Temperature dependency of the cation separation on the Metrosep C 6 - 150/4.0 column

AN-C-156

en

The column temperature influences the duration of the cation separation on the high-performance Metrosep C 6 - 150/4.0 column. The retention times of lithium, sodium, ammonium, magnesium, and calcium remain practically constant with increasing column temperature, whereas those of potassium, strontium, and barium are considerably shortened. This means that the temperature can be used to reduce analysis time considerably on the Metrosep C 6 - 150/4.0.

The column temperature influences the duration of the cation separation on the high-performance Metrosep C 6 - 150/4.0 column. The retention times of lithium, sodium, ammonium, magnesium, and calcium remain practically constant with increasing column temperature, whereas those of potassium, strontium, and barium are considerably shortened. This means that the temperature can be used to reduce analysis time considerably on the Metrosep C 6 - 150/4.0.