure due to reproducibility difficulties. Since this parameter is most frequently used to determine the PE type, alternative methods which are sensitive to molecular structure such as FT-IR spectroscopy are also common. However, measurement via FT-IR can also be problematic when larger sample sizes must be analyzed due to sample inhomogeneity.

This application note demonstrates that the DS2500 Solid Analyzer operating in the visible and near-infrared spectral region (Vis-NIR) provides a reliable and fast solution for determination of the density of PE. With no sample preparation or chemicals needed, Vis-NIR spectroscopy allows the analysis of larger, inhomogeneous sample sizes of PE in less than a minute.

Figure 1. DS2500 Solid Analyzer and PE pellets present in the rotating DS2500 Large Sample Cup.

PE pellets were measured in reflection mode over the full wavelength range (400–2500 nm) of the DS2500 Solid Analyzer. A rotating DS2500 Large Sample Cup was employed to overcome the distribution of varied particle sizes and chemical components. This allowed automated measurements at different sample locations for a reproducible spectrum acquisition. As displayed in Figure 1, samples were measured without any preparation. The Metrohm software package Vision Air Complete was used for all data acquisition and prediction model development.

Table 1. Hardware and software equipment overview
Equipment Metrohm number
DS2500 Solid Analyzer 2.922.0010
DS2500 Large Sample Cup 6.7402.050
Vision Air 2.0 Complete 6.6072.208

The obtained Vis-NIR spectra (Figure 2) were used to create prediction models for quantification of the density content. The quality of the prediction models was evaluated using correlation diagrams, which display the relationship between Vis-NIR prediction and primary method values. The respective figures of merit (FOM) display the expected precision of a prediction during routine analysis.

Figure 2. Selection of PE Vis-NIR spectra obtained using a DS2500 Analyzer and a rotating DS2500 Large Sample Cup. For display reasons a spectra offset was applied.
Figure 3. Correlation diagram for the prediction of the density of PE using a DS2500 Solid Analyzer. The density lab value was evaluated using densimetry.
Table 2. Figures of merit for the prediction of the density of PE using a DS2500 Solid Analyzer.
Figures of merit Value
R2 0.991
Standard error of calibration 0.0005 g/mL
Standard error of cross-validation 0.0005 g/mL

This application note demonstrates that the density of PE can be determined easily with NIR spectroscopy. Since no sample preparation is needed, samples are analyzed as they are, which allows for simple operation leading to highly precise results (0.0005 g/mL, see Table 2).

咨询

瑞士万通中国

北京市海淀区上地东路1号院1号楼7层702
100085 北京

咨询