Primatrode with NTC (IP67, fixed cable 1.2 m)

Primatrode with NTC (IP67, fixed cable 1.2 m)
Order number: 6.0228.020

Combined pH electrode with integrated temperature sensor (NTC) and fixed cable conforming to IP67 (1.2 m).
This electrode is suitable for an introduction to GLP-compliant pH measurements in solutions that do not contain precipitates, proteins, or sulfides. This electrode is mechanically resistant thanks to the robust/unbreakable plastic shaft made of polypropylene and impact protection for the glass membrane.
Additionally, this electrode offers a waterproof plug for mobile use with Metrohm pH meters.
Reference electrolyte: c(KCl) = 3 mol/L, storage in storage solution.

  • Tech specs
  • Documents
  • Applications
  • Included parts
  • Opt. equipment
Diaphragm Ceramic diaphragm
Electrode plug-in head Fixed cable with plug I (IP67)
Electrode slope > 0.97
Electrode zero point (mV) ± 15
Electrolyte flow rate (μl/h) 3...10
Indicator electrode shape Hemisphere
Max. installation length (mm) 113
Measuring range 0 ... 14
Measuring unit pH
Membrane glass type T
Membrane resistance 200 ... 400
Min. immersion depth (mm) 15
pH range 0 ... 14
Reference electrolyte type c(KCl) = 3 mol/L
Reference system LL system
Resistance (kOhm) 5
SGJ sleeve Flexible SGJ sleeve
Shaft diameter bottom (mm) 12
Shaft diameter top (mm) 12
Shaft material PP
Shaft material (completion) Glass/Viton
Temperature range, long-term (°C) 0 ... 80
Temperature range, short-term (°C) 0 ... 80
Temperature sensor NTC (2 mm)

Download the complete Scope of Delivery and Optional Accessories for as PDF:

  • Sleeve with SGJ 14/12 mm
    6.1236.050
    Sleeve with SGJ 14/12 mm
    Sleeve with SGJ 14/12 mm, polyethylene.

    1 piece(s)

  • 
            
    Storage vessel
            
     
    6.2008.040
    Storage vessel  


    Together with 6.2008.050 storage vessel holder provides a support for the electrode on 6.3032.XXX dosing units.


    1 piece(s)

Download the complete Scope of Delivery and Optional Accessories for as PDF:

  • Electrolyte  3 mol/L KCl (250 mL)
    6.2308.020
    Electrolyte 3 mol/L KCl (250 mL)
    Electrolyte solution c(KCl) = 3 mol/L, (for Ag/AgCl reference systems)
  • Storage solution
    6.2323.000
    Storage solution
    Storage solution for all combined pH glass electrodes with reference electrolyte c(KCl) = 3 mol/L
  • pHit kit
    6.2325.000
    pHit kit

    Maintenance kit for pH electrodes

    The kit contains:

    • 50 mL cleaning solution
    • 50 mL 3M KCl solution
    • 50 mL storage solution
    • 2 Storage vessels
    • Instructions for use
  • Cleaning solution 3 x 50 mL
    6.2325.100
    Cleaning solution 3 x 50 mL

    Reliable measuring results over long periods of time can be guaranteed only if the pH glass membrane and the diaphragm receive preventive and regular care. Cleaning by etching with toxic chemicals or applying mechanical treatment to the diaphragm is not only complicated and expensive, it also accelerates the ageing of the pH glass electrode.
    The cleaning solution was developed for easy and gentle cleaning of pH glass electrodes. Regular use can considerably prolong their service life.

    This cleaning solution is also part of the pHit kit.