Hamilton RCX-30 - 150/4.6

Hamilton RCX-30 - 150/4.6
Order number: 6.1018.010

The Hamilton RCX-30 - 150/4.6 is a column for the separation of monosaccharides, disaccharides, oligosaccharides and sugar alcohols. It is an anion exchange column based on polystyrene/divinylbenzene resin. The RCX-30 - 150/4.6 can be used for universal applications.

The Hamilton RCX-30 - 150/4.6 separation column exhibits an outstanding separation of fructose and lactose. The column also offers the advantage that flows of up to 2 mL/min can be used in order to accelerate the chromatography. The column is used both for the rapid separation of small carbohydrates and for the separation of oligosaccharides.

  • Tech specs
  • Documents
  • Applications
  • Opt. equipment
  • Chromatogram
Carrier material Styrene-divinylbenzene copolymer
Column dimensions 150 x 4.6 mm
Eluents

Hydroxide eluent (standard eluent)

Sodium hydroxide (c = 20 mol/L): 150 mmol/L; 15 mL/2 L
(column temperature 32 °C)

Hydroxide eluent modified

Sodium hydroxide (c = 20 mol/L): 200 mmol/L; 20 mL/2 L
(column temperature 32 °C)
Housing material PEEK
Max. pressure 34 MPa
Maximum flow 2.0 mL/min
Particle size 7 µm
pH range 1...13 (T > 35°C max. pH 8)
Standard flow 1.0 mL/min
Temperatur range 20...60 °C