Metrosep Carb 2 - 100/4.0

Metrosep Carb 2 - 100/4.0
Order Number: 6.1090.410

The Metrosep Carb 2 - 100/4.0 IC column is particularly suitable for the determination of carbohydrates using alkaline eluents and pulsed amperometric detection. The high-capacity anion exchanger column is based on a styrol-divinylbenzene copolymer. It is stable in the range of pH = 0 to pH = 14 and separates glucose, fructose and sucrose. It is also suitable for the analysis of some sugar alcohols and oligosaccharides. Short analysis times can be achieved on the 100 mm version of the Metrosep Carb 2 separation column.

  • Tech specs
  • Documents
  • Included parts
  • Options
  • Chromatogram
Carrier material Polystyrene-divinylbenzene copolymer with quaternary ammonium groups
Column dimensions 100 x 4.0 mm
Eluents Hydroxide/acetate eluent (standard eluent)

Sodium hydroxide (c = 20 mol/L): 100 mmol/L; 10 mL/2 L
Sodium acetate: 10 mmol/L; 1640.7 mg/2 L
Housing material PEEK
Max. pressure 20 MPa
Maximum flow 1.6 mL/min
Organic modifier In the eluent: 0 - 50% acetonitrile or methanol
In the sample: 0 - 100% acetone, acetonitrile or methanol
Particle size 5.0 µm
pH range 0...14
Standard flow 0.8 mL/min
Standard temperature 30°C
Temperatur range 20...60 °C