Metrosep C 3 - 250/4.0

Metrosep C 3 - 250/4.0
Beställningsnummer: 6.1010.430

The innovative substrate on a polyvinyl alcohol base increases selectivity for monovalent and divalent cations significantly. A characteristic of this is the number of "theoretical plates per meter". On the Metrosep C 3 - 250/4.0, for example, 42,000 plates are achieved for sodium, 51,000 for ammonium, and 31,000 for barium with its delayed elution. The peak forms on this cation column are highly symmetrical.

The selectivity of the Metrosep C 3 - 250/4.0 also permits the separation of transition metals. Because Metrohm ion chromatographs generally determine the cations without chemical suppression, the transition metals can be analyzed on the Metrosep C 3 - 250/4.0 together with the alkaline and earth alkaline metals.

  • Tekniska specifikationer
  • Dokument
  • Applikationer
  • Alternativ
  • Kromatogram
Carrier material Polyvinyl alcohol with carboxyl groups
Column capacity 30 µmol (K+)
Column dimensions 250 x 4.0 mm
Eluents

Nitric acid eluent (standard eluent)

Nitric acid (c = 1 mol/L): 5.0 mmol/L; 10 mL/2 L
(column temperature 40 °C)

Nitric acid/crown ether eluent

Nitric acid (c = 1 mol/L): 3.5 mmol/L; 7 mL/2 L
18-crown-6: 0.5 mmol/L; 264 mg/2 L
(column temperature 40 °C)
Housing material PEEK
Max. pressure 15 MPa
Maximum flow 1.5 mL/min
Organic modifier 0 - 50% acetonitrile, 0 - 30% acetone, no methanol
Particle size 5 µm
pH range 2...12
Standard flow 1.0 mL/min
Temperatur range 20...40 °C

Ladda ner hela leveransomfattningen och extra tillbehör som PDF

  • Metrosep C 3 Guard/4.0
    6.1010.450
    Metrosep C 3 Guard/4.0


    The Metrosep C 3 S-Guard/4.0 contains the C-3 column material and is used to protect Metrosep C-3 cation columns. Particles and contaminations are reliably retained, allowing the lifetime of the analytical separation column to be prolonged considerably. The Metrosep C 3 Guard/4.0 also functions according to the "On Column Guard System" and is screwed directly onto the separation column with nearly no dead volume.


  • Metrosep C 3 S-Guard/4.0
    6.1010.460
    Metrosep C 3 S-Guard/4.0


    The Metrosep C 3 S-Guard/4.0 contains the C-3 column material and is used to protect Metrosep C-3 cation columns. Particles and contaminations are reliably retained, allowing the lifetime of the analytical separation column to be prolonged considerably.