The bridge electrolyte forms an electrical bridge between the measuring solution and the reference electrode. In many cases a strongly acidic measuring solution is used in CVS analysis. Over time, the acid may also reach the bridge electrolyte and thereby change the reference potential. This is why changing the bridge electrolyte every day is recommended.
The reference electrodes filled with KCl gel are maintenance-free, which means that no electrolyte needs to/may be changed.