Column stability of the Metrosep A Supp 5 - 250/2.0 with Inline Ultrafiltration and Inline Eluent Preparation

Main document

The column stability of the microbore version of the Metrosep A Supp 5 - 250/2.0 was determined in long-term laboratory tests. Two injection series each were run on six days in a row. Each series was comprised of nine tap water injections, three check standard injections and six tap water injections. The IC system was shut down on the seventh day of each series. As a whole, the IC system ran over 12 weeks and counted a total of 2,650 injections. The results show an outstanding reproducibility and verify the high column stability.