Aquatrode Plus with Pt1000 (fixed cable)

Aquatrode Plus with Pt1000 (fixed cable)
Order number: 6.00257.000

Combined pH electrode with integrated Pt1000 temperature sensor and fixed cable (2.0 m, banana plug 2 mm) for pH measurement/titration in low-ion aqueous media (e.g., drinking water, process water), and also exhibits an exceptionally rapid response time in these samples.
The fixed ground-joint diaphragm is insensitive to contamination.
Storage in storage solution is recommended when c(KCl) = 3 mol/L is used as bridge electrolyte.
The bridge electrolyte can be replaced with a chloride-free electrolyte (e.g., potassium nitrate c(KNO3) = 1 mol/L (6.2310.010)), storage in used electrolyte.

  • Tech specs
  • Documents
  • Applications
  • Included parts
  • Opt. equipment
Bridge electrolyte type c(KCl) = 3 mol/L
Diaphragm Ground-joint diaphragm
Electrode plug-in head Fixed cable (l = 2 m) with plug F + 2 x B
Electrode slope > 0.97
Electrode zero point (mV) ± 15
Electrolyte flow rate (μl/h) 5...20
Indicator electrode shape Sphere
Max. installation length (mm) 135
Measuring range 0...13
Measuring unit pH
Membrane glass type A
Membrane resistance 80...200
Min. immersion depth (mm) 20
pH range 0...13
Reference electrolyte type c(KCl) = 3 mol/L (thickened)
Reference system LL system
Resistance (kOhm) 50
SGJ sleeve Flexible SGJ sleeve
Shaft diameter bottom (mm) 12
Shaft diameter top (mm) 12
Shaft material Glass
Temperature range, long-term (°C) 0...60
Temperature range, short-term (°C) 0...60
Temperature sensor Pt1000 (2 mm)