Quality control of deodorant by near-infrared spectroscopy
AN-NIR-146
2025-12
en
Determination of pH, density, viscosity, and aluminum content in a few seconds
Near-infrared spectroscopy (NIRS) can simultaneously determine different quality parameters in deodorants like viscosity, pH, density, and aluminum content.
Personal care products like deodorants are tested using various time-consuming analytical methods during production as well as after the final product is made. This Application Note demonstrates how quality parameters such as viscosity, pH, density, and aluminum content in deodorant can be measured with near-infrared spectroscopy (NIRS) simultaneously and without any sample preparation. NIRS is a fast, chemical-free technology that is very user-friendly. NIR spectroscopy can be used either atline during the production process or offline in a quality control laboratory.
Sixty samples of deodorant were measured in an OMNIS NIR Analyzer Solid (Figure 1). All measurements were performed in transflection mode (1000–2250 nm) using a 1 mm gap size large reflector and a transflection vessel.
Reference values of aluminum content were measured with titration (AN-T-228), pH values with a pH meter, and viscosity and density with a viscometer and density meter, respectively.
OMNIS Software was used for all data acquisition and prediction model development.
| Equipment | Article number |
|---|---|
| OMNIS NIR Analyzer solid | 2.1071.0010 |
| Large reflector OMNIS NIR, 1mm | 6.07402.940 |
| Transflection vessel OMNIS NIR, 60mm | 6.07402.260 |
| OMNIS Stand-Alone license | 6.06003.010 |
| Software license Quant Development | 6.06008.002 |
The obtained NIR spectra of deodorant samples (Figure 2) were used to create prediction models for pH, viscosity, density, and aluminum content. The quality of the prediction models was evaluated using correlation diagrams which display a very high correlation between the NIR prediction and the reference values. The respective figures of merit (FOM) display the expected precision of a prediction during routine analysis (Figures 3–6).
Result aluminum content
| R2 | SEC (%) | SECV (%) | SEP (%) |
|---|---|---|---|
| 0.908 | 0.12 | 0.16 | 0.14 |
Result density
| R2 | SEC (g/mL) | SECV (g/mL) | SEP (g/mL) |
|---|---|---|---|
| 0.979 | 0.0017 | 0.0020 |
0.0023 |
Result viscosity
| R2 | SEC (cP) | SECV (cP) | SEP (cP) |
|---|---|---|---|
| 0.941 | 147.05 | 176.05 | 133.74 |
Result pH
| R2 | SEC | SECV | SEP |
|---|---|---|---|
| 0.832 | 0.13 | 0.18 | 0.10 |
This Application Note shows the feasibility of using NIR spectroscopy for the analysis of aluminum content, pH, viscosity, and density in deodorants. With NIRS, the measurement can be conducted within seconds and requires no chemicals. This solution is easier, much faster than other conventional techniques, and can be used atline in several steps of the production chain or during quality control of the final product, saving manufacturers time and money.