Metrosep A Supp 10 - 250/2.0

Metrosep A Supp 10 - 250/2.0
Order number: 6.1020.230

The Metrosep A Supp 10 - 250/2.0 separation column is based on a high-capacity polystyrene-divinylbenzene copolymer with a particle size of only 4.6 µm. This proven column concept optimized by Metrohm is characterized by its robust construction, great selectivity and outstanding separating efficiencies. The 2 mm Metrosep A Supp 10 separation columns are packed with the same material as the corresponding 4 mm separation columns. Temperature, flow and eluent composition can be used to modify the properties of the column to accommodate the current application directly.

The Metrosep A Supp 10 - 250/2.0 has a very high capacity for a microbore column. It is suitable for samples with high ionic strength, for complex separation tasks and for analyses in which great differences in concentration between the individual components are present. Thanks to its low flow, this microbore separation column is ideally suitable for IC-MS applications.

  • Tech specs
  • Documents
  • Included parts
  • Opt. equipment
  • Chromatogram
Carrier material Polystyrene-divinylbenzene copolymer with quaternary ammonium groups
Column capacity 24 µmol (Cl)
Column dimensions 250 x 2.0 mm
Eluents

Carbonate eluent (standard eluent)

Sodium hydrogen carbonate: 5.0 mmol/L; 840 mg/2 L
Sodium carbonate: 5.0 mmol/L; 1,060 mg/2 L
(column temperature 45 °C)

Ammonium sulfate eluent

Ammonium sulfate: 250 mmol/L; 66.0 g/2 L
Ammonium hydroxide: (c = 5.0 mol/L): 100 mmol/L; 40 mL/2 L
Housing material PEEK
Max. pressure 25 MPa
Maximum flow 0.7 mL/min
Organic modifier 0 - 100%
Particle size 4.6 µm (mono dispersive)
pH range 0...14
Standard flow 0.25 mL/min
Standard temperature 45 °C
Temperatur range 10...70 °C