Metrosep Carb 2 - 100/2.0

Metrosep Carb 2 - 100/2.0
Katalogové číslo: 6.01090.210

The Metrosep Carb 2 - 100/2.0 IC column is the short Microbore version of the Metrosep Carb 2 columns and is particularly suitable for the determination of carbohydrates using alkaline eluents and pulsed amperometric detection. The high-capacity anion exchanger column is based on a styrol-divinylbenzene copolymer. It is stable in the range of pH = 0 to pH = 14 and separates glucose, fructose and sucrose. It is also suitable for the analysis of some sugar alcohols and oligosaccharides. Short analysis times can be achieved on the 100 mm version of the Metrosep Carb 2 separation column.

With its low eluent flow, this column is particularly suitable for IC-MS coupling.
 

  • Technické specifikace
  • Dokumenty
  • Součástky
  • Možnosti
  • Chromatogram
Carrier material Polystyrene-divinylbenzene copolymer with quaternary ammonium groups
Column dimensions 100 x 2.0 mm
Eluents Hydroxide/acetate eluent (standard eluent)

Sodium hydroxide (c = 20 mol/L): 100 mmol/L; 10 mL/2 L
Sodium acetate: 10 mmol/L; 1,640.7 mg/2 L

Sodium chloride eluent

Sodium chloride: 10 g/L; 20 g/2 L
Housing material PEEK
Max. pressure 20 MPa
Maximum flow 0.7 mL/min
Organic modifier In the eluent: 0 - 50% acetonitrile or methanol
In the sample: 0 - 100% acetone, acetonitrile or methanol
Particle size 5.0 µm
pH range 0...14
Standard flow 0.2 mL/min
Standard temperature 30 °C
Temperatur range 20...60 °C