TiNet (latest version: 2.5)
Updates and Downloads |
TiNet 2.5 - new driver for Coulometer and
pH Meter (756_780.zip,
227 KB)
The new drivers for 756 and 831 Coulometer and the
780 pH Meter is available for download.
- Support of 756 firmware version 12.
- Support of 831 firmware version 11.
- 780: Calibration data are stored correctly in
the TiNet configuration.
Installation: Download,
unzip and copy all files to the directory \tinet2\bin.
|
TiNet 2.4 and new models of Titrino and Coulometer
(24_titrino.zip, 1.11
MB) The new drivers for Titrino model 794, 795, 798, 799
and Coulometer 831 are available for download.
Installation: Download,
unzip and follow the instruction in the readme file.
|
TiNet 2.4 - New Titrino drivers
(titrino_driver.zip, 922
kB) The new Titrino drivers should be used with TiNet
2.4 including SR2. The new drivers
- avoid a parser error during pH calibration on a
sample changer.
- can use a DIS block for external Dosimat with a
Titrino without external Dosimat.
- use a correct determination time for the automatic
drift correction in Conditioning mode.
Installation: Download,
unzip and copy all files to the directory \tinet2\bin.
|
TiNet 2.4 SR2 / 756 Coulometer: "Error 42" bug
fixed
When a KF method is being sent to the 756 Coulometer
by TiNet 2.4 SR2 (A.701.0005), the instrument displays "Error
42". TiNet activates the function to read out the oven
parameter by means of the second COM. The system halts.
Installation: Download
the zipped
new driver dating from 30.08.2000 and unzip it
to get the actual driver (756.mdv). Copy this file
to your \tinet2\bin\ directory.
|
Technical
Information |
|
|
Possible Errors with TiNet 2.3
1. Database read error
| Problem: |
On PC systems with country settings
English (British): During read results
a read error appears. Results can not be read in.
|
| Temporary solution: |
Use country settings German (Swiss). |
| Solution: |
Service Pack (see below). |
2. Time out error with 758 KFD Tirino
| Problem: |
During a Karl Fischer titration with
the 758 KFD Titrino a "time out" error
appears. |
| Temporary solution: |
Use the 736 driver with the 758 KFD
Titrino. |
| Solution: |
Service Pack (see below). |
3. Installation of a MOXA card
| Problem: |
Installation of the MOXA card driver
from the CD fails. |
| Solution: |
Use the latest driver (version 3.7)
which can be dowloaded from www.moxa.com. |
Service Pack
There is a Service Pack that will fix these
problems. The Service Pack will be distributed via our
Web site. Its release will be announced on this News
Board.
Sorry for the trouble |
|
|
 |
|
|
|
|
|