Spectroelectrochemical VIS-NIR Instrument (350-1050 nm)

Spectroelectrochemical VIS-NIR Instrument (350-1050 nm)
Order number: SPELEC1050

SPELEC1050 combines in only one box a Lightsource, a Bipotentiostat/Galvanostat and a Spectrometer (UV-VIS wavelength range: 350-1050 nm) with a software that allows optical and electrochemical experiments synchronization.

  • Tech specs
  • Documents
  • Opt. equipment
Current accuracy ≤0.5% of current range at 100 nA to 10 mA
Current resolution 0.1% of current output range
Dimensions in mm (W/H/D) 24 x 11 x 25
Light Source (LS) wavelength range 215-400 nm (deuterium) ; 360-2500 nm (tungsten halogen)
LS Bulb life >1,000 hours @ 240 nm (time) ; < 50% @ 240 nm (decrease of intensity) ; Continuous operation (testing conditions)
LS Fiber optic connector SMA 905
LS Ignition delay <2.0 seconds (delay for cold start-up may be longer)
LS Stability 1.0% peak-to-peak (over 4 hours) after 30-minute warm-up
LS Time to stable output 10 minutes (deuterium) ; 1 minute (tungsten halogen)
Maximum current in Ampère ±40 mA
Maximum number of channels 1
Maximum number of WE sharing AUX and REF 2
Measured current resolution 0.025 % of current range
Measured potential resolution 0.012 % of potential range
Multichannel instrument NO
Number of current ranges 8
Number of current ranges remarks ±1 nA to ±10 mA
Number of potential ranges (galvanostat) 2
Number of potential ranges remarks (galvanostat) ±100 mV to ±1 V
Operating modes Bipotentiostat ; Potentiostat ; Galvanostat
PC interface USB
Potential accuracy ±0.2%
Potential range in volt ±4V
Potential resolution 1 mV
Power 12 V DC
Spectrometer (SPM) Detector Linear silicon CCD array
SPM A/D resolution 16 bit
SPM Dark noise 50 RMS counts
SPM Dynamic range 8.5 x 10 rised to 7 (system) ; 1300:1 for a single acquisition
SPM Integration time 1 ms to 65 seconds
SPM Optical resolution 0.3+10.0 nm FWHM
SPM Pixel size 14 µm x 200 µm
SPM Pixel well depth 62,500 electrons
SPM Pixels 2048
SPM Signal-to-noise ratio 250:1 (at full signal)
SPM Stray light ≤0.05 % at 600 nm; <0.10 % at 435 nm
SPM Wavelength range 350-1050 nm
Weight of the instrument in kg 1.95

Download the complete Scope of Delivery and Optional Accessories for as PDF:

  • DropView SPELEC Software
    DropView SPELEC
    DropView SPELEC Software

    DropView SPELEC is a Spectroelectrochemical software that controls SPELEC instrument, offering a perfect synchronization of the optical and electrochemical measurements, as well as advanced tools for data treatment.