Metrosep C Supp 2 - 100/4.0

Metrosep C Supp 2 - 100/4.0
Order number: 6.01053.410

The Metrosep C Supp 2 separation material is based on a polystyrene/divinylbenzene copolymer with carboxyl groups. It is suitable for the separation and determination of monovalent and divalent cations. The Metrosep C Supp 2 - 100/4.0 column is the shortest separation column in the Metrosep C Supp 2 product range. It is especially suitable for trace analysis of standard cations. Limits of quantification below the µg/L range are achieved thanks to the extremely low baseline noise following sequential suppression.

  • Tech specs
  • Documents
  • Included parts
  • Opt. equipment
Carrier material Polystyrene-divinylbenzene copolymer with carboxyl groups
Column capacity 23 µmol (K+)
Column dimensions 100 x 4.0 mm
Eluents Nitric acid eluent (standard eluent)

Nitric acid (c = 1 mol/L): 5.0 mmol/L; 10 mL/2 L
Rubidium (from rubidium nitrate): 50 µg/L Rb+; 172.5 µg/2 L (RbNO3)
Housing material PEEK
Max. pressure 25 MPa
Maximum flow 3.8 mL/min
Organic modifier Eluent: 0 - 100% acetone and acetonitrile (no alcohol)

Sample: 0 - 100% acetone, acetonitrile and alcohols
Particle size 5 µm
pH range 0...12
Standard flow 1.0 mL/min
Standard temperature 40 °C
Temperatur range 10...60 °C
  • Brochure: Cation suppression in ion chromatography – Cation determination in the trace range (english)
    • Brochure: Cation suppression in ion chromatography – Cation determination in the trace range (français)
    • Brochure: Cation suppression in ion chromatography – Cation determination in the trace range (deutsch)
  • Catalog: The column program 2018 (english)
    • Catalog: The column program 2018 (deutsch)
  • Flyer: Metrosep C Supp 2 Cations in Power Plants – Ultratrace level cation analysis in power plant water  (english)
  • Leaflet for 6.01053.4x0 Metrosep C Supp 2 - XXX/4.0, German, English, French, Spanish
  • Manual Metrosep C Supp 2 (english)
    • Manual Metrosep C Supp 2 (deutsch)