NIRS XDS Process Analyzer – MicroBundle SinglePoint

NIRS XDS Process Analyzer – MicroBundle SinglePoint
Artikkelnummer: 2.928.0110

The NIRS XDS Process Analyzer – MicroBundle provides the next generation of process analyzers for real-time analysis in the pharmaceutical and chemical industries Non-destructive, accurate measurements of practically any sample type (powders, slurries, liquids,..) are performed directly in the process line, granulator, dryer or reaction vessel. The micro bundle fiber optics allows the measuring point to be up to 35 meter away from the XDS NIR process analyzer.

  • Spesifikasjoner
  • Tilleggsutstyr

Tekniske data

Technical specifications
Wavelength range: 800 - 2,200 nm
Detector: Extended InGaAs (Indium Gallium Arsenide)
Data Increment: 0.5 nm
Average time per result: 3 - 30 seconds

General
Lifetime of the light source: = 8000 h
Power supply: 100 - 240 VAC, 50 - 60 Hz, 1 A,
Power consumption: 100 W typical, max. 300 W
Software package: Vision for instrument operation and method development
Standard Operating temperature: 0 - 35 °C (32 - 95 °F)
Humidity: 10 - 90% relative
Weight: 42.2 kg (93 lbs)
Dimensions (W x H x D): 407 x 610 x 362 mm (16 x 24 x 14.25 inches)
Housing: basic device IP65/Nema 4X, stainless steel
Various enclosure options for hazardous areas or environmental conditions outside the above mentioned conditions available.

Fiber Optics and Probes

  • Single Fiber Optic Cable with SMA connectors at the analyzer and probe (requires probe or flow-through cell for use)
  • Stainless Steel or Hastalloy Probes (other materials available as well)
    Micro Transmission Probe Pair
    Single-sided Transmission Probe
    Purge and 45 degree angle probe
    Transmission flow-through cell


Process Communication Options

  • OPC
  • Digital protocols
    MODBUS, PROFIBUS, etc.
  • VISION I/O Modules
    Available with 8, 16, or 24 Analog Channels, 24 Digital Channels

Last ned en komplett deleliste i PDF:

  • 6.7440.480
    NIRS Process, I/O Communication box

    I/O Communication box for process instruments with 24 analog and 24 digital channels

  • NIRS 99% reflection and wavelength standard process, 1 inch diameter
    6.7450.070
    NIRS 99% reflection and wavelength standard process, 1 inch diameter

    NIRS 99% reflection and wavelength standard for process instrument with 1 inch probe diameter

  • 6.7450.080
    NIRS 99% reflection and wavelength standard for process instrument, 0.5 inch probe diameter

    NIRS 99% reflection and wavelength standard for process instrument, 0.5 inch probe diameter

  • 6.7450.090
    NIRS 99% reflection and wavelength standard for process instrument, 1 inch diameter, 45° bevel

    NIRS 99% reflection and wavelength standard for process instrument with 1 inch diameter and 45° angle